Range and dose verification in proton therapy using proton-induced positron emitters and recurrent neural networks (RNNs)

被引:27
作者
Liu, Chuang [1 ]
Li, Zhongxing [1 ]
Hu, Wenbin [2 ]
Xing, Lei [3 ]
Peng, Hao [1 ,3 ,4 ]
机构
[1] Wuhan Univ, Dept Med Phys, Wuhan 430072, Hubei, Peoples R China
[2] Wuhan Univ, Dept Comp Sci, Wuhan 430072, Hubei, Peoples R China
[3] Stanford Univ, Dept Radiat Oncol, Stanford, CA 94305 USA
[4] Hokkaido Univ, Global Inst Collaborat Res & Educ GI CoRE, Sapporo, Hokkaido, Japan
关键词
proton therapy; positron emitter; range verification; dose verification; recurrent neural networks; IN-BEAM PET; PROMPT-GAMMA; VIVO VERIFICATION; PATIENT; SYSTEM; DISTRIBUTIONS; SIMULATION; DELIVERY; CAMERA; YIELDS;
D O I
10.1088/1361-6560/ab3564
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Online proton range/dose verification based on measurements of proton-induced positron emitters is a promising strategy for quality assurance in proton therapy. Because of the nonlinear correlation between the dose distribution and the activity distribution of positron emitters, we aim to establish their relationship using recurrent neural network models (LSTM, BiLSTM, GRU, BiGRU and Seq2seq). Simulations were carried out with a spot-scanning proton system using Geant4- 10.3 toolkit and a CT-based patient phantom. The 1D distributions of positron emitters and radiation dose were obtained. Training data were modeled for different beam energy, irradiation positions and counting statistics. The prediction accuracy of range and dose were quantitatively studied. The impact of including anatomical information (HU values in CT images) on the prediction performance was investigated. The BiGRU demonstrates the most stable and accurate performance with good capability of generalization, especially with the inclusion of anatomical information. When the signal-to-noise ratio (SNR) of the 1 D activity profiles is about 3, the range accuracy can be within 0.5 mm and the dose accuracy close to the peak region is <5% (relative uncertainty between prediction and raw input for all datasets). The feasibility of proton range and dose verification using the RNN-based framework was demonstrated. The RNN-based framework promises to provide a reliable and effective way for online monitoring, quality assurance and ultimately allows for adaptive proton therapy.
引用
收藏
页数:16
相关论文
共 42 条
  • [1] Extension and validation of an analytical model for in vivo PET verification of proton therapy-a phantom and clinical study
    Attanasi, F.
    Knopf, A.
    Parodi, K.
    Paganetti, H.
    Bortfeld, T.
    Rosso, V.
    Del Guerra, A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) : 5079 - 5098
  • [2] Attix H F, 2004, INTRO RADIOLOGICAL P
  • [3] Implementation and initial clinical experience of offline PET/CT-based verification of scanned carbon ion treatment
    Bauer, Julia
    Unholtz, Daniel
    Sommerer, Florian
    Kurz, Christopher
    Haberer, Thomas
    Herfarth, Klaus
    Welzel, Thomas
    Combs, Stephanie E.
    Debus, Juergen
    Parodi, Katia
    [J]. RADIOTHERAPY AND ONCOLOGY, 2013, 107 (02) : 218 - 226
  • [4] LEARNING LONG-TERM DEPENDENCIES WITH GRADIENT DESCENT IS DIFFICULT
    BENGIO, Y
    SIMARD, P
    FRASCONI, P
    [J]. IEEE TRANSACTIONS ON NEURAL NETWORKS, 1994, 5 (02): : 157 - 166
  • [5] Real-time prompt gamma monitoring in spot-scanning proton therapy using imaging through a knife-edge-shaped slit
    Bom, Victor
    Joulaeizadeh, Leila
    Beekman, Freek
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (02) : 297 - 308
  • [6] Direct time-of-flight for quantitative, real-time in-beam PET: a concept and feasibility study
    Crespo, Paulo
    Shakirin, Georgy
    Fiedler, Fine
    Enghardt, Wolfgang
    Wagner, Andreas
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (23) : 6795 - 6811
  • [7] Dose quantification from in-beam positron emission tomography
    Enghardt, W
    Parodi, K
    Crespo, P
    Fiedler, F
    Pawelke, J
    Pönisch, F
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S96 - S98
  • [8] The reliability of proton-nuclear interaction cross-section data to predict proton-induced PET images in proton therapy
    Espana, S.
    Zhu, X.
    Daartz, J.
    El Fakhri, G.
    Bortfeld, T.
    Paganetti, H.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (09) : 2687 - 2698
  • [9] Fernandes O, 2014, SYMP LARG DATA ANAL, P59, DOI 10.1109/LDAV.2014.7013205
  • [10] On the effectiveness of ion range determination from in-beam PET data
    Fiedler, Fine
    Shakirin, Georgy
    Skowron, Judith
    Braess, Henning
    Crespo, Paulo
    Kunath, Daniela
    Pawelke, Joerg
    Poenisch, Falk
    Enghardt, Wolfgang
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (07) : 1989 - 1998