Fluorescence suppression in Raman spectroscopy using a time-gated CMOS SPAD

被引:123
作者
Kostamovaara, Juha [1 ]
Tenhunen, Jussi [2 ]
Kogler, Martin [2 ]
Nissinen, Ilkka [1 ]
Nissinen, Jan [1 ]
Keranen, Pekka [1 ]
机构
[1] Univ Oulu, Dept Elect Engn, Elect Lab, FIN-90014 Oulu, Finland
[2] VTT Elect, Oulu 90570, Finland
来源
OPTICS EXPRESS | 2013年 / 21卷 / 25期
基金
芬兰科学院;
关键词
REJECTION;
D O I
10.1364/OE.21.031632
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A Raman spectrometer technique is described that aims at suppressing the fluorescence background typical of Raman spectra. The sample is excited with a high power (65W), short (300ps) laser pulse and the time position of each of the Raman scattered photons with respect to the excitation is measured with a CMOS SPAD detector and an accurate time-to-digital converter at each spectral point. It is shown by means of measurements performed on an olive oil sample that the fluorescence background can be greatly suppressed if the sample response is recorded only for photons coinciding with the laser pulse. A further correction in the residual fluorescence baseline can be achieved using the measured fluorescence tails at each of the spectral points. (C) 2013 Optical Society of America
引用
收藏
页码:31632 / 31645
页数:14
相关论文
共 20 条
  • [1] Bronzi D., 2012, ESSDERC 2012 - 42nd European Solid State Device Research Conference, P230, DOI 10.1109/ESSDERC.2012.6343375
  • [2] Fluorescence rejection in resonance Raman spectroscopy using a picosecond-gated intensified charge-coupled device camera
    Efremov, Evtim V.
    Buijs, Joost B.
    Gooijer, Cees
    Ariese, Freek
    [J]. APPLIED SPECTROSCOPY, 2007, 61 (06) : 571 - 578
  • [3] Is photobleaching necessary for Raman imaging of bone tissue using a green laser?
    Golcuk, Kurtulus
    Mandair, GuiJit S.
    Callender, Andrew F.
    Sahar, Nadder
    Kohn, David H.
    Morris, Michael D.
    [J]. BIOCHIMICA ET BIOPHYSICA ACTA-BIOMEMBRANES, 2006, 1758 (07): : 868 - 873
  • [4] Prospects for in vivo Raman spectroscopy
    Hanlon, EB
    Manoharan, R
    Koo, TW
    Shafer, KE
    Motz, JT
    Fitzmaurice, M
    Kramer, JR
    Itzkan, I
    Dasari, RR
    Feld, MS
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (02) : R1 - R59
  • [5] Short-wave infrared excited spatially offset Raman spectroscopy (SORS) for through-barrier detection
    Hopkins, Rebecca J.
    Pelfrey, Suzanne H.
    Shand, Neil C.
    [J]. ANALYST, 2012, 137 (19) : 4408 - 4410
  • [6] Itzler M. A., 2010, P SPIE, V7780
  • [7] Jestel N. L., 2005, PROCESS ANAL TECHNOL
  • [8] Wide-Range Time-to-Digital Converter With 1-ps Single-Shot Precision
    Keranen, Pekka
    Maatta, Kari
    Kostamovaara, Juha
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (09) : 3162 - 3172
  • [9] Khalil OS, 1999, CLIN CHEM, V45, P165
  • [10] Effective suppression of fluorescence light in Raman measurements using ultrafast time gated charge coupled device camera
    Martyshkin, DV
    Ahuja, RC
    Kudriavtsev, A
    Mirov, SB
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (03) : 630 - 635