Micro-dose Paranasal Computed Tomography is Valuable When Evaluating Diseases of the Paranasal Sinuses

被引:0
作者
Ozbayrak, Mustafa [1 ]
Ertas, Burak [2 ]
Ulus, Ozden Sila [3 ]
Ozdilek, Alper [1 ]
Karaarslan, Ercan [2 ]
机构
[1] Istanbul Univ, Cerrahpasa Fac Med, Dept Radiol, Istanbul, Turkey
[2] Acibadem Maslak Hosp, Clin Otorhinolaryngol, Istanbul, Turkey
[3] Acibadem Mehmet Ali Aydinlar Univ, Dept Radiol, Fac Med, Istanbul, Turkey
来源
HASEKI TIP BULTENI-MEDICAL BULLETIN OF HASEKI | 2019年 / 57卷 / 02期
关键词
Paranasal sinus; radiation; computed tomography; microdose; MULTIDETECTOR CT; RECONSTRUCTION; RELIABILITY;
D O I
10.4274/haseki.galenos.2018.4594
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim: We explored whether a reduction in the radiation dose affected the quality of diagnostic images generated using third-generation dual-source computed tomography (CT) running advanced third-generation iterative reconstruction software. Methods: From February to May 2015, 72 consecutive patients aged 16 to 64 years with clinically suspected sinusitis were randomly and prospectively enrolled and divided into two groups (standard or micro-dose CT). Anatomical landmarks and mucosal structures were independently evaluated by a radiologist and an otorhinolaryngologist using a four-point grading scale. Mean scores were calculated and compared. Results: All images were of diagnostic quality. The mean effective radiation dose was significantly lower in patients evaluated via the micro-dose protocol (group 2) than in those evaluated using the standard protocol (group 1) (p<0.05). The mean anatomical scores showed that all landmarks were very well-defined, with the exceptions of the uncinate attachment point and the ethmoid foramen (used to identify the ethmoid artery). The Wilcoxon signed rank test revealed no significant difference between the mean scores of the two observers for either group (p>0.05). Conclusion: The radiation dose to the paranasal sinus delivered during CT in patients with suspected sinusitis can be reduced to 0.003 mSv using third-generation dual-source CT running advanced iterative reconstruction software.
引用
收藏
页码:142 / 146
页数:5
相关论文
共 18 条
  • [1] Low-dose computed tomography of the paranasal sinuses: radiation doses and reliability analysis
    Abul-Kasim, Kasim
    Strombeck, Anita
    Sahlstrand-Johnson, Pernilla
    [J]. AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2011, 32 (01) : 47 - 51
  • [2] Reliability of High-Pitch Ultra-Low-Dose Paranasal Sinus Computed Tomography for Evaluating Paranasal Sinus Anatomy and Sinus Disease
    Aksoy, Elif Ayanoglu
    Ozden, Sila Ulus
    Karaarslan, Ercan
    Unal, Omer Faruk
    Tanyeri, Hasan
    [J]. JOURNAL OF CRANIOFACIAL SURGERY, 2014, 25 (05) : 1801 - 1804
  • [3] Computed Tomography of the Lung in the High-Pitch Mode Is Breath Holding Still Required?
    Baumueller, Stephan
    Alkadhi, Hatem
    Stolzmann, Paul
    Frauenfelder, Thomas
    Goetti, Robert
    Schertler, Thomas
    Plass, Andre
    Falk, Volkmar
    Feuchtner, Gudrun
    Scheffel, Hans
    Desbiolles, Lotus
    Leschka, Sebastian
    [J]. INVESTIGATIVE RADIOLOGY, 2011, 46 (04) : 240 - 245
  • [4] Bongartz G., 2004, European Guidelines for Multislice Computed Tomography Funded by the European Commission. Contract number FIGMCT2000- 20078-CT-TIP
  • [5] Multidetector CT of the paranasal sinus: Potential for radiation dose reduction
    Brem, Matthias H.
    Zamani, Amir A.
    Riva, Roberto
    Zou, Kelly H.
    Rumboldt, Zoran
    Hennig, Friedrich F.
    Kikinis, Ron
    Norbash, Alexander M.
    Schoepf, U. Joseph
    [J]. RADIOLOGY, 2007, 243 (03) : 847 - 852
  • [6] Reducing the radiation dose for low-dose CT of the paranasal sinuses using iterative reconstruction: Feasibility and image quality
    Bulla, Stefan
    Blanke, Philipp
    Hassepass, Frederike
    Krauss, Tobias
    Winterer, Jan Thorsten
    Breunig, Christine
    Langer, Mathias
    Pache, Gregor
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (09) : 2246 - 2250
  • [7] Radiation dose to the lens in sequential and spiral CT of the facial bones and sinuses
    Czechowski, J
    Janeczek, J
    Kelly, G
    Johansen, J
    [J]. EUROPEAN RADIOLOGY, 2001, 11 (04) : 711 - 713
  • [8] Tailored low dose three-dimensional CT of paranasal sinuses
    Damilakis, J
    Prassopoulos, P
    Mazonakis, M
    Bizakis, J
    Papadaki, E
    Gourtsoyiannis, N
    [J]. CLINICAL IMAGING, 1998, 22 (04) : 235 - 239
  • [9] Dammann F, 2007, RADIOLOGE, V47, P576, DOI 10.1007/s00117-007-1502-z
  • [10] Ultralow-Dose Chest Computed Tomography for Pulmonary Nodule Detection First Performance Evaluation of Single Energy Scanning With Spectral Shaping
    Gordic, Sonja
    Morsbach, Fabian
    Schmidt, Bernhard
    Allmendinger, Thomas
    Flohr, Thomas
    Husarik, Daniela
    Baumueller, Stephan
    Raupach, Rainer
    Stolzmann, Paul
    Leschka, Sebastian
    Frauenfelder, Thomas
    Alkadhi, Hatem
    [J]. INVESTIGATIVE RADIOLOGY, 2014, 49 (07) : 465 - 473