A simple technique to improve calculated skin dose accuracy in a commercial treatment planning system

被引:33
作者
Wang, Lilie [1 ]
Cmelak, Anthony J. [1 ]
Ding, George X. [1 ]
机构
[1] Vanderbilt Univ, Med Ctr, Dept Radiat Oncol, Nashville, TN 37232 USA
关键词
Eclipse AAA; model-based dose calculation; Monte Carlo calculations; skin dose; MODULATED RADIATION-THERAPY; MEGAVOLTAGE PHOTON BEAMS; CANCER; RADIOTHERAPY; SURFACE; NECK; HEAD; TOMOTHERAPY; FILM;
D O I
10.1002/acm2.12275
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiation dermatitis during radiotherapy is correlated with skin dose and is a common clinical problem for head and neck and thoracic cancer patients. Therefore, accurate prediction of skin dose during treatment planning is clinically important. The objective of this study is to evaluate the accuracy of skin dose calculated by a commercial treatment planning system (TPS). We evaluated the accuracy of skin dose calculations by the anisotropic analytical algorithm (AAA) implemented in Varian Eclipse (V.11) system. Skin dose is calculated as mean dose to a contoured structure of 0.5 cm thickness from the surface. The EGSnrc Monte Carlo (MC) simulations are utilized for the evaluation. The 6, 10 and 15 MV photon beams investigated are from a Varian TrueBeam linear accelerator. The accuracy of the MC dose calculations was validated by phantom measurements with optically stimulated luminescence detectors. The calculation accuracy of patient skin doses is studied by using CT based radiotherapy treatment plans including 3D conformal, static gantry IMRT, and VMAT treatment techniques. Results show the Varian Eclipse system underestimates skin doses by up to 14% of prescription dose for the patients studied when external body contour starts at the patient's skin. The external body contour is used in a treatment planning system to calculate dose distributions. The calculation accuracy of skin dose with Eclipse can be considerably improved to within 4% of target dose by extending the external body contour by 1 to 2 cm from the patient's skin. Dose delivered to deeper target volumes or organs at risk are not affected. Although Eclipse treatment planning system has its limitations in predicting patient skin dose, this study shows the calculation accuracy can be considerably improved to an acceptable level by extending the external body contour without affecting the dose calculation accuracy to the treatment target and internal organs at risk. This is achieved by moving the calculation entry point away from the skin.
引用
收藏
页码:191 / 197
页数:7
相关论文
共 29 条
[1]   Validation of Monte Carlo calculated surface doses for megavoltage photon beams [J].
Abdel-Rahman, W ;
Seuntjens, JP ;
Verhaegen, F ;
Deblois, F ;
Podgorsak, EB .
MEDICAL PHYSICS, 2005, 32 (01) :286-298
[2]   Evaluation of superficial dosimetry between treatment planning system and measurement for several breast cancer treatment techniques [J].
Akino, Yuichi ;
Das, Indra J. ;
Bartlett, Gregory K. ;
Zhang, Hualin ;
Thompson, Elizabeth ;
Zook, Jennifer E. .
MEDICAL PHYSICS, 2013, 40 (01)
[3]  
[Anonymous], 2014, VARIAN PHASE SPACE F
[4]   The impact of patient immobilization devices on skin dose during IMRT: A radiochromic EBT film dosimetry study in phantom [J].
Chiu-Tsao, S. ;
Sim, S. ;
Chan, M. F. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01) :S659-S659
[5]   Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer [J].
Chung, HT ;
Jin, HS ;
Dempsey, JF ;
Liu, CH ;
Palta, J ;
Suh, TS ;
Kim, SY .
MEDICAL PHYSICS, 2005, 32 (08) :2682-2689
[6]   Modeling the TrueBeam linac using a CAD to Geant4 geometry implementation: Dose and IAEA-compliant phase space calculations [J].
Constantin, Magdalena ;
Perl, Joseph ;
LoSasso, Tom ;
Salop, Arthur ;
Whittum, David ;
Narula, Anisha ;
Svatos, Michelle ;
Keall, Paul J. .
MEDICAL PHYSICS, 2011, 38 (07) :4018-4024
[7]  
Court LE, 2008, J APPL CLIN MED PHYS, V9, P2008
[8]   Accurate skin dose measurements using radiochromic film in clinical applications [J].
Devic, S ;
Seuntjens, J ;
Abdel-Rahman, W ;
Evans, M ;
Olivares, M ;
Podgorsak, EB ;
Vuong, T ;
Soares, CG .
MEDICAL PHYSICS, 2006, 33 (04) :1116-1124
[9]  
Gagnon W F, 1979, Med Phys, V6, P285, DOI 10.1118/1.594583
[10]   Influence of the Couch and Immobilization Device used during Stereotactic Body Radiotherapy on Skin Dose [J].
Gao, S. ;
Chang, J. Y. ;
Welsh, J. ;
Komaki, R. ;
Balter, P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03) :S682-S682