A Retrospective Study of 1526 Cases of Transcatheter Occlusion of Patent Ductus Arteriosus

被引:22
作者
Jin, Mei [1 ]
Liang, Yong-Mei [1 ]
Wang, Xiao-Fang [1 ]
Guo, Bao-Jing [1 ]
Zheng, Ke [1 ]
Gu, Yan [1 ]
Lyu, Zhen-Yu [1 ]
机构
[1] Capital Med Univ, Beijing Anzhen Hosp, Dept Pediat Cardiol, Beijing 100029, Peoples R China
关键词
Congenital Heart Disease; Devices; Patent Ductus Arteriosus; Transcatheter Occlusion; PERCUTANEOUS CLOSURE; OCCLUDER; CHILDREN; EXPERIENCE; DEVICE;
D O I
10.4103/0366-6999.163398
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Patent ductus arteriosus (PDA) is one of the most common congenital heart diseases and began to get treated by transcatheter occlusion since 1997 in China. Since then, several devices have been invented for occluding PDA. This study aimed to evaluate the technical feasibility, safety, and efficacy of transcatheter occlusion of PDA with different devices. Methods: One thousand five hundred and twenty-six patients (537 boys, 989 girls) with PDA from January 1997 to September 2014 underwent descending aortogram and transcatheter occlusion procedure. We retrospectively analyzed data of these patients, including gender, age, weight, size and morphology of PDA, and devices used in transcatheter occlusion, outcomes, and postoperational complications. Results: Median age and median weight were 4.0 years (range: 0.3-52.0 years old) and 15.3 kg (range: 4.5-91.0 kg), respectively. Mean ductal diameter, aortic ductal diameter, ductal length, and pulmonary artery pressure were 3.50 +/- 2.15 mm, 10.08 +/- 2.46 mm, 7.49 +/- 3.02 mm, and 30.21 +/- 17.28 mmHg, respectively. Morphology of PDA assessed by descending aortogram was of type A in 1428 patients, type B in 6 patients, type C in 79 patients, type D in 4 patients, and type E in 9 patients according to the classification of Krichenko. Of all the 1526 patients, 1497 patients underwent transcatheter PDA closure, among which 1492 were successful. Devices used were Amplatzer duct occluder I (ADO I, 1280, 85.8%), Cook detachable coils (116, 7.8%), ADO II (ADO II, 68, 4.6%), muscular VSD occluder (12, 0.8%), and Amplatzer vascular plug (16, 1.0%). Conclusions: Excellent occlusion rates with low complication rates were achieved with all devices regardless of PDA types. With transcatheter occlusion technique and devices developing, more patients with PDA can be treated with transcatheter closure both safely and efficiently.
引用
收藏
页码:2284 / 2289
页数:6
相关论文
共 21 条
[1]   Pulmonary Hypertension Associated With Congenital Heart Disease Pulmonary Vascular Disease: The Global Perspective [J].
Adatia, Ian ;
Kothari, Shyam S. ;
Feinstein, Jeffrey A. .
CHEST, 2010, 137 (06) :52S-61S
[2]   Comparison of the Efficacy of Different-Sized Amplatzer Duct Occluders (I, II, and II AS) in Children Weighing Less Than 10 kg [J].
Baspinar, Osman ;
Irdem, Ahmet ;
Sivasli, Ercan ;
Sahin, Derya Aydin ;
Kilinc, Metin .
PEDIATRIC CARDIOLOGY, 2013, 34 (01) :88-94
[3]   Transcatheter closure of large patent ductus arteriosus with severe pulmonary arterial hypertension: Short and intermediate term results [J].
Bhalgat, Parag S. ;
Pinto, Robin ;
Dalvi, Bharat V. .
ANNALS OF PEDIATRIC CARDIOLOGY, 2012, 5 (02) :135-140
[4]   The Amplatzer duct occluder: Experience in 209 patients [J].
Bilkis, AA ;
Alwi, M ;
Hasri, S ;
Haifa, AL ;
Geetha, K ;
Rehman, MA ;
Hasanah, I .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (01) :258-261
[5]   PERCUTANEOUS CLOSURE OF THE SMALL (LESS-THAN-2.5 MM) PATENT DUCTUS-ARTERIOSUS USING COIL EMBOLIZATION [J].
CAMBIER, PA ;
KIRBY, WC ;
WORTHAM, DC ;
MOORE, JW .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 69 (08) :815-816
[6]  
de Mello Junior Walter Teixeira, 2008, Interact Cardiovasc Thorac Surg, V7, P1183, DOI 10.1510/icvts.2008.180430
[7]   Trans-Catheter Closure of Patent Ductus Arteriosus-What Is the Best Device? [J].
Ghasemi, Abdolrahim ;
Pandya, Swati ;
Reddy, Surendranath Veeram ;
Turner, Daniel R. ;
Du, Wei ;
Navabi, Mohamad Ali ;
Mirzaaghayan, Mohammad Reza ;
Kiani, Abdolrazagh ;
Sloan, Katherine ;
Forbes, Thomas J. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (05) :687-695
[8]   RESULTS OF ANTEROGRADE TRANSCATHETER CLOSURE OF PATENT DUCTUS-ARTERIOSUS USING SINGLE OR MULTIPLE GIANTURCO COILS [J].
HIJAZI, ZM ;
GEGGEL, RL .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 74 (09) :925-929
[9]   Patent ductus arteriosus associated with pulmonary hypertension and desaturation [J].
Khositseth, Anant ;
Wanitkun, Suthep .
CARDIOLOGY JOURNAL, 2012, 19 (05) :543-546
[10]   Age-Related Clinical Characteristics of Isolated Congenital Unilateral Absence of a Pulmonary Artery [J].
Koga, Hiroshi ;
Hidaka, Tomoko ;
Miyako, Kenichi ;
Suga, Naohiro ;
Takahashi, Noboru .
PEDIATRIC CARDIOLOGY, 2010, 31 (08) :1186-1190