Clinical accuracy of non-contact infrared thermometer from umbilical region in children: A new side

被引:5
作者
Apa, Hursit [4 ]
Gozmen, Salih [1 ]
Keskin-Gozmen, Sukran [2 ]
Aslan, Fatma [4 ]
Bayram, Nuri [3 ]
Devrim, Ilker [3 ]
机构
[1] Dr Behcet Uz Childrens Training & Res Hosp, Div Pediat Hematol, Izmir, Turkey
[2] Dr Behcet Uz Childrens Training & Res Hosp, Div Pediat Nephrol, Izmir, Turkey
[3] Dr Behcet Uz Childrens Training & Res Hosp, Div Pediat Infect Dis, Izmir, Turkey
[4] Dr Behcet Uz Childrens Training & Res Hosp, Dept Pediat, Izmir, Turkey
关键词
umbilical region; non-contact infrared thermometer; children; TEMPERATURE-MEASUREMENT; PEDIATRIC PRACTICE; SKIN THERMOMETER; AXILLARY; MANAGEMENT; GUIDELINE; INFANTS; FEVER;
D O I
10.24953/turkjped.2016.02.009
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Measurement from axillary site with digital thermometer has been accepted as the most accurate method. But this method is time consuming. Tympanic and forehead measurements are often used but don't always seem to be more appropriate. Another site, umbilical region, could be an alternative site. This study aims to compare the measurements with axillary digital thermometer and non-contact infrared thermometers at sites from umbilicus and forehead to determine whether umbilical site could be used accurately in children. For each method, 2,048 measurements in total were performed. Using axillary method as gold standard, with a cut-off temperature of 38 degrees C, the sensitivities and specificities, positive and negative predictive values of umbilical and forehead temperatures and area under the ROC curve were determined in non obese children. There was a significant positive correlation between axillary and umbilical temperatures with a correlation coefficient of 0.78. The average difference between the mean of both axillary and umbilical temperatures was -0.47 +/- 0.65 degrees C. The Bland-Altman plot showed good accuracy with only 2.5 % of the readings falling outside the 95% level of confidence. Umbilical measurements showed sensitivity of 71.7% and specificity of 95.8%. The area under the ROC curve was 0.93. The easy application may lead noncontact measurements from umbilicus site to be the preferable method for health care providers, but agreement limits mentioned in this study should be considered.
引用
收藏
页码:180 / 186
页数:7
相关论文
共 21 条
[1]  
Apa H, 2013, PEDIATR EMERG CARE, V29, P992, DOI 10.1097/PEC.0b013e3182a2d419
[2]   PRACTICE GUIDELINE FOR THE MANAGEMENT OF INFANTS AND CHILDREN 0 TO 36 MONTHS OF AGE WITH FEVER WITHOUT SOURCE [J].
BARAFF, LJ ;
BASS, JW ;
FLEISHER, GR ;
KLEIN, JO ;
MCCRACKEN, GH ;
POWELL, KR ;
SCHRIGER, DL .
ANNALS OF EMERGENCY MEDICINE, 1993, 22 (07) :1198-1210
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Elemental mercury exposure: An evidence-based consensus guideline for out-of-hospital management [J].
Caravati, E. Martin ;
Erdman, Andrew R. ;
Christianson, Gwenn ;
Nelson, Lewis S. ;
Woolf, Alan D. ;
Booze, Lisa L. ;
Cobaugh, Daniel J. ;
Chyka, Peter A. ;
Scharman, Elizabeth J. ;
Manoguerra, Anthony S. ;
Troutman, William G. .
CLINICAL TOXICOLOGY, 2008, 46 (01) :1-21
[5]   Performance of non-contact infrared thermometer for detecting febrile children in hospital and ambulatory settings [J].
Chiappini, Elena ;
Sollai, Sara ;
Longhi, Riccardo ;
Morandini, Liana ;
Laghi, Anna ;
Osio, Catia Emilia ;
Persiani, Mario ;
Lonati, Silvia ;
Picchi, Raffaella ;
Bonsignori, Francesca ;
Mannelli, Francesco ;
Galli, Luisa ;
de Martino, Maurizio .
JOURNAL OF CLINICAL NURSING, 2011, 20 (9-10) :1311-1318
[6]  
CORK RC, 1983, ANESTH ANALG, V62, P211
[7]  
Falzon A, 2003, ACTA PAEDIATR, V92, P309
[8]  
Fischler MP, 1997, SCHWEIZ MED WSCHR, V127, P864
[9]   CORE TEMPERATURE-MEASUREMENT IN ADULTS - A LITERATURE-REVIEW [J].
FULBROOK, P .
JOURNAL OF ADVANCED NURSING, 1993, 18 (09) :1451-1460
[10]   CORE TEMPERATURE MEASURED IN THE AURICULAR CANAL - COMPARISON BETWEEN 4 DIFFERENT TYMPANIC THERMOMETERS [J].
JAKOBSSON, J ;
NILSSON, A ;
CARLSSON, L .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1992, 36 (08) :819-824