Feasibility of improving cone-beam CT number consistency using a scatter correction algorithm

被引:19
|
作者
Li, Jun [1 ]
Yao, Weiguang [2 ]
Xiao, Ying [1 ]
Yu, Yan [1 ]
机构
[1] Thomas Jefferson Univ, Dept Radiat Oncol, Philadelphia, PA 19107 USA
[2] Eastern Hlth Canc Care Program, Dept Med Phys, St John, NF, Canada
来源
关键词
cone-beam computed tomography; cone-beam CT number; scatter correction; DOSE CALCULATION; COMPUTED-TOMOGRAPHY; RADIOTHERAPY;
D O I
10.1120/jacmp.v14i6.4346
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The study was to explore the feasibility of improving cone-beam CT (CBCT) number (corresponding to the Hounsfield units in computed tomography) consistency using a scatter-correction algorithm, with the aim of using CBCT images for treatment planning with density correction. A scatter-correction algorithm was applied to a Varian OBI CBCT and an Elekta XVI CBCT, and was evaluated for improving CBCT number consistency. CBCT numbers of phantom materials were compared between images with and without bolus, which introduced additional scatter, and with and without scatter-correction processing. It was observed that CBCT numbers were different in the images with and without bolus in both CBCT studies, and the differences were reduced remarkably after scatter-correction processing. Results showed that CBCT number consistency was significantly improved by use of the scatter-correction algorithm.
引用
收藏
页码:167 / 176
页数:10
相关论文
共 50 条
  • [1] Feasibility of Improving Cone Beam CT Number Consistency with a Scatter-Correction Algorithm
    Li, J.
    Yao, W.
    Xiao, Y.
    Yu, Y.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [2] Correction of scatter in megavoltage cone-beam CT
    Spies, L
    Ebert, M
    Groh, BA
    Hesse, BM
    Bortfeld, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03): : 821 - 833
  • [3] Scatter Correction for Cone-Beam CT in Radiation Therapy
    Zhu, L.
    Xing, L.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [4] Correction of veiling glare and scatter in cone-beam CT
    Baba, R
    Ueda, K
    Takagi, H
    RADIOLOGY, 1999, 213P : 453 - 453
  • [5] A Hybrid Scatter Correction Method for Cone-Beam CT
    Zhang, Feng
    Yan, Bin
    Li, Jianxin
    Li, Lei
    Jia, Pengxiang
    Hu, Guoen
    2010 3RD INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2010), VOLS 1-7, 2010, : 23 - 27
  • [6] Noise suppression in scatter correction for cone-beam CT
    Zhu, Lei
    Wang, Jing
    Xing, Lei
    MEDICAL PHYSICS, 2009, 36 (03) : 741 - 752
  • [7] Scatter correction for cone-beam CT in radiation therapy
    Zhu, Lei
    Xie, Yaoqin
    Wang, Jing
    Xing, Lei
    MEDICAL PHYSICS, 2009, 36 (06) : 2258 - 2268
  • [8] Noise Suppression in Scatter Correction for Cone-Beam CT
    Xing, L.
    Wang, J.
    Zhu, L.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [9] Scatter Corrected Cone-Beam CT for Radiomics in SBRT: A Feasibility Study
    Jiang, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E676 - E676
  • [10] An efficient Monte Carlo-based algorithm for scatter correction in keV cone-beam CT
    Poludniowski, G.
    Evans, P. M.
    Hansen, V. N.
    Webb, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (12): : 3847 - 3864