Aerosol Delivery Through an Adult High-Flow Nasal Cannula Circuit Using Low-Flow Oxygen

被引:44
作者
Madney, Yasmin M. [1 ]
Fathy, Maha [2 ]
Elberry, Ahmed A. [3 ]
Rabea, Hoda [1 ]
Abdelrahim, Mohamed E. A. [1 ,4 ]
机构
[1] Beni Suef Univ, Clin Pharm Dept, Fac Pharm, Bani Suwayf, Egypt
[2] Beni Suef Univ, Resp Dept, Fac Med, Bani Suwayf, Egypt
[3] Beni Suef Univ, Clin Pharmacol Dept, Fac Med, Bani Suwayf, Egypt
[4] Modern Univ Technol & Informat, Clin Pharm Dept, Fac Pharm, Cairo, Egypt
关键词
nebulizers; large spacer; HFNC; inhalation therapy; nasal cannula; low-flow oxygen; VIBRATING MESH NEBULIZER; METERED-DOSE INHALER; NONINVASIVE VENTILATION; IN-VITRO; RESPIRATORY-FAILURE; JET NEBULIZER; LUNG DELIVERY; THERAPY; PERFORMANCE; SALBUTAMOL;
D O I
10.4187/respcare.06345
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: There has been a growing trend toward delivering aerosolized medications using high-flow nasal cannula (HFNC). In some cases, patients who do not require high-flow oxygen to maintain adequate oxygenation may benefit from aerosol delivery while receiving low-flow oxygen via HFNC. The objective of this study was to quantify and compare the relative pulmonary and systemic delivery of salbutamol, with 2 different nebulizers, in patients with COPD receiving low-flow oxygen therapy through an HFNC. METHODS: Subjects were randomized to receive study doses of 5 mg salbutamol nebulized by either a jet nebulizer or a vibrating mesh nebulizer with a T-piece or spacer on days 1, 3, and 5 of admission. Subjects using the large spacer also received 2 puffs (100 mu g each) of salbutamol via a pressurized metered-dose-inhaler prior to the nebulizer dose. Urinary salbutamol excretion 30 min post-inhalation and pooled samples of urinary salbutamol excretion up to 24 h post-inhalation were measured. On day 2, ex vivo studies were performed with salbutamol collected on filters placed between the HFNC and nebulizer, with drug eluted from filters and analyzed to determine inhaled dose. RESULTS: Twelve subjects (6 females), age 51.3 +/- 11.2 y, were included. The vibrating mesh nebulizer demonstrated higher urinary salbutamol excretion at 30 min and 24 h post-inhalation compared to a jet nebulizer (P =.001 and P =.02, respectively). No significant difference was found between the T-piece and large-spacer configurations, even though the spacer provided a significantly larger emitted aerosol dose at the opening of the HFNC (P =.002). CONCLUSIONS: Aerosolized medication could be efficiently combined with low-flow oxygen, via HFNC, in COPD subjects without the need to interrupt the gas supply. The vibrating mesh nebulizer delivered larger doses to subjects compared to the jet nebulizer. However, there was no benefit of using the large spacer with HFNC in low-flow delivery, because the small inner diameter of the HFNC does not allow larger aerosol droplet sizes (preserved by the spacer) to reach the subject.
引用
收藏
页码:453 / 461
页数:9
相关论文
共 45 条
[1]   The relative lung and systemic bioavailability of terbutaline following nebulisation in non-invasively ventilated patients [J].
Abdelrahim, M. E. ;
Plant, P. K. ;
Chrystyn, H. .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2011, 420 (02) :313-318
[2]   In-vitro characterisation of the nebulised dose during non-invasive ventilation [J].
Abdelrahim, Mohamed E. ;
Plant, Paul ;
Chrystyn, Henry .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 2010, 62 (08) :966-972
[3]  
[Anonymous], 2017, ARCH PULMONOLOGY RES, DOI DOI 10.17352/APRC.000017
[4]  
[Anonymous], [No title captured]
[5]   Effect of high flow oxygen on mortality in chronic obstructive pulmonary disease patients in prehospital setting: randomised controlled trial [J].
Austin, Michael A. ;
Wills, Karen E. ;
Blizzard, Leigh ;
Walters, Eugene H. ;
Wood-Baker, Richard .
BMJ-BRITISH MEDICAL JOURNAL, 2010, 341 :927
[6]   Pediatric Aerosol Therapy [J].
Berlinski, Ariel .
RESPIRATORY CARE, 2017, 62 (06) :662-677
[7]   In Vitro Performance of Spacers for Aerosol Delivery during Adult Mechanical Ventilation [J].
Boukhettala, Nabile ;
Poree, Thierry ;
Diot, Patrice ;
Vecellio, Laurent .
JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2015, 28 (02) :130-136
[8]   Nasal highflow improves ventilation in patients with COPD [J].
Braeunlich, Jens ;
Koehler, Marcus ;
Wirtz, Hubert .
INTERNATIONAL JOURNAL OF CHRONIC OBSTRUCTIVE PULMONARY DISEASE, 2016, 11 :1077-1085
[9]   Effects of Nasal High Flow on Ventilation in Volunteers, COPD and Idiopathic Pulmonary Fibrosis Patients [J].
Braeunlich, Jens ;
Beyer, Denise ;
Mai, David ;
Hammerschmidt, Stefan ;
Seyfarth, Hans-Juergen ;
Wirtz, Hubert .
RESPIRATION, 2013, 85 (04) :319-325
[10]   The effects of high-flow vs low-flow oxygen on exercise in advanced obstructive airways disease [J].
Chatila, W ;
Nugent, T ;
Vance, G ;
Gaughan, J ;
Criner, GJ .
CHEST, 2004, 126 (04) :1108-1115