Frame-based radiosurgery: Is it relevant in the era of IGRT?

被引:11
作者
Kataria, Tejinder [1 ]
Gupta, Deepak [1 ]
Karrthick, K. P. [1 ]
Bisht, Shyam Singh [1 ]
Goyal, Shikha [1 ]
Abhishek, Ashu [1 ]
Govardhan, H. B. [1 ]
Sharma, Kuldeep [1 ]
Pareek, Puneet [1 ]
Gupta, Aditya [2 ]
机构
[1] Medanta Canc Inst, Div Radiat Oncol, Gurgaon 122001, Haryana, India
[2] Medanta Inst Neurosci, Div Neurosurg, Gurgaon, Haryana, India
关键词
Frame-based radiosurgery; frameless radiosurgery; hexapod couch; radiosurgery; STEREOTACTIC RADIOSURGERY; QUALITY-ASSURANCE; PATIENT FIXATION; BRAIN METASTASES; SETUP ACCURACY; SYSTEM; HEAD;
D O I
10.4103/0028-3886.115068
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Purpose: To assess the setup errors and intrafraction motion in patients treated with frame-based and frameless stereotactic radiosurgery (SRS). Materials and Methods: Ten patients treated with frame-based and six patients treated with frameless radiosurgery were prospectively enrolled in the study. Leksell frame was used for frame-based and a customized uniframe orfit cast for frameless techniques. Cone beam computed tomography (CBCT) scans were taken immediately before and after each treatment to evaluate the positional accuracy and corrections applied with the use of hexapod couch for both groups. Results: The mean translational shifts with frame-based SRS were 1.00 +/- 0.30 mm in the lateral direction (X), 0.20 +/- 1.20 mm in craniocaudal direction (Y) and -0.10 +/- 0.31 mm in the anteroposterior direction (Z). The rotational shifts for frame-based treatments were as follows: roll 0.32 +/- 0.70, pitch 0.44 +/- 0.66 and yaw 0.20 +/- 0.4. For frameless SRS, translational shifts were -0.40 +/- 0.90, 1.10 +/- 1.10, and 0.50 +/- 1.30 mm in X, Y, and Z directions, respectively, and rotational shifts were -0.11 +/- 0.78, 0.20 +/- 0.44, and 0.29 +/- 0.35 in roll, pitch, and yaw, respectively. Intrafraction shifts with frame-based SRS were: X = 0.60 +/- 1.80 mm, Y = 0.20 +/- 0.60 mm, and Z = 0.00 +/- 0.05 mm; and rotational shifts were: roll 0.01 +/- 0.27, pitch 0.06 +/- 0.15, and yaw 0.01 +/- 0.09. For frameless SRS, these were: X = 0.11 +/- 0.20 mm, Y = 0.20 +/- 0.40 mm, and Z = 0.20 +/- 0.20 mm and rotational shifts were: roll 0.09 +/- 0.23, pitch 0.00 +/- 0.12, and yaw 0.00 +/- 0.09. Conclusions: In our experience, set up accuracy of frameless SRS is as good as frame-based SRS. With availability of verification methods such as CBCT and hexapod couch, an accurate and precise treatment delivery is feasible with frameless techniques.
引用
收藏
页码:277 / 281
页数:5
相关论文
共 20 条
[1]   Whole brain radiation therapy with or without stereotactic radiosurgery boost for patients with one to three brain metastases: phase III results of the RTOG 9508 randomised trial [J].
Andrews, DW ;
Scott, CB ;
Sperduto, PW ;
Flanders, AE ;
Gaspar, LE ;
Schell, MC ;
Werner-Wasik, M ;
Demas, W ;
Ryu, J ;
Bahary, JP ;
Souhami, L ;
Rotman, M ;
Mehta, MP ;
Curran, WJ .
LANCET, 2004, 363 (9422) :1665-1672
[2]   FRAMELESS IMAGE-GUIDED INTRACRANIAL STEREOTACTIC RADIOSURGERY: CLINICAL OUTCOMES FOR BRAIN METASTASES [J].
Breneman, John C. ;
Steinmetz, Ryan ;
Smith, Aaron ;
Lamba, Michael ;
Warnick, Ronald E. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03) :702-706
[3]   Preliminary experience with frameless stereotactic radiotherapy [J].
Buatti, JM ;
Bova, FJ ;
Friedman, WA ;
Meeks, SL ;
Marcus, RB ;
Mickle, JP ;
Ellis, TL ;
Mendenhall, WM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :591-599
[4]   Repositioning accuracy of a commercially available thermoplastic mask system [J].
Fuss, M ;
Salter, BJ ;
Cheek, D ;
Sadeghi, A ;
Hevezi, JM ;
Herman, TS .
RADIOTHERAPY AND ONCOLOGY, 2004, 71 (03) :339-345
[5]   Is mask-based stereotactic head-and-neck fixation as precise as stereotactic head fixation for precision radiotherapy? [J].
Georg, Dietmar ;
Bogner, Joachim ;
Dieckmann, Karin ;
Potter, Richard .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (04) :S61-S66
[6]   SETUP ACCURACY OF THE NOVALIS EXACTRAC 6DOF SYSTEM FOR FRAMELESS RADIOSURGERY [J].
Gevaert, Thierry ;
Verellen, Dirk ;
Tournel, Koen ;
Linthout, Nadine ;
Bral, Samuel ;
Engels, Benedikt ;
Collen, Christine ;
Depuydt, Tom ;
Duchateau, Michael ;
Reynders, Truus ;
Storme, Guy ;
De Ridder, Mark .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05) :1627-1635
[7]   Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors [J].
Gilbeau, L ;
Octave-Prignot, M ;
Loncol, T ;
Renard, L ;
Scalliet, P ;
Grégoire, V .
RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) :155-162
[8]  
HODAPP N, 1994, ACT NEUR S, V62, P101
[9]   A FRAMELESS METHOD FOR STEREOTAXIC RADIOTHERAPY [J].
JONES, D ;
CHRISTOPHERSON, DA ;
WASHINGTON, JT ;
HAFERMANN, MD ;
RIEKE, JW ;
TRAVAGLINI, JJ ;
VERMEULEN, SS .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (792) :1142-1150
[10]   Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system [J].
Karger, CP ;
Jäkel, O ;
Debus, J ;
Kuhn, S ;
Hartmann, GH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (05) :1493-1504