'Severe' wear challenge to 'as-cast' and 'double heat-treated' large-diameter metal-on-metal hip bearings

被引:41
作者
Bowsher, JG
Nevelos, J
Williams, PA
Shelton, JC
机构
[1] Univ London Queen Mary Coll, Dept Engn, London E1 4NS, England
[2] Univ London Queen Mary Coll, IRC Biomed Mat, London E1 4NS, England
[3] Loma Linda Univ, Med Ctr, Dept Orthopaed, Loma Linda, CA USA
关键词
metal-on-metal bearings; total hip replacement; heat treatment; metallurgy; lubrication; hip joint simulator; jogging;
D O I
10.1243/09544119JEIM81
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The wear generation of double-heat-treated and as-cast large-diameter metal-on-metal (MOM) hip bearings was investigated using standard- and 'severe'-gait simulations. The test hypothesis was that double heat treatment would change MOM hip wear compared with the as-cast condition. Two groups of high-carbon MOM bearings of 40 mm diameter were manufactured and subjected to either hot isostatic pressing (HIP) and solution annealing (SA) or no heat treatment (as cast). The results showed no statistical difference between the two groups under both running-in and steady state conditions. Even under the Most 'severe'-gait simulation published to date, the mean volumetric wear rates were 2.9 and 3.9 mm(3) per 10(6) cycles for the HIP-SA and as-cast bearings respectively, showing a ten-fold increase in wear compared with walking. These differences were not statistically different; therefore our hypothesis was negated. Changes in alloy microstructure do not appear to influence the wear behaviour of high-carbon cast MOM articulations with similar chemical compositions. This is in sharp contrast with the published significance of bearing diameter and radial clearance on the wear of MOM hip bearings.
引用
收藏
页码:135 / 143
页数:9
相关论文
共 27 条
[1]   The wear pattern in metal-on-metal hip prostheses [J].
Anissian, HL ;
Stark, A ;
Good, V ;
Dahlstrand, H ;
Clarke, IC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (06) :673-678
[2]  
[Anonymous], 1966, P I MECH ENG C P
[3]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[4]   Hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions [J].
Bowsher, JG ;
Shelton, JC .
WEAR, 2001, 250 :167-179
[5]  
BOWSHER JG, 2004, T 50 ORTHOP RES SOC, V29, P1453
[6]  
BUTTERFIELD M, 2002, T 48 ORTH RES SOC M, V27, P128
[7]  
Chan FW, 1996, CLIN ORTHOP RELAT R, P96
[8]  
CHAN FW, 1999, T ORTHOP RES SOC, V24, P310
[9]   Charnley wear model for validation of hip simulators - Ball diameter versus polytetrafluoroethylene and polyethylene wear [J].
Clarke, IC ;
Good, V ;
Anissian, L ;
Gustafson, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :25-36
[10]   A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design [J].
Dowson, D ;
Hardaker, C ;
Flett, M ;
Isaac, GH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :124-130