Evaluation of bone plate with low-stiffness material in terms of stress distribution

被引:105
作者
Benli, Semih [1 ]
Aksoy, Sami [1 ]
Havitcioglu, Hasan [2 ]
Kucuk, Mumin [3 ]
机构
[1] Dokuz Eylul Univ, Fac Engn, TR-35100 Izmir, Turkey
[2] Dokuz Eylul Univ, Fac Med, TR-35340 Izmir, Turkey
[3] Ege Univ, Vocat Sch, Izmir, Turkey
关键词
Bone-plate system; Stress analysis; Low-stiffness material;
D O I
10.1016/j.jbiomech.2008.08.003
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The aim of this study is to evaluate a newly developed bone plate with low-stiffness material in terms of stress distribution. In this numerical study, 3D finite element models of the bone plate with low-stiffness material and traditional bone plates made of stainless steel and Ti alloy have been developed by using the ANSYS software. Stress analyses have been carried out for all three models under the same loading and boundary conditions. Compressive stresses occurring in the intact portion of the bone (tibia) and at the fractured interface at different stages of bone healing have been investigated for all three types of bone-plate systems. The results obtained have been compared and presented in graphs. It has been seen that the bone plate with low-stiffness material offers less stress-shielding to the bone, providing a higher compressive stress at the fractured interface to induce accelerated healing in comparison with Ti alloy and stainless-steel bone plate. In addition, the effects of low-stiffness materials with different Young's modulus on stress distribution at the fractured interface have been investigated in the newly developed bone-plate system. The results showed that when a certain value of Young's modulus of low-stiffness material is exceeded, increase in stiffness of the bone plate does not occur to a large extent and stress distributions and micro-motions at the fractured interface do not change considerably. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:3229 / 3235
页数:7
相关论文
共 12 条
[1]   CARBON-FIBER COMPOSITE BONE PLATES - DEVELOPMENT, EVALUATION AND EARLY CLINICAL-EXPERIENCE [J].
ALI, MS ;
FRENCH, TA ;
HASTINGS, GW ;
RAE, T ;
RUSHTON, N ;
ROSS, ERS ;
WYNNJONES, CH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (04) :586-591
[2]   EFFECT OF ANNEALING TEMPERATURE ON THE DEGRADATION OF REINFORCING FIBERS OR ABSORBABLE IMPLANTS [J].
CHOUEKA, A ;
CHARVET, JL ;
ALEXANDER, H ;
OH, YH ;
JOSEPH, G ;
BLUMENTHAL, NC ;
LACOURSE, WC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (11) :1309-1315
[3]   ARAMID-EPOXY COMPOSITE INTERNAL-FIXATION PLATES - A PILOT-STUDY [J].
COCHRAN, GVB ;
PALMIERI, VR ;
ZICKEL, RE .
CLINICAL BIOMECHANICS, 1994, 9 (05) :315-322
[4]   Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws and the friction between plate and bone [J].
Cordey, J ;
Borgeaud, M ;
Perren, SM .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 :21-28
[5]   Quantification of fracture healing with three-dimensional computed tomography [J].
den Boer, FC ;
Bramer, JAM ;
Patka, P ;
Bakker, FC ;
Barentsen, RH ;
Feilzer, AJ ;
de Lange, ESM ;
Haarman, HJTM .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1998, 117 (6-7) :345-350
[6]   Feasibility of knitted carbon/PEEK composites for orthopedic bone plates [J].
Fujihara, K ;
Huang, ZM ;
Ramakrishna, S ;
Satknanantham, K ;
Hamada, H .
BIOMATERIALS, 2004, 25 (17) :3877-3885
[7]  
GANESH VK, 2005, BIOMEDICAL ENG
[8]  
Pohler OEM, 2000, INJURY, V31, pS7
[9]   PROTECTION FROM STRESS IN BONE AND ITS EFFECTS - EXPERIMENTS WITH STAINLESS-STEEL AND PLASTIC PLATES IN DOGS [J].
TONINO, AJ ;
DAVIDSON, CL ;
KLOPPER, PJ ;
LINCLAU, LA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1976, 58 (01) :107-113
[10]   ULTRA-HIGH-STRENGTH ABSORBABLE SELF-REINFORCED POLYGLYCOLIDE (SR-PGA) COMPOSITE RODS FOR INTERNAL-FIXATION OF BONE-FRACTURES - INVITRO AND INVIVO STUDY [J].
TORMALA, P ;
VASENIUS, J ;
VAINIONPAA, S ;
LAIHO, J ;
POHJONEN, T ;
ROKKANEN, P .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1991, 25 (01) :1-22