Distal Femur Locking Plates Fit Poorly Before and After Total Knee Arthroplasty

被引:14
作者
Campbell, Sean T. [1 ]
Bosch, Liam C. [1 ]
Swinford, Steven [1 ]
Amanatullah, Derek F. [1 ]
Bishop, Julius A. [1 ]
Gardner, Michael J. [1 ]
机构
[1] Stanford Univ Hosp, Dept Orthopaed Surg, Stanford, CA 94305 USA
关键词
distal femur fracture; periprosthetic distal femur fracture; distal femur locking plate; translational deformity; total knee arthroplasty; FRACTURES; FIXATION;
D O I
10.1097/BOT.0000000000001431
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To evaluate the fit of distal femur locking plates. Secondarily, we sought to compare plate fit among patients with and without a total knee arthroplasty (TKA). Design: Retrospective. Setting: University hospital. Intervention: Standard length precontoured distal femur locking plates from 4 manufacturers were digitally templated onto each patient's pre-TKA and post-TKA radiographs. Main Outcome Measurements: The maximum distance from the plate to the lateral femoral cortex (plate-bone distance) was measured in the metaphyseal region. Mean plate-bone distances were compared between manufacturers and between pre-TKA and postTKA radiographs. Results: All implants tested were undercontoured in all patients. Plate-bone distances ranged from 6.6 0.4 mm to 8.0 0.4 mm (mean SE) pre-TKA and 8.2 0.3 mm to 8.6 0.3 mm after TKA, indicating worse fit after arthroplasty (P < 0.001). There were also intermanufacturer differences, with Synthes and Smith & Nephew implants demonstrating the lowest plate-bone distances in the pre-TKA and post-TKA groups, respectively. Proportionally, plate-bone increase was greater in the female cohort (16%) compared with the male cohort (8%). Conclusions: There was plate-bone mismatch for the distal femur locking plates tested in this study, due to undercontouring of the implants. After patients underwent TKA, poor implant fit was exacerbated. Surgeons must be aware of the potential for deformity if the proximal segment is brought into contact with the implant. These finding may help optimize implant design for the treatment of periprosthetic distal femur fractures.
引用
收藏
页码:239 / 243
页数:5
相关论文
共 12 条
[1]   Periprosthetic supracondylar femur fractures following total knee arthroplasty [J].
Bezwada, HP ;
Neubauer, P ;
Baker, J ;
Israelite, CL ;
Johanson, NA .
JOURNAL OF ARTHROPLASTY, 2004, 19 (04) :453-458
[2]   Radiographic and Clinical Outcomes of Periprosthetic Distal Femur Fractures Treated With Open Reduction Internal Fixation [J].
Christ, Alexander B. ;
Chawla, Harshvardhan ;
Gausden, Elizabeth B. ;
Villa, Jordan C. ;
Wellman, David S. ;
Lorich, Dean G. ;
Helfet, David L. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (10) :515-520
[3]   High Complication Rate in Locking Plate Fixation of Lower Periprosthetic Distal Femur Fractures in Patients With Total Knee Arthroplasties [J].
Ebraheim, Nabil A. ;
Liu, Jiayong ;
Hashmi, Sohaib Z. ;
Sochacki, Kyle R. ;
Moral, Muhammad Z. ;
Hirschfeld, Adam G. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (05) :809-813
[4]   Working length and proximal screw constructs in plate osteosynthesis of distal femur fractures [J].
Harvin, William H. ;
Oladeji, Lasun O. ;
Della Rocca, Gregory J. ;
Murtha, Yvonne M. ;
Volgas, David A. ;
Stannard, James P. ;
Crist, Brett D. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2017, 48 (11) :2597-2601
[5]   Locking Plates for Distal Femur Fractures: Is There a Problem With Fracture Healing? [J].
Henderson, Christopher E. ;
Kuhl, Lori L. ;
Fitzpatrick, Daniel C. ;
Marsh, J. L. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 :S8-S14
[6]   Outcome of periprosthetic distal femoral fractures following knee arthroplasty [J].
Hoffmann, M. F. ;
Jones, C. B. ;
Sietsema, D. L. ;
Koenig, S. J. ;
Tornetta, P., III .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2012, 43 (07) :1084-1089
[7]   Implant Material, Type of Fixation at the Shaft, and Position of Plate Modify Biomechanics of Distal Femur Plate Osteosynthesis [J].
Kandemir, Utku ;
Augat, Peter ;
Konowalczyk, Stefanie ;
Wipf, Felix ;
von Oldenburg, Geert ;
Schmidt, Ulf .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (08) :E241-E246
[8]   Patient Mortality in Geriatric Distal Femur Fractures [J].
Myers, Philip ;
Laboe, Patrick ;
Johnson, Kory J. ;
Fredericks, Peter D. ;
Crichlow, Renn J. ;
Maar, Dean C. ;
Weber, Timothy G. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (03) :111-115
[9]   Cadaveric Analysis of Proximal Humerus Locking Plate Fit: Contour Mismatch May Lead to Malreduction [J].
Ravindra, Amy ;
Roebke, Austin ;
Goyal, Kanu S. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2017, 31 (12) :663-667
[10]   Predictors of Functional Recovery Following Periprosthetic Distal Femur Fractures [J].
Ruder, John A. ;
Hart, Gavin P. ;
Kneisl, Jeffrey S. ;
Springer, Bryan D. ;
Karunakar, Madhav A. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (05) :1571-1575