Reproducibility of Tumor Perfusion Measurements Using 15O-Labeled Water and PET

被引:46
作者
de Langen, Adrianus J. [2 ]
Lubberink, Mark [1 ]
Boellaard, Ronald [1 ]
Spreeuwenberg, Marieke D. [3 ]
Smit, Egbert F. [2 ]
Hoekstra, Otto S. [1 ]
Lammertsma, Adriaan A. [1 ]
机构
[1] Vrije Univ Amsterdam Med Ctr, Dept Nucl Med & PET Res, NL-1007 MB Amsterdam, Netherlands
[2] Vrije Univ Amsterdam Med Ctr, Dept Resp Med, NL-1007 MB Amsterdam, Netherlands
[3] Vrije Univ Amsterdam Med Ctr, Dept Clin Epidemiol & Biostat, NL-1007 MB Amsterdam, Netherlands
关键词
positron emission tomography; blood flow; perfusion; non-small-cell lung cancer;
D O I
10.2967/jnumed.108.053454
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PET and O-15-labeled water ((H2O)-O-15) can be used to noninvasively monitor tumor perfusion. This allows evaluation of the direct target of antiangiogenic drugs, that is, tumor vasculature. Because these drugs often result in consolidation rather than regression of the tumor mass, a change in perfusion might be a more sensitive way to evaluate response than are indirect size measures on a CT scan. However, to use the technique for serial imaging of individual patients, good reproducibility is essential. The purpose of the present study was to evaluate the reproducibility of quantitative (H2O)-O-15 measurements. Methods: Nine patients with non-small-cell lung cancer (NSCLC) were scanned twice within 7 d and before any therapy. All (H2O)-O-15 scans were followed by an F-18-fluorothymidine scan to allow for adequate volume-of-interest (VOI) definition. VOIs were defined using a 3-dimensional threshold technique. Tumor perfusion and the volume of distribution (V-T) were obtained using a 1-tissue-compartment model including an arterial blood volume component and an image-derived input function. The level of agreement between test and retest values was assessed using the intraclass correlation coefficient (ICC) and Bland-Altman analyses. Possible dependency on absolute values and lesion size was assessed by linear regression. Results: All primary tumors and more than 90% of clinically suspected locoregional metastases could be delineated. In total, 14 lesions in 9 patients were analyzed. Tumor perfusion showed excellent reproducibility, with an ICC of 0.95 and SD of 9%. The V-T was only moderately reproducible, with an ICC of 0.52 and SD of 16%. No dependency was found on absolute values of perfusion (P = 0.14) and V-T (P = 0.15). In addition, tumor volume did not influence the reproducibility of perfusion (P = 0.46) and V-T (P = 0.25). Conclusion: Quantitative measurements of tumor perfusion using (H2O)-O-15 and PET are reproducible in NSCLC. When patients are repeatedly being scanned during therapy, changes of more than 18% in tumor perfusion and 32% in V-T (>1.96 x SD) are likely to represent treatment effects.
引用
收藏
页码:1763 / 1768
页数:6
相关论文
共 50 条
[31]   Cerebrovascular reactivity measurements using simultaneous 15O-water PET and ASL MRI: Impacts of arterial transit time, labeling efficiency, and hematocrit [J].
Zhao, Moss Y. ;
Fan, Audrey P. ;
Chen, David Yen-Ting ;
Sokolska, Magdalena J. ;
Guo, Jia ;
Ishii, Yosuke ;
Shin, David D. ;
Khalighi, Mohammad Mehdi ;
Holley, Dawn ;
Halbert, Kim ;
Otte, Andrea ;
Williams, Brittney ;
Rostami, Taghi ;
Park, Jun-Hyung ;
Shen, Bin ;
Zaharchuk, Greg .
NEUROIMAGE, 2021, 233
[32]   Clinical Value of Absolute Quantification of Myocardial Perfusion With 15O-Water in Coronary Artery Disease [J].
Kajander, Sami A. ;
Joutsiniemi, Esa ;
Saraste, Markku ;
Pietila, Mikko ;
Ukkonen, Heikki ;
Saraste, Antti ;
Sipila, Hannu T. ;
Teras, Mika ;
Maki, Maija ;
Airaksinen, Juhani ;
Hartiala, Jaakko ;
Knuuti, Juhani .
CIRCULATION-CARDIOVASCULAR IMAGING, 2011, 4 (06) :678-684
[33]   Preliminary protocol for measuring the reproducibility and accuracy of flow values on digital PET/CT systems in [15O]H2O myocardial perfusion imaging using a flow phantom [J].
Siekkinen, Reetta ;
Partanen, Heidi ;
Kukola, Linda ;
Tolvanen, Tuula ;
Fenwick, Andrew ;
Smith, Nadia A. S. ;
Teras, Mika ;
Saraste, Antti ;
Teuho, Jarmo .
EJNMMI PHYSICS, 2024, 11 (01)
[34]   Whole-body parametric mapping of tumour perfusion in metastatic prostate cancer using long axial field-of-view [15O]H2O PET [J].
Jochumsen, Mads Ryo ;
Christensen, Nana L. ;
Iversen, Peter ;
Gormsen, Lars C. ;
Sorensen, Jens ;
Tolbod, Lars P. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 51 (13) :4134-4140
[35]   Assessment of myocardial viability with [15O]water PET: A validation study in experimental myocardial infarction [J].
Gronman, Maria ;
Tarkia, Miikka ;
Stark, Christoffer ;
Vahasilta, Tommi ;
Kiviniemi, Tuomas ;
Lubberink, Mark ;
Halonen, Paavo ;
Kuivanen, Antti ;
Saunavaara, Virva ;
Tolvanen, Tuula ;
Teuho, Jarmo ;
Teras, Mika ;
Savunen, Timo ;
Pietila, Mikko ;
Yla-Herttuala, Seppo ;
Roivainen, Anne ;
Knuuti, Juhani ;
Saraste, Antti .
JOURNAL OF NUCLEAR CARDIOLOGY, 2021, 28 (04) :1271-1280
[36]   Comparison of simultaneous arterial spin labeling MRI and 15O-H2O PET measurements of regional cerebral blood flow in rest and altered perfusion states [J].
Puig, Oriol ;
Henriksen, Otto M. ;
Vestergaard, Mark B. ;
Hansen, Adam E. ;
Andersen, Flemming L. ;
Ladefoged, Claes N. ;
Rostrup, Egill ;
Larsson, Henrik B. W. ;
Lindberg, Ulrich ;
Law, Ian .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2020, 40 (08) :1621-1633
[37]   Pharmacodynamic analysis of tumour perfusion assessed by 15O-water-PET imaging during treatment with sunitinib malate in patients with advanced malignancies [J].
Scott, Andrew M. ;
Mitchell, Paul L. ;
O'Keefe, Graeme ;
Saunder, Timothy ;
Hicks, Rodney J. ;
Poon, Aurora ;
Baum, Charles ;
Brega, Nicoletta ;
McCarthy, Timothy J. ;
Toner, Guy C. .
EJNMMI RESEARCH, 2012, 2
[38]   Reproducibility of FDG PET based metabolic tumor volume measurements and of their FDG distribution within [J].
Kruse, V. ;
Mees, G. ;
Maes, A. ;
D'Asseler, Y. ;
Borms, M. ;
Cocquyt, V. ;
Van De Wiele, C. .
QUARTERLY JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 59 (04) :462-468
[39]   Assessment of endothelial function and myocardial flow reserve using 15O-water PET without attenuation correction [J].
Tuffier, Stephane ;
Legallois, Damien ;
Belin, Annette ;
Joubert, Michael ;
Bailliez, Alban ;
Redonnet, Michel ;
Agostini, Denis ;
Manrique, Alain .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 (02) :288-295
[40]   Quantitative analysis of coronary endothelial function with generator-produced 82Rb PET: comparison with 15O-labelled water PET [J].
Yoshinaga, Keiichiro ;
Manabe, Osamu ;
Katoh, Chietsugu ;
Chen, Li ;
Klein, Ran ;
Naya, Masanao ;
deKemp, Robert A. ;
Williams, Kathryn ;
Beanlands, Rob S. B. ;
Tamaki, Nagara .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (12) :2233-2241