A field-deployable mobile molecular diagnostic system for malaria at the point of need

被引:41
作者
Choi, Gihoon [1 ]
Song, Daniel [2 ]
Shrestha, Sony [3 ]
Miao, Jun [3 ]
Cui, Liwang [3 ]
Guan, Weihua [1 ,2 ]
机构
[1] Penn State Univ, Dept Elect Engn, University Pk, PA 16802 USA
[2] Penn State Univ, Dept Biomed Engn, University Pk, PA 16802 USA
[3] Penn State Univ, Dept Entomol, University Pk, PA 16802 USA
基金
美国国家卫生研究院;
关键词
MEDIATED ISOTHERMAL AMPLIFICATION; PLASMODIUM-FALCIPARUM; ELIMINATION; DEVICES; DNA; LAB; SENSITIVITY; VIVAX; LAMP; KIT;
D O I
10.1039/c6lc01078d
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In response to the urgent need of a field-deployable and highly sensitive malaria diagnosis, we developed a standalone, "sample-in-answer-out" molecular diagnostic system (AnyMDx) to enable quantitative molecular analysis of blood-borne malaria in low resource areas. The system consists of a durable battery-powered analyzer and a disposable microfluidic compact disc loaded with reagents ready for use. A low power thermal module and a novel fluorescence-sensing module are integrated into the analyzer for real-time monitoring of loop-mediated isothermal nucleic acid amplification ( LAMP) of target parasite DNA. With 10 mu L of raw blood sample, the AnyMDx system automates the nucleic acid sample preparation and subsequent LAMP and real-time detection. Under laboratory conditions with whole-blood samples spiked with cultured Plasmodium falciparum, we achieved a detection limit of similar to 0.6 parasite per mu L, much lower than those for the conventional microscopy and rapid diagnostic tests (similar to 50-100 parasites per mu L). The turnaround time from sample to answer is less than 40 minutes. The AnyMDx is user-friendly requiring minimal technological training. The analyzer and the disposable reagent compact discs are cost-effective, making AnyMDx a potential tool for malaria molecular diagnosis under field settings for malaria elimination.
引用
收藏
页码:4341 / 4349
页数:9
相关论文
共 47 条
[1]   Current status and future prospects of point-of-care testing around the globe [J].
Abel, Gyorgy .
EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2015, 15 (07) :853-855
[2]  
Aly HAA, 2015, WHO TECH REP SER, V994, P1
[3]   Loop Mediated Isothermal Amplification (LAMP) Accurately Detects Malaria DNA from Filter Paper Blood Samples of Low Density Parasitaemias [J].
Aydin-Schmidt, Berit ;
Xu, Weiping ;
Gonzalez, Iveth J. ;
Polley, Spencer D. ;
Bell, David ;
Shakely, Deler ;
Msellem, Mwinyi I. ;
Bjorkman, Anders ;
Martensson, Andreas .
PLOS ONE, 2014, 9 (08)
[4]   The effect of malaria control on Plasmodium falciparum in Africa between 2000 and 2015 [J].
Bhatt, S. ;
Weiss, D. J. ;
Cameron, E. ;
Bisanzio, D. ;
Mappin, B. ;
Dalrymple, U. ;
Battle, K. E. ;
Moyes, C. L. ;
Henry, A. ;
Eckhoff, P. A. ;
Wenger, E. A. ;
Briet, O. ;
Penny, M. A. ;
Smith, T. A. ;
Bennett, A. ;
Yukich, J. ;
Eisele, T. P. ;
Griffin, J. T. ;
Fergus, C. A. ;
Lynch, M. ;
Lindgren, F. ;
Cohen, J. M. ;
Murray, C. L. J. ;
Smith, D. L. ;
Hay, S. I. ;
Cibulskis, R. E. ;
Gething, P. W. .
NATURE, 2015, 526 (7572) :207-+
[5]   Rotationally controlled magneto-hydrodynamic particle handling for bead-based microfluidic assays [J].
Burger, Robert ;
Reith, Patrick ;
Akujobi, Victor ;
Ducree, Jens .
MICROFLUIDICS AND NANOFLUIDICS, 2012, 13 (04) :675-681
[6]   Powering point-of-care diagnostic devices [J].
Choi, Seokheun .
BIOTECHNOLOGY ADVANCES, 2016, 34 (03) :321-330
[7]   Malaria Management: Past, Present, and Future [J].
Enayati, A. ;
Hemingway, J. .
ANNUAL REVIEW OF ENTOMOLOGY, 2010, 55 :569-591
[8]   Microfluidic designs and techniques using lab-on-a-chip devices for pathogen detection for point-of-care diagnostics [J].
Foudeh, Amir M. ;
Didar, Tohid Fatanat ;
Veres, Teodor ;
Tabrizian, Maryam .
LAB ON A CHIP, 2012, 12 (18) :3249-3266
[9]   Contact angle hysteresis explained [J].
Gao, Lichao ;
McCarthy, Thomas J. .
LANGMUIR, 2006, 22 (14) :6234-6237
[10]   Centrifugal microfluidics for biomedical applications [J].
Gorkin, Robert ;
Park, Jiwoon ;
Siegrist, Jonathan ;
Amasia, Mary ;
Lee, Beom Seok ;
Park, Jong-Myeon ;
Kim, Jintae ;
Kim, Hanshin ;
Madou, Marc ;
Cho, Yoon-Kyoung .
LAB ON A CHIP, 2010, 10 (14) :1758-1773