Microfabricated Implantable Parylene-Based Wireless Passive Intraocular Pressure Sensors

被引:194
作者
Chen, Po-Jui [1 ]
Rodger, Damien C. [2 ]
Saati, Saloomeh [2 ,3 ]
Humayun, Mark S. [2 ,3 ]
Tai, Yu-Chong [1 ,4 ]
机构
[1] CALTECH, Div Engn & Appl Sci, Dept Elect Engn, Pasadena, CA 91125 USA
[2] Univ So Calif, Keck Sch Med, Los Angeles, CA 90033 USA
[3] Doheny Eye Inst, Los Angeles, CA 90033 USA
[4] CALTECH, Div Engn & Appl Sci, Dept Bioengn, Pasadena, CA 91125 USA
基金
美国国家科学基金会;
关键词
Implantable microdevice; intraocular pressure (IOP); parylene; pressure sensor; wireless sensing;
D O I
10.1109/JMEMS.2008.2004945
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents an implantable parylene-based wireless pressure sensor for biomedical pressure sensing applications specifically designed for continuous intraocular pressure (IOP) monitoring in glaucoma patients. It has an electrical LC tank resonant circuit formed by an integrated capacitor and an inductor coil to facilitate passive wireless sensing using an external interrogating coil connected to a readout unit. Two surface-micromachined sensor designs incorporating variable capacitor and variable capacitor/inductor resonant circuits have been implemented to realize the pressure-sensitive components. The sensor is monolithically microfabricated by exploiting parylene as a biocompatible structural material in a suitable form factor for minimally invasive intraocular implantation. Pressure responses of the microsensor have been characterized to demonstrate its high pressure sensitivity (> 7000 ppm/mmHg) in both sensor designs, which confirms the feasibility of pressure sensing with smaller than 1 mmHg of resolution for practical biomedical applications. A six-month animal study verifies the in vivo bioefficacy and biostability of the implant in the intraocular environment with no surgical or postoperative complications. Preliminary ex vivo experimental results verify the TOP sensing feasibility of such device. This sensor will ultimately be implanted at the pars plana or on the iris of the eye to fulfill continuous, convenient, direct, and faithful TOP monitoring.
引用
收藏
页码:1342 / 1351
页数:10
相关论文
共 34 条
  • [1] A wireless batch sealed absolute capacitive pressure sensor
    Akar, O
    Akin, T
    Najafi, K
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2001, 95 (01) : 29 - 38
  • [2] Asrani S, 2000, J GLAUCOMA, V9, P134
  • [3] A self-resonant frequency-modulated micromachined passive pressure transensor
    Baldi, A
    Choi, W
    Ziaie, B
    [J]. IEEE SENSORS JOURNAL, 2003, 3 (06) : 728 - 733
  • [4] Boresi A.P., 2003, Advanced Mechanics of Materials, V6th
  • [5] Chen PJ, 2008, PROC IEEE MICR ELECT, P58
  • [7] A readout circuit for an intra-ocular pressure sensor
    Coosemans, J
    Catrysse, M
    Puers, R
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2004, 110 (1-3) : 432 - 438
  • [8] DeHennis A, 2002, PROC IEEE MICR ELECT, P252, DOI 10.1109/MEMSYS.2002.984250
  • [9] Ocular biomechanics and biotransport
    Ethier, CR
    Johnson, M
    Ruberti, J
    [J]. ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2004, 6 : 249 - 273
  • [10] Fonseca M. A., 2006, SOL STAT SENS ACT MI, P37, DOI DOI 10.31438/TRF.HH2006.9