A comparison of the INVOS 4100 and the NIRO 300 near-infrared spectrophotometers

被引:117
作者
Yoshitani, K
Kawaguchi, M
Tatsumi, K
Kitaguchi, K
Furuya, H
机构
[1] Nara Med Univ, Dept Anesthesiol, Nara 6348522, Japan
[2] Seikeikai Hosp, Dept Anesthesia, Osaka, Japan
关键词
D O I
10.1097/00000539-200203000-00020
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We determined whether two different devices for measuring near-infrared spectroscopy (NIRS)-the INVOS 4100 and the NIRO 300-produce similar cerebral oxygenation data during the CO2 challenge test. Nineteen patients anesthetized with sevoflurane, 67% nitrous oxide in oxygen, and fentanyl were studied. A series of measurements of regional cerebral oxygen saturation (rSo(2)), measured by the INVOS 4100, and tissue oxygen index (TOI), measured by the NIRO 300, were performed in the following conditions: 1) normocapnia (Paco(2), 35-45 mm Hg); 2) hypocapnia (Paco(2), 2535 mm Hg); 3) normocapnia; and 4) hypercapnia (Paco(2), 45-55 mm Hg). Hemodynamic variables, including arterial blood gases and cerebral blood flow velocity, were measured at the same time with transcranial Doppler. The values and percentage changes of rSo(2) and TOI were compared by using regression analysis and Bland and Altman analysis. The rSo(2) showed a significant positive correlation with TOI (r = 0.58, P < 0.01). The percentage change of rSo(2) also showed a significant positive correlation with the percentage change of TOI during the CO2 challenge (r = 0.85, P < 0.01). Bland and Altman analysis revealed a bias of -0.5% with 2 SD of 15.6% when comparing the rSo(2) value with the TOI value, and it showed a bias of -3.4% with 2 SD of 15.2% when comparing the percentage change of rSo2 with the percentage change of TOI, indicating unacceptable disagreement of these data. These results indicate that cerebral oxygen saturation and its relative change during the CO2 challenge may vary depending on the type of NIRS used. Because the measurement technique and algorithm were different in each device, we should carefully consider the clinical application of the values produced by NIRS.
引用
收藏
页码:586 / 590
页数:5
相关论文
共 18 条
[1]   Assessment of spatially resolved spectroscopy during cardiopulmonary bypass [J].
Al-Rawi, PG ;
Smielewski, P ;
Hobbiger, H ;
Ghosh, S ;
Kirkpatrick, PJ .
JOURNAL OF BIOMEDICAL OPTICS, 1999, 4 (02) :208-216
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
BRAZY JE, 1985, PEDIATRICS, V75, P217
[4]   Cerebral monitoring by means of oximetry and somatosensory evoked potentials during carotid endarterectomy [J].
Cho, H ;
Nemoto, EM ;
Yonas, H ;
Balzer, J ;
Sclabassi, RJ .
JOURNAL OF NEUROSURGERY, 1998, 89 (04) :533-538
[5]   Comparison of two commercially available near-infrared spectroscopy instruments for cerebral oximetry - Technical note [J].
Cho, H ;
Nemoto, EM ;
Sanders, M ;
Fernandez, K ;
Yonas, H .
JOURNAL OF NEUROSURGERY, 2000, 93 (02) :351-354
[6]  
CORP M, 1998, MED BIOL ENG COMPUT, V26, P289
[7]   MEASUREMENT OF HEMOGLOBIN FLOW AND BLOOD-FLOW BY NEAR-INFRARED SPECTROSCOPY [J].
EDWARDS, AD ;
RICHARDSON, C ;
VANDERZEE, P ;
ELWELL, C ;
WYATT, JS ;
COPE, M ;
DELPY, DT ;
REYNOLDS, EOR .
JOURNAL OF APPLIED PHYSIOLOGY, 1993, 75 (04) :1884-1889
[8]   NEAR-INFRARED SPECTROSCOPY IN ADULTS - EFFECTS OF EXTRACRANIAL ISCHEMIA AND INTRACRANIAL HYPOXIA ON ESTIMATION OF CEREBRAL OXYGENATION [J].
GERMON, TJ ;
KANE, NM ;
MANARA, AR ;
NELSON, RJ .
BRITISH JOURNAL OF ANAESTHESIA, 1994, 73 (04) :503-506
[9]   A comparison of the hamamatsu NIRO 500 and the INVOS 3100 near-infrared spectrophotometers [J].
Gomersall, CD ;
Leung, PL ;
Gin, T ;
Joynt, GM ;
Young, RJ ;
Poon, WS ;
Oh, TE .
ANAESTHESIA AND INTENSIVE CARE, 1998, 26 (05) :548-557
[10]   A comparison of the monitors INVOS3100 and NIRO500 in detecting changes in cerebral oxygenation [J].
Grubhofer, G ;
Tonninger, W ;
Keznicki, P ;
Skyllouriotis, P ;
Ehrlich, M ;
Hiesmayr, M ;
Lassnigg, A .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1999, 43 (04) :470-475