Optical Coherence Tomography after Carotid Stenting: Rate of Stent Malapposition, Plaque Prolapse and Fibrous Cap Rupture According to Stent Design

被引:96
作者
de Donato, G. [1 ]
Setacci, F. [1 ]
Sirignano, P. [1 ]
Galzerano, G. [1 ]
Cappelli, A. [1 ]
Setacci, C. [1 ]
机构
[1] Univ Siena, Dept Surg, Vasc & Endovasc Surg Unit, I-53100 Siena, Italy
关键词
lntravascular imaging; Optical coherence tomography; Carotid stenting; Stent apposition; Plaque prolapse; ELUTING STENTS; BARE-METAL; CELL; IMPLANTATION; FEASIBILITY; EXPERIENCE; COVERAGE; DOCUMENT; SAFETY; RISK;
D O I
10.1016/j.ejvs.2013.03.005
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objectives: This study aims to evaluate the rate of stent malapposition, plaque prolapse and fibrous cap rupture detected by optical coherence tomography (OCT) imaging according to carotid stent design. Design: It was a prospective single-centre study. Materials and methods: Forty consecutive patients undergoing protected carotid artery stenting (CAS) and high-definition OCT image acquisition were enrolled in the study. OCT frames were analysed off-line, in a dedicated core laboratory by two independent physicians. Cross-sectional OCT images within the stented segment of the internal carotid artery were evaluated at 1-mm intervals for the presence of strut malapposition, plaque prolapse and fibrous cap rupture according to stent design. Results: Closed-cell design stents (CC) were used in 17 patients (42.5%), open-cell design stents (OC) in 13 (32.5%) and hybrid design stents (Hyb) in 10 (25%). No procedural or post-procedural neurological complications occurred (stroke/death 0% at 30 days). On OCT analysis the frequencies of malapposed struts were higher with CC compared to OC and Hyb (34.5% vs 15% and 16.3%, respectively; p < 0.01). Plaque prolapse was more frequent with OC vs CC (68.6% vs 23.3%; p < 0.01) and vs Hyb stents (30.8%; p < 0.01). Significant differences were also noted in the rates of fibrous cap rupture between CC and OC (24.2% vs 43.8%; p < 0.01), and between CC and Hyb (24.2% vs 39.6%; p < 0.01), but not between OC and Hyb stents (p = 0.4). Conclusion: Intravascular OCT after CAS revealed that micro-defects after stent deployment are frequent and are related to the design of implanted stents. Stent malapposition is more frequent with CC stents, while plaque prolapse is more common with OC stents. It remains, however, unknown whether these figures now detected with OCT are of any clinical and prognostic significance. (C) 2013 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:579 / 587
页数:9
相关论文
共 25 条
[1]  
Bonati LH, 2010, LANCET NEUROL, V9, P353, DOI 10.1016/S1474-4422(10)70057-0
[2]   Does free cell area influence the outcome in carotid artery stenting? [J].
Bosiers, M. ;
de Donato, G. ;
Deloose, K. ;
Verbist, J. ;
Peeters, P. ;
Castriota, F. ;
Cremonesi, A. ;
Setacci, C. .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2007, 33 (02) :135-141
[3]   Carotid artery stenting - First consensus document of the ICCS-SPREAD joint committee [J].
Cremonesi, Alberto ;
Setacci, Carlo ;
Bignamini, Angelo ;
Bolognese, Leonardo ;
Briganti, Francesco ;
Di Sciascio, Germano ;
Inzitari, Domenico ;
Lanza, Gaetano ;
Lupattelli, Luciano ;
Mangiafico, Salvatore ;
Pratesi, Carlo ;
Reimers, Bernard ;
Ricci, Stefano ;
de Donato, Gianmarco ;
Ugolotti, Ugo ;
Zaninelli, Augusto ;
Gensini, Gian Franco .
STROKE, 2006, 37 (09) :2400-2409
[4]  
Cremonesi Alberto, 2005, EuroIntervention, V1, P289
[5]   Pathological correlates of late drug-eluting stent thrombosis - Strut coverage as a marker of endothelialization [J].
Finn, Aloke V. ;
Joner, Michael ;
Nakazawa, Gaku ;
Kolodgie, Frank ;
Newell, John ;
John, Mike C. ;
Gold, Herman K. ;
Virmani, Renu .
CIRCULATION, 2007, 115 (18) :2435-2441
[6]   Differential response of delayed healing and persistent inflammation at sites of overlapping sirolimus- or paclitaxel-eluting stents [J].
Finn, AV ;
Kolodgie, FD ;
Harnek, J ;
Guerrero, LJ ;
Acampado, E ;
Tefera, K ;
Skorija, K ;
Weber, DK ;
Gold, HK ;
Virmani, R .
CIRCULATION, 2005, 112 (02) :270-278
[7]  
GRAYWEALE AC, 1988, J CARDIOVASC SURG, V29, P676
[8]   Optical Coherence Tomography Assessment of In Vivo Vascular Response After Implantation of Overlapping Bare-Metal and Drug-Eluting Stents [J].
Guagliumi, Giulio ;
Musumeci, Giuseppe ;
Sirbu, Vasile ;
Bezerra, Hiram G. ;
Suzuki, Nobuaki ;
Fiocca, Luigi ;
Matiashvili, Aleksandre ;
Lortkipanidze, Nikoloz ;
Trivisonno, Antonio ;
Valsecchi, Orazio ;
Biondi-Zoccai, Giuseppe ;
Costa, Marco A. .
JACC-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (05) :531-539
[9]   Late stent malapposition risk is higher after drug-eluting stent compared with bare-metal stent implantation and associates with late stent thrombosis [J].
Hassan, Ayman K. M. ;
Bergheanu, Sandrin C. ;
Stijnen, Theo ;
van der Hoeven, Bas L. ;
Snoep, Jaapjan D. ;
Plevier, Josepha W. M. ;
Schalij, Martin J. ;
Jukema, J. Wouter .
EUROPEAN HEART JOURNAL, 2010, 31 (10) :1172-1180
[10]   Intravascular Frequency-Domain Optical Coherence Tomography Assessment of Atherosclerosis and Stent-Vessel Interactions in Human Carotid Arteries [J].
Jones, M. R. ;
Attizzani, G. F. ;
Given, C. A., II ;
Brooks, W. H. ;
Costa, M. A. ;
Bezerra, H. G. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (08) :1494-1501