Calculation of beam quality correction factors for various thimble ionization chambers using the Monte Carlo code PENELOPE

被引:24
作者
Erazo, Fabian [1 ,2 ]
Lallena, Antonio M. [1 ]
机构
[1] Univ Granada, Dept Fis Atom Mol & Nucl, E-18071 Granada, Spain
[2] SOLCA, Inst Canc, Cuenca, Ecuador
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2013年 / 29卷 / 02期
关键词
Ionization chamber perturbation factors; PENELOPE; NE2571; A19; A12S; HIGH-ENERGY PHOTON; REFERENCE DOSIMETRY; K(Q) FACTORS;
D O I
10.1016/j.ejmp.2012.01.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The beam quality correction factor k(Q,Q0) and the perturbation factor p(Q), commonly considered in dosimetry with ionization chambers, were calculated for the NE2571 and the Standard Imaging A19 and Al2S chambers, using the Monte Carlo simulation code PENELOPE. For the NE2571 chamber, the values of k(Q,Q0) obtained are in very good agreement with those found in previous works by Wulff et al. and Muir and Rogers with the code EGSnrc and also with the experimental results summarized in the NCS code of practice. For p(Q), a difference of similar to 0.4% has been found between our results and those obtained with EGSnrc for Co-60 and this difference increases slightly with TPR1020 values. These factors have been calculated also for the A19 and Al2S chambers of Standard Imaging. The values of k(Q,Q0) show reasonable agreement with those recently calculated by Muir and Rogers and the measurements of McEwen. (C) 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:163 / 170
页数:8
相关论文
共 24 条
[1]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]   DESIGN OF A THIMBLE CHAMBER FOR FARMER DOSIMETER [J].
AIRD, EGA ;
FARMER, FT .
PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (02) :169-&
[3]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[4]  
Andreo P, 1994, P S MEAS ASS DOS
[5]   A package of Linux scripts for the parallelization of Monte Carlo simulations [J].
Badal, Andreu ;
Sempau, Josep .
COMPUTER PHYSICS COMMUNICATIONS, 2006, 175 (06) :440-450
[6]   An empirical relationship for determining photon beam quality in TG-21 from a ratio of percent depth doses [J].
Followill, DS ;
Tailor, RC ;
Tello, VM ;
Hanson, WF .
MEDICAL PHYSICS, 1998, 25 (07) :1202-1205
[7]  
International Atomic Energy Agency, 2000, IAEA TECHN REP SER, V398
[8]   Which accelerator photon beams are "clinic-like" for reference dosimetry purposes? [J].
Kalach, NI ;
Rogers, DWO .
MEDICAL PHYSICS, 2003, 30 (07) :1546-1555
[9]  
Kawrakow I., 2007, PIRS701 NRC
[10]   On the effective point of measurement in megavoltage photon beams [J].
Kawrakow, Iwan .
MEDICAL PHYSICS, 2006, 33 (06) :1829-1839