MR Thermometry in the Human Prostate Gland at 3.0T for Transurethral Ultrasound Therapy

被引:28
作者
Ramsay, Elizabeth [1 ]
Mougenot, Charles [2 ]
Kohler, Max [3 ]
Bronskill, Michael [1 ,4 ]
Klotz, Laurence [5 ]
Haider, Masoom A. [6 ]
Chopra, Rajiv [1 ,4 ,7 ]
机构
[1] Sunnybrook Res Inst, Toronto, ON M4N 3M5, Canada
[2] Philips Healthcare, Toronto, ON, Canada
[3] Philips Healthcare, Vantaa, Uusimaa, Finland
[4] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
[5] Sunnybrook Hlth Sci, Urol, Toronto, ON, Canada
[6] Sunnybrook Hlth Sci, Med Imaging, Toronto, ON, Canada
[7] UT Southwestern Med Ctr, Dept Radiol, Dallas, TX USA
基金
美国国家卫生研究院;
关键词
MR thermometry; thermal therapy; segmented EPI; proton resonance frequency shift; interventional MRI; INTENSITY FOCUSED ULTRASOUND; THERMAL ABLATION; IN-VIVO; TEMPERATURE CONTROL; CANINE PROSTATE; CANCER; TISSUE; EXPERIMENTATION; HYPERTHERMIA; GUIDANCE;
D O I
10.1002/jmri.24063
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo investigate the spatial, temporal, and temperature resolution of a segmented gradient echo echo-planar imaging (EPI) technique as applied to proton resonance frequency (PRF) shift thermometry at 3 T in the human prostate gland, and to determine appropriate sequence parameters for magnetic resonance imaging (MRI)-controlled transurethral ultrasound thermal therapy. Materials and MethodsEleven healthy volunteers (age range 23-58) were scanned at 3 T with a 16-channel torso coil to study the behavior of a gradient echo EPI thermometry sequence. The temperature stability and geometric distortion were assessed for 11 different parameter sets. In a further five volunteers, the prostate T2* was measured. ResultsFor all scan parameters investigated, the temperature standard deviation within the prostate was less than 1 degrees C, while the distortion was less than 1 mm. Temperature stability was best with higher TE values (up to 25 msec), larger voxel sizes and lower EPI factors, but this had to be balanced against requirements for good spatial and temporal resolution. Prostate T2* values ranged from 30-50 msec. ConclusionA good balance between temperature stability and temporal/spatial resolution is obtained with TE = 15 msec, voxel size = 1.14 mm, and EPI factor = 9, resulting in a dynamic scan time of 7.2 seconds for the nine slices. J. Magn. Reson. Imaging 2013;38:1564-1571. (c) 2013 Wiley Periodicals, Inc.
引用
收藏
页码:1564 / 1571
页数:8
相关论文
共 31 条
[1]   Quantitative analysis of 3-D conformal MRI-guided transurethral ultrasound therapy of the prostate: Theoretical simulations [J].
Burtnyk, Mathieu ;
Chopra, Rajiv ;
Bronskill, Michael J. .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2009, 25 (02) :116-131
[2]   Prostate cancer: MR imaging and thermometry during microwave thermal ablation-initial experience [J].
Chen, JC ;
Moriarty, JA ;
Derbyshire, JA ;
Peters, RD ;
Trachtenberg, J ;
Bell, SD ;
Doyle, J ;
Arrelano, R ;
Wright, GA ;
Henkelman, RM ;
Hinks, RS ;
Lok, SY ;
Toi, A ;
Kucharczyk, W .
RADIOLOGY, 2000, 214 (01) :290-297
[3]   Analysis of factors important for transurethral ultrasound prostate heating using MR temperature feedback [J].
Chopra, R ;
Wachsmuth, J ;
Burtnyk, M ;
Haider, MA ;
Bronskill, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (04) :827-844
[4]   MR Imaging-controlled Transurethral Ultrasound Therapy for Conformal Treatment of Prostate Tissue: Initial Feasibility in Humans [J].
Chopra, Rajiv ;
Colquhoun, Alexandra ;
Burtnyk, Mathieu ;
N'djin, William A. ;
Kobelevskiy, Ilya ;
Boyes, Aaron ;
Siddiqui, Kashif ;
Foster, Harry ;
Sugar, Linda ;
Haider, Masoom A. ;
Bronskill, Michael ;
Klotz, Laurence .
RADIOLOGY, 2012, 265 (01) :303-313
[5]   MRI-controlled transurethral ultrasound therapy for localised prostate cancer [J].
Chopra, Rajiv ;
Burtnyk, Mathieu ;
N'Djin, W. Apoutou ;
Bronskill, Michael .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2010, 26 (08) :804-821
[6]   The changing face of prostate cancer [J].
Cooperberg, MR ;
Moul, JW ;
Carroll, PR .
JOURNAL OF CLINICAL ONCOLOGY, 2005, 23 (32) :8146-8151
[7]   Fast magnetic-resonance temperature imaging [J].
deZwart, JA ;
vanGelderen, P ;
Kelly, DJ ;
Moonen, CTW .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 112 (01) :86-90
[8]  
Diederich CJ, 1996, INT J HYPERTHER, V12, P305, DOI 10.3109/02656739609022519
[9]  
GELET A, 1993, J UROLOGIE, V99, P350
[10]  
GELET A, 1993, EUR UROL, V23, P44