Inverse Monte Carlo method in a multilayered tissue model for diffuse reflectance spectroscopy

被引:62
作者
Fredriksson, Ingemar [1 ,2 ]
Larsson, Marcus [1 ]
Stromberg, Tomas [1 ]
机构
[1] Linkoping Univ, Univ Hosp, Dept Biomed Engn, S-58185 Linkoping, Sweden
[2] Perimed AB, Datavagen 9A, S-17543 Jarfalla, Sweden
关键词
inverse problems; multilayers; simulations; spectrometers; tissues; OPTICAL-PROPERTIES; DEPTH; VALIDATION; DIAMETER;
D O I
10.1117/1.JBO.17.4.047004
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Model based data analysis of diffuse reflectance spectroscopy data enables the estimation of optical and structural tissue parameters. The aim of this study was to present an inverse Monte Carlo method based on spectra from two source-detector distances (0.4 and 1.2 mm), using a multi layered tissue model. The tissue model variables include geometrical properties, light scattering properties, tissue chromophores such as melanin and hemoglobin, oxygen saturation and average vessel diameter. The method utilizes a small set of presimulated Monte Carlo data for combinations of different levels of epidermal thickness and tissue scattering. The path length distributions in the different layers are stored and the effect of the other parameters is added in the post-processing. The accuracy of the method was evaluated using Monte Carlo simulations of tissue-like models containing discrete blood vessels, evaluating blood tissue fraction and oxygenation. It was also compared to a homogeneous model. The multi layer model performed better than the homogeneous model and all tissue parameters significantly improved spectral fitting. Recorded in vivo spectra were fitted well at both distances, which we previously found was not possible with a homogeneous model. No absolute intensity calibration is needed and the algorithm is fast enough for real-time processing. (C) 2012 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI: 10.1117/1.JBO.17.4.047004]
引用
收藏
页数:12
相关论文
共 31 条
[1]   White Monte Carlo for time-resolved photon migration [J].
Alerstam, Erik ;
Andersson-Engels, Stefan ;
Svensson, Tomas .
JOURNAL OF BIOMEDICAL OPTICS, 2008, 13 (04)
[2]   Non-invasive measurement of the microvascular properties of non-dysplastic and dysplastic oral leukoplakias by use of optical spectroscopy [J].
Amelink, A. ;
Sterenborg, H. J. C. M. ;
Roodenburg, J. L. N. ;
Witjes, M. J. H. .
ORAL ONCOLOGY, 2011, 47 (12) :1165-1170
[3]   In vivo determination of optical properties of normal and tumor tissue with white light reflectance and an empirical light transport model during endoscopy -: art. no. 034018 [J].
Bargo, PR ;
Prahl, SA ;
Goodell, TT ;
Sleven, RA ;
Koval, G ;
Blair, G ;
Jacques, SL .
JOURNAL OF BIOMEDICAL OPTICS, 2005, 10 (03)
[4]   Determination of epithelial tissue scattering coefficient using confocal microscopy [J].
Collier, T ;
Arifler, D ;
Malpica, A ;
Follen, M ;
Richards-Kortum, R .
IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2003, 9 (02) :307-313
[5]  
Du H, 1998, PHOTOCHEM PHOTOBIOL, V68, P141, DOI 10.1562/0031-8655(1998)068<0141:PACADA>2.3.CO
[6]  
2
[7]   Optical microcirculatory skin model:: assessed by Monte Carlo simulations paired with in vivo laser Doppler flowmetry [J].
Fredriksson, Ingemar ;
Larsson, Marcus ;
Stromberg, Tomas .
JOURNAL OF BIOMEDICAL OPTICS, 2008, 13 (01)
[8]   Accuracy of vessel diameter estimated from a vessel packaging compensation in diffuse reflectance spectroscopy [J].
Fredriksson, Ingemar ;
Larsson, Marcus ;
Stromberg, Tomas .
CLINICAL AND BIOMEDICAL SPECTROSCOPY AND IMAGING II, 2011, 8087
[9]   Model-based quantitative laser Doppler flowmetry in skin [J].
Fredriksson, Ingemar ;
Larsson, Marcus ;
Stromberg, Tomas .
JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (05)
[10]   Measurement depth and volume in laser Doppler flowmetry [J].
Fredriksson, Ingemar ;
Larsson, Marcus ;
Stromberg, Tomas .
MICROVASCULAR RESEARCH, 2009, 78 (01) :4-13