Data-Driven Motion Detection and Event-by-Event Correction for Brain PET: Comparison with Vicra

被引:33
作者
Lu, Yihuan [1 ]
Naganawa, Mika [1 ]
Toyonaga, Takuya [1 ]
Gallezot, Jean-Dominique [1 ]
Fontaine, Kathryn [1 ]
Ren, Silin [2 ]
Revilla, Enette Mae [1 ]
Mulnix, Tim [1 ]
Carson, Richard E. [1 ,2 ]
机构
[1] Yale Univ, Dept Radiol & Biomed Imaging, POB 208048, New Haven, CT 06520 USA
[2] Yale Univ, Dept Biomed Engn, New Haven, CT USA
基金
美国国家卫生研究院;
关键词
PET; data-driven; motion detection; motion correction; event-by-event;
D O I
10.2967/jnumed.119.235515
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Head motion degrades image quality and causes erroneous parameter estimates in tracer kinetic modeling in brain PET studies. Existing motion correction methods include frame-based image registration (FIR) and correction using real-time hardware-based motion tracking (HMT) information. However, FIR cannot correct for motion within 1 predefined scan period, and HMT is not readily available in the clinic since it typically requires attaching a tracking device to the patient. In this study, we propose a motion correction framework with a data-driven algorithm, that is, using the PET raw data itself, to address these limitations. Methods: We propose a data-driven algorithm, centroid of distribution (COD), to detect head motion. In COD, the central coordinates of the line of response of all events are averaged over 1-s intervals to generate a COD trace. A point-to-point change in the COD trace in 1 direction that exceeded a user-defined threshold was defined as a time point of head motion, which was followed by manually adding additional motion time points. All the frames defined by such time points were reconstructed without attenuation correction and rigidly registered to a reference frame. The resulting transformation matrices were then used to perform the final motion-compensated reconstruction. We applied the new COD framework to 23 human dynamic datasets, all containing large head motion, with F-18-FDG (n = 13) and 11C-UCB-J ((R)-1-((3-(11C-methyl-11C)pyridin-4-yl)methyl)-4-(3,4,5-trifl uorophenyl)pyrrolidin-2-one) (n = 10) and compared its performance with FIR and with HMT using Vicra (an optical HMT device), which can be considered the gold standard. Results: The COD method yielded a 1.0% +/- 3.2% (mean +/- SD across all subjects and 12 gray matter regions) SUV difference for F-18-FDG (3.7% +/- 5.4% for 11CUCB-J) compared with HMT, whereas no motion correction (NMC) and FIR yielded -15.7% +/- 12.2% (-20.5% +/- 15.8%) and -4.7% +/- 6.9% (-6.2% +/- 11.0%), respectively. For F-18-FDG dynamic studies, COD yielded differences of 3.6% +/- 10.9% in Ki value as compared with HMT, whereas NMC and FIR yielded -18.0% +/- 39.2% and -2.6% +/- 19.8%, respectively. For 11C-UCB-J, COD yielded 3.7% +/- 5.2% differences in VT compared with HMT, whereas NMC and FIR yielded -20.0% +/- 12.5% and -5.3% +/- 9.4%, respectively. Conclusion: The proposed COD-based data-driven motion correction method outperformed FIR and achieved comparable or even better performance than the Vicra HMT method in both static and dynamic studies.
引用
收藏
页码:1397 / 1403
页数:7
相关论文
共 24 条
[1]   Design of a motion-compensation OSEM list-mode algorithm for resolution-recovery reconstruction for the HRRT [J].
Carson, RE ;
Barker, WC ;
Liow, JS ;
Johnson, CA .
2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, :3281-3285
[2]   Assessing Synaptic Density in Alzheimer Disease With Synaptic Vesicle Glycoprotein 2A Positron Emission Tomographic Imaging [J].
Chen, Ming-Kai ;
Mecca, Adam P. ;
Naganawa, Mika ;
Finnema, Sjoerd J. ;
Toyonaga, Takuya ;
Lin, Shu-fei ;
Najafzadeh, Soheila ;
Ropchan, Jim ;
Lu, Yihuan ;
McDonald, Julia W. ;
Michalak, Hannah R. ;
Nabulsi, Nabeel B. ;
Arnsten, Amy F. T. ;
Huang, Yiyun ;
Carson, Richard E. ;
van Dyck, Christopher H. .
JAMA NEUROLOGY, 2018, 75 (10) :1215-1224
[3]   Motion correction of multi-frame PET data in neuroreceptor mapping: Simulation based validation [J].
Costes, Nicolas ;
Dagher, Alain ;
Larcher, Kevin ;
Evans, Alan C. ;
Collins, D. Louis ;
Reilhac, Anthonin .
NEUROIMAGE, 2009, 47 (04) :1496-1505
[4]  
Dinelle K, 2006, IEEE NUCL SCI CONF R, P3283
[5]  
Feng T, 2016, 2016 IEEE NUCL SCI S, P1
[6]   Kinetic evaluation and test-retest reproducibility of [11C]UCB-J, a novel radioligand for positron emission tomography imaging of synaptic vesicle glycoprotein 2A in humans [J].
Finnema, Sjoerd J. ;
Nabulsi, Nabeel B. ;
Mercier, Joel ;
Lin, Shu-fei ;
Chen, Ming-Kai ;
Matuskey, David ;
Gallezot, Jean-Dominique ;
Henry, Shannan ;
Hannestad, Jonas ;
Huang, Yiyun ;
Carson, Richard E. .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2018, 38 (11) :2041-2052
[7]   Imaging synaptic density in the living human brain [J].
Finnema, Sjoerd J. ;
Nabulsi, Nabeel B. ;
Eid, Tore ;
Detyniecki, Kamil ;
Lin, Shu-fei ;
Chen, Ming-Kai ;
Dhaher, Roni ;
Matuskey, David ;
Baum, Evan ;
Holden, Daniel ;
Spencer, Dennis D. ;
Mercier, Joel ;
Hannestad, Jonas ;
Huang, Yiyun ;
Carson, Richard E. .
SCIENCE TRANSLATIONAL MEDICINE, 2016, 8 (348)
[8]   Parametric Imaging With PET and SPECT [J].
Gallezot, Jean-Dominique ;
Lu, Yihuan ;
Naganawa, Mika ;
Carson, Richard E. .
IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2020, 4 (01) :1-23
[9]   Different partial volume correction methods lead to different conclusions: An 18F-FDG-PET study of aging [J].
Greve, Douglas N. ;
Salat, David H. ;
Bowen, Spencer L. ;
Izquierdo-Garcia, David ;
Schultz, Aaron P. ;
Catana, Ciprian ;
Becker, J. Alex ;
Svarer, Claus ;
Knudsen, Gitte M. ;
Sperling, Reisa A. ;
Johnson, Keith A. .
NEUROIMAGE, 2016, 132 :334-343
[10]   Lower synaptic density is associated with depression severity and network alterations [J].
Holmes, Sophie E. ;
Scheinost, Dustin ;
Finnema, Sjoerd J. ;
Naganawa, Mika ;
Davis, Margaret T. ;
DellaGioia, Nicole ;
Nabulsi, Nabeel ;
Matuskey, David ;
Angarita, Gustavo A. ;
Pietrzak, Robert H. ;
Duman, Ronald S. ;
Sanacora, Gerard ;
Krystal, John H. ;
Carson, Richard E. ;
Esterlis, Irina .
NATURE COMMUNICATIONS, 2019, 10 (1)