Impact of flat panel-imager veiling glare on scatter-estimation accuracy and image quality of a commercial on-board cone-beam CT imaging system

被引:21
|
作者
Lazos, Dimitrios [1 ]
Williamson, Jeffrey F. [1 ]
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
基金
美国国家卫生研究院;
关键词
low frequency drop; veiling glare; point spread function; flat-panel imager; cone-beam CT; beam-stop method; MODULATION TRANSFER-FUNCTION; MONTE-CARLO EVALUATION; EDGE-TEST DEVICE; X-RAY; COMPUTED-TOMOGRAPHY; DIGITAL MAMMOGRAPHY; SPATIAL-RESOLUTION; DETECTOR; ALGORITHM; REMOVAL;
D O I
10.1118/1.4747260
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purposes of this study is to measure the low frequency drop (LFD) of the modulation transfer function (MTF), associated with the long tails of the detector point spread function (PSF) of an on-board flat panel imager and study its impact on cone-beam CT (CBCT) image quality and scatter measurement accuracy. Methods: Two different experimental methods were used to characterize LFD and its associated PSF of a Varian OBI flat-panel detector system: the edge response function (ERF) method and the disk transfer function (DTF) method. PSF was estimated by fitting parametric models to these measurements for four values of the applied voltage (kVp). The resultant PSF was used to demonstrate the effect of LFD on image contrast and CT number accuracy in CBCT images reconstructed from synthetic datasets, as well as, accuracy of scatter measurements with the beam-stop method. Results: The MTFs derived from the measured ERF data revealed LFDs varying from 8% (at 60 kVp) to 10.5% (at 120 kVp), while the intensity of the long PSF tails was found to increase with increasing kVp. The veiling glare line spread functions derived from the ERF and DTF methods were in excellent agreement. Uncorrected veiling glare reduced contrast and the image intensity in CBCT reconstruction, near the phantom periphery (by 67 Hounsfield units in a 20 cm-in-diameter water phantom) and (to a smaller degree) near inhomogeneities. Use of the bow-tie filter mitigated these effects. Veiling glare also resulted in about 10%-15% overestimation of the scatter-to-primary ratio when measured with the beam-stop or beam-stop array method. Conclusions: The long tails of the detector PSF were found to have a modest dependence of beam spectrum, which is reflected on the MTF curve LFD. Our findings show that uncorrected veiling glare can affect quantitative accuracy and contrast in CBCT imaging, based on flat panel imager. In addition, it results in overestimation of the scatter-to-primary ratio, measured with the beam-stop methods. (c) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4747260]
引用
收藏
页码:5639 / 5651
页数:13
相关论文
共 37 条
  • [1] Image quality of flat panel cone-beam CT
    Rose, G
    Wiegert, J
    Schäfer, D
    Fiedler, K
    Conrads, N
    Timmer, J
    Rasche, V
    Noordhoek, N
    Klotz, E
    Koppe, R
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 677 - 683
  • [2] Image quality advances in flat panel cone-beam CT
    Sowards-Emmerd, David
    Elenbaas, Thijs
    Hansis, Eberhard
    van de Haar, Peter
    Shao, Lingxiong
    Timmer, Jan
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [3] A quality assurance procedure to monitor mechanical stability and image quality of an on-board KV cone-beam CT imager
    Ali, I.
    Lovelock, D.
    LoSasso, T.
    Amols, H.
    MEDICAL PHYSICS, 2006, 33 (06) : 2282 - 2283
  • [4] Image quality assessment of synthetic CT from on-board cone-beam CT
    Guo, Alan
    Blais, Danis
    Carrier, Jean-Francois
    Bedwani, Stephane
    MEDICAL PHYSICS, 2021, 48 (08) : 4681 - 4682
  • [5] Image Quality of On-board Cone-beam CT Acquired during VMAT Delivery
    Marchant, T.
    Rowbottom, C.
    Moore, C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S97 - S98
  • [6] Detector Characterization of a New Flat-Panel Imager: Prospects for Image Quality Improvement in Fluoroscopy and Cone-Beam CT
    Mail, N.
    Moseley, D.
    Siewerdsen, J.
    Jaffray, D.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [7] Removal and effects of scatter-glare in cone-beam CT with an amorphous-silicon flat-panel detector
    Poludniowski, G.
    Evans, P. M.
    Kavanagh, A.
    Webb, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (06): : 1837 - 1851
  • [8] Image quality of an investigational imaging panel for use with the imaging beam line cone-beam CT
    Beltran, Chris
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (01): : 76 - 83
  • [9] Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone-beam CT from kV on-board imagers
    Ali, Imad
    Ahmad, Salahuddin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (03): : 180 - 194
  • [10] Correction of couch artifacts in KV cone-beam CT from an on-board imaging system
    Ali, I.
    Lovelock, D.
    Kriminski, S.
    Chang, J.
    Amols, H.
    Ali, I.
    MEDICAL PHYSICS, 2006, 33 (06) : 2280 - 2280