Data-driven respiratory gating based on localized diaphragm sensing in TOF PET

被引:5
作者
Kim, Kyungsang [1 ]
Wang, Mengdie [1 ]
Guo, Ning [1 ]
Schaefferkoetter, Joshua [2 ]
Li, Quanzheng [1 ]
机构
[1] Harvard Med Sch, Gordon Ctr Med Imaging, Dept Radiol, Massachusetts Gen Hosp, Boston, MA 02114 USA
[2] Natl Univ Singapore Hosp, Singapore, Singapore
基金
美国国家卫生研究院;
关键词
positron emission tomography; respiratory gating; diaphragm sensing; time-of-flight PET; EMISSION TOMOGRAPHY/COMPUTED TOMOGRAPHY; TUMOR MOTION; LUNG; CT;
D O I
10.1088/1361-6560/ab9660
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is important to measure the respiratory cycle in positron emission tomography (PET) to enhance the contrast of the tumor as well as the accuracy of its localization in organs such as the lung and liver. Several types of data-driven respiratory gating methods, such as center of mass and principal component analysis, have been developed to directly measure the breathing cycle from PET images and listmode data. However, the breathing cycle is still hard to detect in low signal-to-noise ratio (SNR) data, particularly in low dose PET/CT scans. To address this issue, a time-of-flight (TOF) PET is currently utilized for the data-driven respiratory gating because of its higher SNR and better localization of the region of interest. To further improve the accuracy of respiratory gating with TOF information, we propose an accurate data-driven respiratory gating method, which retrospectively derives the respiratory signal using a localized sensing method based on a diaphragm mask in TOF PET data. To assess the accuracy of the proposed method, the performance is evaluated with three patient datasets, and a pressure-belt signal as the ground truth is compared. In our experiments, we validate that the respiratory signal using the proposed data-driven gating method is well matched to the pressure-belt respiratory signal with less than 5% peak time errors and over 80% trace correlations. Based on gated signals, the respiratory-gated image of the proposed method provides more clear edges of organs compared to images using conventional non-TOF methods. Therefore, we demonstrate that the proposed method can achieve improvements for the accuracy of gating signals and image quality.
引用
收藏
页数:10
相关论文
共 27 条
[11]   Real-time system for respiratory-cardiac gating in positron tomography [J].
Klein, GJ ;
Reutter, BW ;
Ho, MH ;
Reed, JH ;
Huesman, RH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (04) :2139-2143
[12]   Fine-scale motion detection using intrinsic list mode PET information [J].
Klein, GJ ;
Reutter, BW ;
Botvinick, EH ;
Budinger, TF ;
Huesman, RH .
IEEE WORKSHOP ON MATHEMATICAL METHODS IN BIOMEDICAL IMAGE ANALYSIS, PROCEEDINGS, 2001, :71-78
[14]   A PIEZOELECTRIC RESPIRATORY MONITOR FOR INVIVO NMR [J].
MCKIBBEN, CK ;
REO, NV .
MAGNETIC RESONANCE IN MEDICINE, 1992, 27 (02) :338-342
[15]   Quantitation of respiratory motion during 4D-PET/CT acquisition [J].
Nehmeh, SA ;
Erdi, YE ;
Pan, T ;
Yorke, E ;
Mageras, GS ;
Rosenzweig, KE ;
Schoder, H ;
Mostafavi, H ;
Squire, O ;
Pevsner, A ;
Larson, SM ;
Humm, JL .
MEDICAL PHYSICS, 2004, 31 (06) :1333-1338
[16]   Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer [J].
Nehmeh, SA ;
Erdi, YE ;
Ling, CC ;
Rosenzweig, KE ;
Squire, OD ;
Braban, LE ;
Ford, E ;
Sidhu, K ;
Mageras, GS ;
Larson, SM ;
Humm, JL .
MEDICAL PHYSICS, 2002, 29 (03) :366-371
[17]   Respiratory motion in positron emission tomography/computed tomography: A review [J].
Nehmeh, Sadek A. ;
Erdi, Yusuf E. .
SEMINARS IN NUCLEAR MEDICINE, 2008, 38 (03) :167-176
[18]   A comparison of the respiratory signals acquired by different respiratory monitoring systems used in respiratory gated radiotherapy [J].
Otani, Yuki ;
Fukuda, Ichirou ;
Tsukamoto, Nobuhiro ;
Kumazaki, Yu ;
Sekine, Hiroshi ;
Imabayashi, Etsuko ;
Kawaguchi, Osamu ;
Nose, Takayuki ;
Teshima, Teruki ;
Dokiya, Takushi .
MEDICAL PHYSICS, 2010, 37 (12) :6178-6186
[19]   4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT [J].
Pan, T ;
Lee, TY ;
Rietzel, E ;
Chen, GTY .
MEDICAL PHYSICS, 2004, 31 (02) :333-340
[20]   Retrospective data-driven respiratory gating for PET/CT [J].
Schleyer, Paul J. ;
O'Doherty, Michael J. ;
Barrington, Sally F. ;
Marsden, Paul K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (07) :1935-1950