Comparative assessment of a real-time particle monitor against the reference gravimetric method for PM10 and PM2.5 in indoor air

被引:45
作者
Tasic, Visa [1 ]
Jovasevic-Stojanovic, Milena [2 ]
Vardoulakis, Sotiris [3 ]
Milosevic, Novica [1 ]
Kovacevic, Renata [1 ]
Petrovic, Jelena [1 ]
机构
[1] Min & Met Inst Bor, Bor 19210, Serbia
[2] Univ Belgrade, Vinca Inst Nucl Sci, Belgrade 11001, Serbia
[3] Hlth Protect Agcy, Ctr Radiat Chem & Environm Hazards, Chilton OX11 0RQ, Oxon, England
关键词
Particulate matter; Optical monitor; Indoor air pollution; Aerosol monitoring; Exposure assessment; PARTICULATE MATTER; PERSONAL EXPOSURE; POLLUTION; QUALITY; PERFORMANCE; URBAN;
D O I
10.1016/j.atmosenv.2012.02.030
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Accurate monitoring of indoor mass concentrations of particulate matter is very important for health risk assessment as people in developed countries spend approximately 90% of their time indoors. The direct reading, aerosol monitoring device, Turnkey, OSIRIS Particle Monitor (Model 2315) and the European reference low volume sampler, LVS3 (Sven/Leckel LVS3) with size-selective inlets for PM10 and PM2.5 fractions were used to assess the comparability of available optical and gravimetric methods for particulate matter characterization in indoor air. Simultaneous 24-hour samples were collected in an indoor environment for 60 sampling periods in the town of Bor, Serbia. The 24-hour mean PM10 levels from the OSIRIS monitor were well correlated with the LVS3 levels (R-2=0.87) and did not show statistically significant bias. The 24-hour mean PM2.5 levels from the OSIRIS monitor were moderately correlated with the LVS3 levels (R-2=0.71), but show statistically significant bias. The results suggest that the OSIRIS monitor provides sufficiently accurate measurements for PM10. The OSIRIS monitor under-estimated the indoor PM10 concentrations by approximately 12%, relative to the reference LVS3 sampler. The accuracy of PM10 measurements could be further improved through empirical adjustment. For the fine fraction of particulate matter, PM2.5, it was found that the OSIRIS monitor underestimated indoor concentrations by approximately 63%, relative to the reference LVS3 sampler. This could lead to exposure misclassification in health effects studies relying on PM2.5 measurements collected with this instrument in indoor environments. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:358 / 364
页数:7
相关论文
共 37 条
[1]   Characterization of indoor particle sources: A study conducted in the metropolitan Boston area [J].
Abt, E ;
Suh, HH ;
Allen, G ;
Koutrakis, P .
ENVIRONMENTAL HEALTH PERSPECTIVES, 2000, 108 (01) :35-44
[2]   Relationships between personal, indoor, and outdoor exposures to trace elements in PM2.5 [J].
Adgate, John L. ;
Mongin, Steven J. ;
Pratt, Gregory C. ;
Zhang, Junfeng ;
Field, M. Paul ;
Ramachandran, Gurumurthy ;
Sexton, Ken .
SCIENCE OF THE TOTAL ENVIRONMENT, 2007, 386 (1-3) :21-32
[3]  
[Anonymous], GUID IND AIR QUAL SE
[4]  
[Anonymous], AIRB PART MON
[5]  
[Anonymous], 2006, AIR QUAL GUID PART M
[6]  
[Anonymous], PM STAND
[7]   The contribution of ambient sources to particulate pollution in spaces and trains of the Prague underground transport system [J].
Branis, M .
ATMOSPHERIC ENVIRONMENT, 2006, 40 (02) :348-356
[8]   Year-long continuous personal exposure to PM2.5 recorded by a fast responding portable nephelometer [J].
Branis, Martin ;
Kolomaznikova, Jana .
ATMOSPHERIC ENVIRONMENT, 2010, 44 (24) :2865-2872
[9]   Effects of travel mode on exposures to particulate air pollution [J].
Briggs, David J. ;
de Hoogh, Kees ;
Morris, Chloe ;
Gulliver, John .
ENVIRONMENT INTERNATIONAL, 2008, 34 (01) :12-22
[10]   Performance evaluation of the active-flow personal DataRAM PM2.5 mass monitor (Thermo Anderson pDR-1200) designed for continuous personal exposure measurements [J].
Chakrabarti, B ;
Fine, PM ;
Delfino, R ;
Sioutas, C .
ATMOSPHERIC ENVIRONMENT, 2004, 38 (20) :3329-3340