Accuracy and the Proper Use of the Point-of-Care Analysis in Hemoglobin A1C Measurement

被引:1
|
作者
Lynn, Lei [1 ]
Kay, Dana [1 ]
Younes, Naji [2 ]
Sen, Sabyasachi [3 ]
Jain, Vivek [4 ]
机构
[1] George Washington Univ, Dept Internal Med, 2150 Penn Ave NW, Washington, DC 20037 USA
[2] George Washington Univ, Milken Inst Sch Publ Hlth, Dept Epidemiol & Biostat, Washington, DC 20037 USA
[3] George Washington Univ, Dept Endocrinol, Washington, DC 20037 USA
[4] George Washington Univ, Dept Pulm Crit Care & Sleep Med, Washington, DC 20037 USA
关键词
point-of-care; hemoglobin A1C; diabetes;
D O I
10.1089/met.2017.0098
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: Diabetes Mellitus is one of the most common chronic diseases in the United States and worldwide. In 2009 the American Diabetes Association (ADA) recommended using A1C greater than 6.5% to diagnose diabetes. Hemoglobin A1C can be measured by laboratory and Point-of-Care (POC) methods. It is important to compare results obtained by the two modalities to determine the proper use of these tests. Methods: One hundred fifteen patients were identified to have both POC and laboratory A1C values obtained on the same day, between August 2013 and August 2014. The differences were analyzed and the average difference and coefficient of determination (R-2) was calculated. In addition, a Bland-Altman plot was generated; Kappa statistic calculated. We also computed the sensitivity and specificity of the POC method in diagnosing diabetes while using laboratory analysis as the gold standard. Results: The differences between the two measurements ranged from 0 to 3.2 with an average difference of 0.2% of A1C. The Kappa statistic is 0.761 and R-2 was 0.938. Laboratory results were greater than POC in 84% of cases. Interestingly, the Bland-Altman graph also indicates that the variability between the two sets of data increases with higher values of A1C especially for A1C greater than 10%. The POC method has a sensitivity of 88.6% and specificity of 96.3% in diagnosing diabetes. Conclusions: Per ADA guidelines, the POC method is not recommended to be used to diagnose diabetes. Based on our findings 10.3% of individuals would be missed if one used the POC method to diagnose diabetes. However, given its accuracy supported by our findings it is adequate to be used during routine diabetes follow-up visits. The finding from our study can help improve healthcare quality by reducing unnecessary blood draws and clinic visits.
引用
收藏
页码:117 / 121
页数:5
相关论文
共 50 条
  • [1] Point-of-Care Hemoglobin A1c
    O'Brien, Matthew J.
    Sacks, David B.
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2019, 322 (14): : 1404 - 1405
  • [2] Retrospective Accuracy Study of Point-of-Care Hemoglobin A1c in Field Conditions
    Manthei, David M.
    Wesener, Nicholas
    Twarkowski, Denise
    Giacherio, Donald A.
    Schroeder, Lee F.
    CLINICAL CHEMISTRY, 2017, 63 (03) : 780 - 781
  • [3] Evaluation of three hemoglobin A1c point-of-care instruments
    Criel, M.
    Jonckheere, S.
    Langlois, M.
    ACTA CLINICA BELGICA, 2015, 70 : S3 - S3
  • [4] Point-of-care hemoglobin A1c testing in postmortem examination
    Joo-Young Na
    Forensic Science, Medicine and Pathology, 2018, 14 : 322 - 326
  • [5] Assessing the Performance of Point-of-Care Hemoglobin A1c Systems
    Irvin, Ben
    Knaebel, Jennifer
    Simmons, David
    CLINICAL CHEMISTRY, 2010, 56 (08) : 1359 - 1360
  • [6] Point-of-Care Assays for Hemoglobin A1c: Is Performance Adequate?
    Little, Randie R.
    Lenters-Westra, Erna
    Rohlfing, Curt L.
    Slingerland, Robbert
    CLINICAL CHEMISTRY, 2011, 57 (09) : 1333 - 1334
  • [7] Evaluation of Three Hemoglobin A1c Point-of-Care Instruments
    Criel, Marieke
    Jonckheere, Stijn
    Langlois, Michel
    CLINICAL LABORATORY, 2016, 62 (03) : 285 - 291
  • [9] Comparison of costs of hemoglobin A1c determination by point-of-care testing (POCT) and standard hemoglobin A1c analyzer
    Wiwanitkit, V.
    ARCHIVES OF HELLENIC MEDICINE, 2010, 27 (06): : 953 - 955
  • [10] Assessing the Performance of Point-of-Care Hemoglobin A1c Systems Reply
    Lenters-Westra, Erna
    Slingerland, Robbert J.
    CLINICAL CHEMISTRY, 2010, 56 (08) : 1360 - 1361