Tuneable Microsecond-Pulsed Glow Discharge Design for the Simultaneous Acquisition of Elemental and Molecular Chemical Information Using a Time-of-Flight Mass Spectrometer

被引:21
作者
Sola-Vazquez, Auristela [1 ]
Martin, Antonio [1 ]
Costa-Fernandez, Jose M. [1 ]
Pereiro, Rosario [1 ]
Sanz-Medel, Alfredo [1 ]
机构
[1] Univ Oviedo, Dept Phys & Analyt Chem, Fac Chem, E-33006 Oviedo, Spain
关键词
DIRECT SOLID ANALYSIS; GAS-CHROMATOGRAPHY; IONIZATION SOURCE; SPECIATION; QUANTIFICATION; PARAMETERS;
D O I
10.1021/ac802520q
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A microsecond-pulsed direct current glow discharge (GD) was interfaced and synchronized to a time-of-flight mass spectrometer MS(TOF) for time-gated generation and detection of elemental, structural, and molecular ions. In this way, sequential collection of the mass spectra at different temporal regimes occurring during the GD pulse cycle is allowed. The capabilities of this setup were explored using bromochloromethane as model analyte. A simple GD chamber, developed in our laboratory and characterized by a low plasma volume minimizing dilution of the sample but showing great robustness to the entrance of organic compounds in the microsecond-pulsed plasma, has been used. An exhaustive analytical characterization of the GD-MS(TOF) prototype has been performed. Calibration curves for bromochloromethane observed at the different time regimes of the GD pulse cycle (that is, for elemental, fragment, and molecular ions from the analyte) showed very good linearity for the measurement of the different involved ions, with precisions in the range of 7-13% (relative standard deviation). Actual detection limits obtained for bromochloromethane were in the range of 1-3 mu g/L for elements monitoring in the GD pulse "prepeak", in the range of 11 - 13 mu g/L when monitoring analyte fragments in the plateau, and about 238 mu g/L when measuring the molecular peak in the afterpeak regime.
引用
收藏
页码:2591 / 2599
页数:9
相关论文
共 23 条
[1]   Potential of radio frequency glow discharge optical emission spectrometry for the analysis of gaseous samples [J].
Centineo, G ;
Fernandez, M ;
Pereiro, R ;
SanzMedel, A .
ANALYTICAL CHEMISTRY, 1997, 69 (18) :3702-3707
[2]   Quantification of aromatic and halogenated hydrocarbons and alcohol mixtures at the elemental, structural, and parent molecular ion level [J].
Fliegel, D ;
Waddell, R ;
Majidi, V ;
Günther, D ;
Lewis, CL .
ANALYTICAL CHEMISTRY, 2005, 77 (06) :1847-1852
[3]   Evaluation of a pulsed glow discharge time-of-flight mass spectrometer as a detector for gas chromatography and the influence of the glow discharge source parameters on the information volume in chemical speciation analysis [J].
Fliegel, Daniel ;
Fuhrer, Katrin ;
Gonin, Marc ;
Guenther, Detlef .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 386 (01) :169-179
[4]   Development of a direct current gas sampling glow discharge ionization source for the time-of-flight mass spectrometer [J].
Guzowski, JP ;
Broekaert, JAC ;
Ray, SJ ;
Hieftje, GM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1121-1127
[5]   Characterization of switched direct current gas sampling glow discharge ionization source for the time-of-flight mass spectrometer [J].
Guzowski, JP ;
Hieftje, GM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (01) :27-36
[6]   Spectral, spatial and temporal characteristics of a millisecond pulsed glow discharge: metastable argon atom production [J].
Jackson, GP ;
Lewis, CL ;
Doorn, SK ;
Majidi, V ;
King, FL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (12) :2449-2464
[7]   FACTORS INFLUENCING ION SIGNAL PROFILES IN PULSED GLOW-DISCHARGE MASS-SPECTROMETRY [J].
KLINGLER, JA ;
BARSHICK, CM ;
HARRISON, WW .
ANALYTICAL CHEMISTRY, 1991, 63 (22) :2571-2576
[8]   Influence of discharge parameters on real-time chemical speciation for gas chromatography pulsed glow discharge plasma time-of-flight mass spectrometry [J].
Lewis, CL ;
Moser, MA ;
Hang, W ;
Dale, DE ;
Hassell, DC ;
Majidi, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (06) :629-636
[9]   Time-gated pulsed glow discharge: Real-time chemical speciation at the elemental, structural, and molecular level for gas chromatography time-of-flight mass spectrometry [J].
Lewis, CL ;
Moser, MA ;
Dale, DE ;
Hang, W ;
Hassell, C ;
King, FL ;
Majidi, V .
ANALYTICAL CHEMISTRY, 2003, 75 (09) :1983-1996
[10]   IONIZATION METHODS FOR ANALYSIS OF GASES AND VAPORS [J].
LOVELOCK, JE .
ANALYTICAL CHEMISTRY, 1961, 33 (02) :162-&