Simultaneous determination of oxymorphone and its active metabolite 6-OH-oxymorphone in human plasma by high performance liquid chromatography-tandem mass spectrometry

被引:2
作者
Zha, Wuyi [1 ]
Shum, Linyee [1 ]
机构
[1] TDM Pharmaceut Res LLC, Newark, DE 19702 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2012年 / 902卷
关键词
Oxymorphone; 6-OH-oxymorphone; LC-MS/MS; SPE; Human plasma; RAT PLASMA; EXTENDED-RELEASE; OXYCODONE; NOROXYCODONE; QUANTIFICATION; PHASE;
D O I
10.1016/j.jchromb.2012.06.035
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A selective high performance liquid chromatography-tandem mass spectrometric (LC-MS/MS) method for the simultaneous determination of oxymorphone and its active metabolite 6-OH-oxymorphone in human plasma was developed and validated using oxymorphone-d(3) as the internal standard. Chromatographic conditions were optimized to separate oxymorphone from the other metabolite, oxymorphone-3-glucuronide, which may convert to oxymorphone in MS ion source, resulting in inaccurate quantitation of oxymorphone. Solid phase extraction (SPE) was used to extract oxymorphone and 6-OH-oxymorphone from plasma. SPE offered the advantage of being able to remove the unwanted metabolite, oxymorphone-3-glucuronide, through the wash step during the extraction. The developed method was precise and reproducible as shown by good linearity of calibration curves (correlation coefficients >= 0.9968 for oxymorphone and >= 0.9967 for 6-OH-oxymorphone) with high intraday assay and interday assay precision (CV% <= 11.0% for oxymorphone and <= 12.6% for 6-OH-oxymorphone) over a range of 35/25 - 5000/5000 pg/mL for oxymorphone/6-OH-oxymorphone. The method has been successfully applied to analyze oxymorphone and 6-OH-oxymorphone in plasma from 19 healthy volunteers in a bioequivalence study. A total of 1026 samples were analyzed. Good linearity (average correlation coefficient 0.9988 for oxymorphone and 0.9966 for 6-OH-oxymorphone) was achieved with calibration curves and high precision (CV% <= 5.9% for oxymorphone and <= 10.9% for 6-OH-oxymorphone) was obtained with QCs. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:116 / 121
页数:6
相关论文
共 17 条