Study of PET intrinsic spatial resolution and contrast recovery improvement for PET/MRI systems

被引:14
作者
Peng, Hao [1 ,2 ,3 ]
Levin, Craig S. [1 ,2 ,4 ,5 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] MIPS, Stanford, CA 94305 USA
[3] McMaster Univ, Dept Med Phys, Hamilton, ON L8S 4K1, Canada
[4] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[5] Stanford Univ, Dept Phys, Stanford, CA 94305 USA
关键词
POSITRON-EMISSION-TOMOGRAPHY; COMPUTED-TOMOGRAPHY; MAGNETIC-FIELD; MRI; QUANTITATION; DETECTOR;
D O I
10.1088/0031-9155/57/9/N101
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper studied PET intrinsic spatial resolution and contrast recovery improvement for PET/MRI dual modality systems. A Monte Carlo simulation tool was developed to study positron diffusion in tissues with and without a magnetic field for six commonly used isotopes (F-18, C-11, N-13, O-15, Ga-68 and Rb-82). A convolution process was implemented to investigate PET intrinsic spatial resolution, taking into account three factors: positron diffusion range, collinear photon annihilation and finite detector element width. The resolution improvement was studied quantitatively as a function of magnetic field strength for three PET system configurations (whole-body, brain-dedicated and small-animal PET). When the magnetic field strength increases up to 10 T, the system spatial resolution in directions orthogonal to the field for O-15, Ga-68 and Rb-82 is comparable to that of F-18 without the magnetic field. Beyond 10 T, no significant improvement of spatial resolution was observed. In addition, the modulation transfer function was studied to predict the intrinsic contrast recovery improvement for several existing and promising PET/MRI configurations.
引用
收藏
页码:N101 / N115
页数:15
相关论文
共 29 条
[1]  
[Anonymous], 1966, BETA DECAY
[2]  
Attix F H, 1986, INTRO RADIOLOGICAL P
[3]  
Bethe H., 1953, EXPT NUCL PHYS
[4]   Simultaneous in vivo positron emission tomography and magnetic resonance imaging [J].
Catana, Ciprian ;
Procissi, Daniel ;
Wu, Yibao ;
Judenhofer, Martin S. ;
Qi, Jinyi ;
Pichler, Bernd J. ;
Jacobs, Russell E. ;
Cherry, Simon R. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2008, 105 (10) :3705-3710
[5]  
Chatziioannou A, 1997, J NUCL MED, V7, P1164
[6]   Multimodality in vivo imaging systems: Twice the power or double the trouble? [J].
Cherry, Simon R. .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2006, 8 :35-62
[7]   ON THE ANGULAR DISTRIBUTION OF 2-PHOTON ANNIHILATION RADIATION [J].
DEBENEDETTI, S ;
COWAN, CE ;
KONNEKER, WR ;
PRIMAKOFF, H .
PHYSICAL REVIEW, 1950, 77 (02) :205-212
[8]  
Derenzo StephenE., 1979, Precision measurement of annihilation point spread distributions for medically important positron emitters, P819
[9]   An account of the discrepancy between MRI and PET cerebral blood flow measures. A high-field MRI investigation [J].
Donahue, Manus J. ;
Lu, Hanzhang ;
Jones, Craig K. ;
Pekar, James J. ;
van Zijl, Peter C. M. .
NMR IN BIOMEDICINE, 2006, 19 (08) :1043-1054
[10]   USE OF A MAGNETIC-FIELD TO INCREASE THE SPATIAL-RESOLUTION POSITRON EMISSION TOMOGRAPHY [J].
HAMMER, BE ;
CHRISTENSEN, NL ;
HEIL, BG .
MEDICAL PHYSICS, 1994, 21 (12) :1917-1920