Metal artefact reduction in CT imaging of hip prostheses-an evaluation of commercial techniques provided by four vendors

被引:99
作者
Andersson, K. M. [1 ]
Nowik, P. [2 ]
Persliden, J. [1 ]
Thunberg, P. [1 ,3 ]
Norrman, E. [1 ]
机构
[1] Univ Orebro, Fac Med & Hlth, Dept Med Phys, SE-70182 Orebro, Sweden
[2] Karolinska Univ Hosp, Dept Med Phys, Stockholm, Sweden
[3] Univ Orebro, Sch Hlth & Med Sci, SE-70182 Orebro, Sweden
关键词
DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; SOFTWARE;
D O I
10.1259/bjr.20140473
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to evaluate commercial metal artefact reduction (MAR) techniques in X-ray CT imaging of hip prostheses. Methods: Monoenergetic reconstructions of dual-energy CT (DECT) data and several different MAR algorithms, combined with single-energy CT or DECT, were evaluated by imaging a bilateral hip prosthesis phantom. The MAR images were compared with uncorrected images based on CT number accuracy and noise in different regions of interest. Results: The three MAR algorithms studied implied a general noise reduction (up to 67%, 74% and 77%) and an improvement in CT number accuracy, both in regions close to the prostheses and between the two prostheses. The application of monoenergetic reconstruction, without any MAR algorithm, did not decrease the noise in the regions close to the prostheses to the same extent as did the MAR algorithms and even increased the noise in the region between the prostheses. Conclusion: The MAR algorithms evaluated generally improved CT number accuracy and substantially reduced the noise in the hip prostheses phantom images, both close to the prostheses and between the two prostheses. The study showed that the monoenergetic reconstructions evaluated did not sufficiently reduce the severe metal artefact caused by large orthopaedic implants. Advances in knowledge: This study evaluates several commercially available MAR techniques in CT imaging of large orthopaedic implants.
引用
收藏
页数:7
相关论文
共 16 条
[1]   Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning [J].
Andersson, Karin M. ;
Ahnesjo, Anders ;
Dahlgren, Christina Vallhagen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (05) :112-119
[2]   Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation [J].
Bamberg, Fabian ;
Dierks, Alexander ;
Nikolaou, Konstantin ;
Reiser, Maximilian F. ;
Becker, Christoph R. ;
Johnson, Thorsten R. C. .
EUROPEAN RADIOLOGY, 2011, 21 (07) :1424-1429
[3]   Artifacts in CT: Recognition and avoidance [J].
Barrett, JF ;
Keat, N .
RADIOGRAPHICS, 2004, 24 (06) :1679-1691
[4]   Metal Artifact Reduction Software Used With Abdominopelvic Dual- Energy CT of Patients With Metal Hip Prostheses: Assessment of Image Quality and Clinical Feasibility [J].
Han, Seung Chol ;
Chung, Yong Eun ;
Lee, Young Han ;
Park, Kwan Kyu ;
Kim, Myeong Jin ;
Kim, Ki Whang .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (04) :788-795
[5]  
Healthcare P., 2012, White Paper
[6]   The CT number accuracy of a novel commercial metal artifact reduction algorithm for large orthopedic implants [J].
Hilgers, Guido ;
Nuver, Tonnis ;
Minken, Andre .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (01) :274-278
[7]   Reduction of dental metallic artefacts in CT: Value of a newly developed algorithm for metal artefact reduction (O-MAR) [J].
Kidoh, M. ;
Nakaura, T. ;
Nakamura, S. ;
Tokuyasu, S. ;
Osakabe, H. ;
Harada, K. ;
Yamashita, Y. .
CLINICAL RADIOLOGY, 2014, 69 (01) :E11-E16
[8]   Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multidetector CT [J].
Lee, Mi-Jung ;
Kim, Sungjun ;
Lee, Sung-Ah ;
Song, Ho-Taek ;
Huh, Yong-Min ;
Kim, Dae-Hong ;
Han, Seung Hwan ;
Suh, Jin-Suck .
RADIOGRAPHICS, 2007, 27 (03) :791-803
[9]   Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software [J].
Lee, Young Han ;
Park, Kwan Kyu ;
Song, Ho-Taek ;
Kim, Sungjun ;
Suh, Jin-Suck .
EUROPEAN RADIOLOGY, 2012, 22 (06) :1331-1340
[10]   Reducing the effects of metal artefact using high keV monoenergetic reconstruction of dual energy CT (DECT) in hip replacements [J].
Lewis, Mark ;
Reid, Karen ;
Toms, Andoni P. .
SKELETAL RADIOLOGY, 2013, 42 (02) :275-282