Fatigue Fracture in Dual Modular Revision Total Hip Arthroplasty Stems Failure Analysis and Computed Tomography Diagnostics in Two Cases

被引:50
作者
Norman, Pontus [1 ]
Iyengar, Srinivasan [2 ]
Svensson, Ingrid [3 ]
Flivik, Gunnar [1 ]
机构
[1] Lund Univ, Dept Orthopaed, Skane Univ Hosp, Lund, Sweden
[2] Lund Univ, Div Mat Engn, Lund, Sweden
[3] Lund Univ, Div Solid Mech, Lund, Sweden
关键词
modular; hip prosthesis; fatigue failure; CT; FEMORAL STEMS; COMPONENT; CORROSION; IMPLANT; REPLACEMENT; JUNCTION;
D O I
10.1016/j.arth.2013.09.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report on two patients with fracture of a modular, tapered and distally fixed, uncemented titanium revision hip stem, not previously described. A failure analysis revealed that the cause of the fractures was the development of fatigue cracks in the mid-stem cobalt-chromium modular junction ending in corrosion-fatigue failure. No material defects or stress risers were found in any of the implants. The diameter of the mid-stem modular junction might be undersized for use in heavy and active patients. We also report a new way of detecting an undisplaced fracture at the modular junction, using the scout image from a computed tomography (CT) scan; a technique that can be used when plain radiographs are inconclusive. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:850 / 855
页数:6
相关论文
共 26 条
[1]   Corrosion-Induced Fracture of a Double-Modular Hip Prosthesis A Case Report [J].
Atwood, Sara A. ;
Patten, Eli W. ;
Bozic, Kevin J. ;
Pruitt, Lisa A. ;
Ries, Michael D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (06) :1522-1525
[2]   Fractures of distally-fixed femoral stems after revision arthroplasty [J].
Busch, CA ;
Charles, MN ;
Bourne, RB ;
Rorabeck, CH ;
MacDonald, SJ ;
McCalden, RW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2005, 87B (10) :1333-1336
[3]   Fatigue fracture of a proximally modular, distally tapered fluted implant with diaphyseal fixation [J].
Buttaro, Martin A. ;
Mayor, Michael B. ;
Van Citters, Douglas ;
Piccaluga, Francisco .
JOURNAL OF ARTHROPLASTY, 2007, 22 (05) :780-783
[4]   FRACTURE OF THE FEMORAL COMPONENT AFTER TOTAL HIP-REPLACEMENT - AN ANALYSIS OF 58 CASES [J].
CHAO, EYS ;
COVENTRY, MB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1981, 63 (07) :1078-1094
[5]  
CHARNLEY J, 1975, CLIN ORTHOP RELAT R, P105
[6]   Use of magnetic resonance imaging in the diagnosis of an occult fracture of the femoral component after total hip arthroplasty [J].
Cook, SM ;
Pellicci, PM ;
Potter, HG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :149-153
[7]   Late fatigue fracture of a modern cemented forged cobalt chrome stem for total hip arthroplasty -: A report of 10 cases [J].
Della Valle, AG ;
Becksaç, B ;
Anderson, J ;
Wright, T ;
Nestor, B ;
Pellicci, PM ;
Salvati, EA .
JOURNAL OF ARTHROPLASTY, 2005, 20 (08) :1084-1088
[8]  
Efe Turgay, 2011, J Arthroplasty, V26, DOI 10.1016/j.arth.2010.05.020
[9]  
Ellman MB, 2013, J ARTHROPLASTY, V28, P196, DOI DOI 10.1016/J.ARTH.2011.05.024
[10]   Short-term results of hip revisions with a curved cementless modular stem in association with the surgical approach [J].
Fink, Bernd ;
Grossmann, Alexandra ;
Schubring, Svenja ;
Schulz, Martin S. ;
Fuerst, Martin .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (01) :65-73