Multisectored interstitial ultrasound applicators for dynamic angular control of thermal therapy

被引:28
作者
Kinsey, Adam M.
Diederich, Chris J. [1 ]
Tyreus, P. Daniel
Nau, William H.
Rieke, Viola
Pauly, Kim Butts
机构
[1] Univ Calif San Francisco, Thermal Therapy Res Grp, Dept Radiat Oncol, San Francisco, CA 94115 USA
[2] Univ Calif Berkeley, Joint Grad Grp Bioengn, San Francisco, CA 94115 USA
[3] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
关键词
multisectored interstitial ultrasound; thermal therapy; thermal ablation; hyperthermia; magnetic resonance temperature imaging;
D O I
10.1118/1.2184443
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Dynamic angular control of thermal ablation, and hyperthermia therapy with current interstitial heating technology is limited in capability, and often relies upon nonadjustable angular power deposition patterns and/or mechanical manipulation of the heating device. The objective of this study was to investigate the potential of multisectored tubular interstitial ultrasound devices to provide control of the angular heating distribution without device manipulation. Multisectored tubular transducers with independent sector power control were incorporated into modified versions of internally cooled (1.9 mm OD) and catheter-cooled (2.4 mm OD) interstitial ultrasound applicators in this work. The heating capabilities of these multisectored devices were evaluated by measurements of acoustic output properties, measurements of thermal lesions produced in ex vivo tissue samples, biothermal simulations of thermal ablation and hyperthermia treatments, and MR temperature imaging of ex vivo and in vivo experiments. Acoustic beam measurements of each applicator type displayed a 35 degrees-40 degrees acoustic dead zone between each independent sector, with negligible mechanical or electrical coupling. Thermal lesions produced in ex vivo liver tissue with one, two, or three sectors activated ranged from 13-18 mm in radius with contiguous zones of coagulation between active sectors. The simulations demonstrated the degree of angular control possible by using variable power levels applied to each sector, variable duration of applied constant power to individual sectors, respectively, or a multipoint temperature controller to vary the power applied to each sector. Despite the acoustic dead zone between sectors, the simulations also showed that the variance from the maximum lesion radius with three elements activated is within 4%-13% for tissue perfusions from 1-10 kg m(-3) s(-1). Simulations of hyperthermia with maximum tissue temperatures of 45 degrees C and 48 degrees C displayed radial penetration up to 2 cm of the 40 degrees C steady-state contour. Thermal characterizations of trisectored applicators in ex vivo and in vivo muscle, using real-time MR thermal imaging, reinforced angular controllability and negligible radial variance of the heating pattern from the applicators, demonstrated effective heating penetration, and displayed MR compatibility. The multisectored interstitial ultrasound applicators developed in this study demonstrated a significant degree of dynamic angular control of a heating pattern without device manipulation, while maintaining heat penetration consistent with previously reported results from other interstitial ultrasound applicators. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1352 / 1363
页数:12
相关论文
共 45 条
[1]   Ablative approaches to the minimally invasive treatment of breast cancer [J].
Agnese, DM ;
Burak, WE .
CANCER JOURNAL, 2005, 11 (01) :77-82
[2]   DIFFERENTIAL THERMAL SENSITIVITY OF TUMOR AND NORMAL TISSUE MICROVASCULAR RESPONSE DURING HYPERTHERMIA [J].
BROWN, SL ;
HUNT, JW ;
HILL, RP .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (04) :501-514
[3]  
Chen L, 2001, J MAGN RESON IMAGING, V13, P23, DOI 10.1002/1522-2586(200101)13:1<23::AID-JMRI1004>3.0.CO
[4]  
2-G
[5]   Method for MRI-guided conformal thermal therapy of prostate with planar transurethral ultrasound heating applicators [J].
Chopra, R ;
Burtnyk, M ;
Haider, MA ;
Bronskill, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21) :4957-4975
[6]   Multifrequency ultrasound transducers for conformal interstitial thermal therapy [J].
Chopra, R ;
Luginbuhl, C ;
Foster, FS ;
Bronskill, MJ .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (07) :881-889
[7]   Interstitial ultrasound heating applicator for MR-guided thermal therapy [J].
Chopra, R ;
Luginbuhl, C ;
Weymouth, AJ ;
Foster, FS ;
Bronskill, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (12) :3133-3145
[8]   Feasibility of linear arrays for interstitial ultrasound thermal therapy [J].
Chopra, R ;
Bronskill, MJ ;
Foster, FS .
MEDICAL PHYSICS, 2000, 27 (06) :1281-1286
[9]  
D'Ippolito Giuseppe, 2002, Tech Vasc Interv Radiol, V5, P141
[10]   Dependence of ultrasonic attenuation and absorption in dog soft tissues on temperature and thermal dose [J].
Damianou, CA ;
Sanghvi, NT ;
Fry, FJ ;
MaassMoreno, R .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 102 (01) :628-634