Double chamber cross-talk, the ultimate double-cross

被引:1
作者
Almehmadi, Fahad [1 ,2 ]
Duffett, Stephen [1 ,3 ]
Khan, Habib R. [1 ]
Klein, George [1 ]
Yee, Raymond [1 ]
机构
[1] Univ Western Ontario, London Hlth Sci Ctr, Heart Rhythm Program, London, ON, Canada
[2] King Saud bin Abdulaziz Univ Hlth Sci, King Faisal Cardiac Ctr, Natl Guard Hlth Affairs, Jeddah, Saudi Arabia
[3] Mem Univ, Dept Med, St John, NF, Canada
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2020年 / 43卷 / 12期
关键词
cross-talk; implantable cardioverter defibrillator; oversensing; PACEMAKER; MODE;
D O I
10.1111/pace.13998
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Cross-talk is a well-described phenomenon in a dual-chamber cardiovascular implantable electronic device. Far-field ventricular events are more commonly sensed in the atrial channel, the reverse is uncommon, and seeing both at the same time has never been reported. We present a case of double cross-talk in a dual-chamber Medtronic (R) implantable cardioverter-defibrillator. In this report, we decipher an unusual device response to the cross-talk and describe the programming changes required to resolve it.
引用
收藏
页码:1599 / 1604
页数:6
相关论文
共 10 条
  • [1] Paradoxical induction of cross-talk in a dual chamber pacemaker by increasing the ventricular output
    Barold, SS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (05): : 906 - 907
  • [2] Sensing and detection in Medtronic implantable cardioverter defibrillators; [Wahrnehmung und Detektion durch implantierbare Kardioverter-Defibrillatoren von Medtronic]
    Brown M.L.
    Swerdlow C.D.
    [J]. Herzschrittmachertherapie + Elektrophysiologie, 2016, 27 (3) : 193 - 212
  • [3] Failure of automatic mode switching: Recognition and management
    Ellenbogen, KA
    Mond, HG
    Wood, MA
    Barold, SS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (02): : 268 - 275
  • [4] Management of far-field R wave sensing for the avoidance of inappropriate mode switch in dual chamber pacemakers:: Results of the FFS-Test Study
    Kolb, Christof
    Wille, Bernd
    Maurer, Dominik
    Schuchert, Andreas
    Weber, Ralf
    Schibgilla, Volker
    Klein, Norbert
    Huemmer, Alexander
    Schmitt, Claus
    Zrenner, Bernhard
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (09) : 992 - 997
  • [5] The managed ventricular pacing algorithm can be misinterpreted as pacemaker malfunction
    Mugnai, Giacomo
    Tomei, Ruggero
    Vassanelli, Corrado
    [J]. EUROPACE, 2013, 15 (12): : 1783 - 1783
  • [6] Pacemaker Syndrome During Managed Ventricular Pacing Mode: What is the Mechanism?
    Pascale, Patrizio
    Pruvot, Etienne
    Graf, Denis
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2009, 20 (05) : 574 - 576
  • [7] Cross-talk in a dual-chamber defibrillator presenting as pacemaker alternans
    Shalaby, AA
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2003, 14 (03) : 325 - 328
  • [8] Troubleshooting Implantable Cardioverter-Defibrillator Sensing Problems II
    Swerdlow, Charles D.
    Asirvatham, Samuel J.
    Ellenbogen, Kenneth A.
    Friedman, Paul A.
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2015, 8 (01) : 212 - 220
  • [9] Implantable Cardioverter-Defibrillator Leads Design, Diagnostics, and Management
    Swerdlow, Charles D.
    Ellenbogen, Kenneth A.
    [J]. CIRCULATION, 2013, 128 (18) : 2062 - 2071
  • [10] Inhibition of bradycardia pacing and detection of ventricular fibrillation due to far-field atrial sensing in a triple cardioverter defibrillator
    Vollmann, D
    Lüthje, L
    Görtler, G
    Unterberg, C
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (10): : 1513 - 1516