Interventional MR-guided excisional biopsy of breast lesions

被引:34
|
作者
Gould, SWT [1 ]
Lamb, G [1 ]
Lomax, D [1 ]
Gedroyc, W [1 ]
Darzi, A [1 ]
机构
[1] St Marys Hosp, NHS Trust, Minimal Access Surg Unit, London W2 1NY, England
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1998年 / 8卷 / 01期
关键词
interventional MR; image-guided surgery; breast lesion;
D O I
10.1002/jmri.1880080110
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Interventional MR (IMR) machines have produced unique opportunities for image-guided surgery. The open configuration design and fast pulse sequences allow intraoperative scanning to monitor procedures. This study was undertaken to assess the potential use of IMR for image-guided surgery. Benign breast lesion excision was chosen as an uncomplicated surgical model. Ten female patients with known benign tumors underwent excision biopsy under general anesthesia in a Signs SP10 .5-T IMR machine (General Electric Medical Systems, Milwaukee, WI), Lesions were localized with precontrast and postcontrast (intravenous gadolinium-diethylenetriamine pentaacetic acid,.2 mmol/kg) fast multiplanar spoiled gradient-recalled acquisition in the steady state (GRASS) sequences. Preoperative "real-time" fast gradient-recalled sequences were also obtained using the Flashpoint (General Electric Medical Systems, Milwaukee, WI) tracker device. The maximum dimensions of each lesion were measured from the resulting images. Excision was performed using titanium instruments and an ultrasonically activated scalpel. Intraoperative real-time scanning demonstrated the resection margin and confirmed complete excision. The maximum dimensions of the macroscopic specimens were compared with those from the MR images. All tumors were visualized with the Signa scanner and real-time imaging and the images were enhanced after intravenous contrast. Maximum dimensions on histologic examination were not significantly different from those measured from Signs (P > .17) or real-time images (p > .4). There was no significant difference between lesion size from Signs and real-time images (P > .25). All postprocedure scans demonstrated complete excision. There were six fibroadenomas, two foci of sclerosing adenosis, one area of fibrocystic disease, and one schwannoma, Intraoperative MR scanning reliably identifies palpable breast tumors and can accurately guide surgical excision. Further work using MR guidance can now be performed in other general surgical areas.
引用
收藏
页码:26 / 30
页数:5
相关论文
共 50 条
  • [1] MR-guided percutaneous excisional and incisional biopsy of breast lesions
    S. H. Heywang-Köbrunner
    A. Heinig
    U. Schaumlöffel
    P. Viehweg
    J. Buchmann
    D. Lampe
    R.-P. Spielmann
    European Radiology, 1999, 9 : 1656 - 1665
  • [2] MR-guided percutaneous excisional and incisional biopsy of breast lesions
    Heywang-Köbrunner, SH
    Heinig, A
    Schaumlöffel, U
    Viehweg, P
    Buehmann, J
    Lampe, D
    Spielmann, RP
    EUROPEAN RADIOLOGY, 1999, 9 (08) : 1656 - 1665
  • [3] Value of MR-guided localization and biopsy in breast lesions
    Obenauer, S
    Grabbe, E
    Knollmann, F
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2006, 178 (05): : 477 - 483
  • [4] MR-guided breast biopsy
    SH Heywang-Köbrunner
    H Sittek
    X Prat
    L Baath
    Breast Cancer Research, 2 (Suppl 2)
  • [5] MR-guided interventional breast procedures considering vacuum biopsy in particular
    Viehweg, P
    Heinig, A
    Amaya, B
    Alberich, T
    Laniado, M
    Heywang-Köbrunner, SH
    EUROPEAN JOURNAL OF RADIOLOGY, 2002, 42 (01) : 32 - 39
  • [6] Stereotaxic add-on device for MR-guided biopsy of breast lesions
    Doler, W
    Fischer, U
    Metzger, I
    Harder, D
    Grabbe, E
    RADIOLOGY, 1996, 200 (03) : 863 - 864
  • [7] Radiologic-pathologic correlation from MR-guided breast biopsy of lesions identified at breast MR
    Langlas, TS
    Pai, VR
    Gregory, NE
    Swinford, A
    Miller, PA
    Rebner, M
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (04) : 72 - 72
  • [8] MR-guided vacuum biopsy of 206 contrast-enhancing breast lesions
    Perlet, C
    Schneider, P
    Amaya, B
    Grosse, A
    Sittek, H
    Reiser, MF
    Heywang-Köbrunner, SH
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (01): : 88 - 95
  • [9] MR-guided percutaneous vacuum biopsy of breast lesions: Experiences with 100 leisons
    Heywang-Koebrunner, SH
    Schaumloeffel-Schulze, U
    Heinig, A
    Beck, RM
    Lampe, D
    Buchmann, J
    RADIOLOGY, 1999, 213P : 289 - 289
  • [10] MR-guided percutaneous vacuum biopsy of breast lesions:: Experiences with 100 cases
    Heywang-Köbrunner, SH
    Schaumlöffel-Schulze, U
    Beck, R
    Lampe, D
    Buchmann, J
    EUROPEAN JOURNAL OF CANCER, 1999, 35 : S307 - S307