An instrument for measuring blood pressure and assessing cardiovascular health from the fingertip

被引:35
作者
Panula, Tuukka [1 ]
Koivisto, Tero [1 ]
Pankaala, Mikko [1 ]
Niiranen, Teemu [2 ]
Kantola, Ilkka [3 ]
Kaisti, Matti [1 ,4 ]
机构
[1] Univ Turku, Dept Future Technol, Turku 20520, Finland
[2] Univ Turku, Dept Clin Med, Turku 20520, Finland
[3] Turku Univ Hosp, Turku 20521, Finland
[4] Imperial Coll London, Dept Bioengn, London SW7 2AZ, England
关键词
Blood pressure; Central blood pressure; Tonometry; Oscillometric method; Fingertip; CLINICAL-ASSESSMENT; MEASURING DEVICES; WAVE-FORM; VALIDATION; DISEASE; HYPERTENSION; PREDICTOR; SOCIETY; SENSOR;
D O I
10.1016/j.bios.2020.112483
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Despite blood pressure being one the leading modifiable risk factors for cardiovascular disease and death, it is severely under-monitored. For this challenge we propose a finger artery non-invasive tono-oscillometric monitor (FANTOM) which is an automated low-cost instrument for measuring blood pressure and hemodynamic parameters from the fingertip. The sensing technology is highly scalable and could be integrated to a pulse oximeter probe for increased patient comfort. A tonometric cuff-less mechatronic system is used to apply pressure on the fingertip for (i) measuring oscillometric blood pressure, (ii) recording arterial waveform and for (iii) constructing central blood pressure (CBP) waveform. Clinical study on volunteers (n = 33) was performed against a commercially available arm cuff device yielding systolic and diastolic readings ((mean +/- SD) mmHg) of (-0.9 +/- 7.3) mmHg and (-3.3 +/- 6.6) mmHg respectively. The results comply with the Association for the Advancement of Medical Instrumentation (AAMI) standard for non-invasive blood pressure monitors. The arterial pulse recording morphology was compared against a volume clamp device (CNSystems CNAP 500) (n = 3) resulting in similar performance. Comparison of CBP against a pulse wave analysis (PWA) device (Atcor Medical Sphygmocor XCEL) (n = 5) revealed central aortic systolic pulse (CASP) and central augmentation index (cAIx) estimates with precision and accuracy of (2.0 +/- 3.7) mmHg and (1.4 +/- 6.2)% respectively. In conclusion, the results indicate that the proposed technology could be useful in the development of new portable or wearable blood pressure monitors. The sensing technology is highly scalable and could be integrated to a pulse oximeter probe for increased patient comfort.
引用
收藏
页数:9
相关论文
共 46 条
[1]   World Medical Association Declaration of Helsinki Ethical Principles for Medical Research Involving Human Subjects [J].
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2013, 310 (20) :2191-2194
[2]  
[Anonymous], 2017, TECHNOLOGY PRACTICAL
[3]   Oscillometric measurement of systolic and diastolic blood pressures validated in a physiologic mathematical model [J].
Babbs, Charles F. .
BIOMEDICAL ENGINEERING ONLINE, 2012, 11
[4]   Review: Semiconductor Piezoresistance for Microsystems [J].
Barlian, A. Alvin ;
Park, Woo-Tae ;
Mallon, Joseph R., Jr. ;
Rastegar, Ali J. ;
Pruitt, Beth L. .
PROCEEDINGS OF THE IEEE, 2009, 97 (03) :513-552
[5]   How to assess mean blood pressure properly at the brachial artery level [J].
Bos, Willem J. W. ;
Verrij, Elisabeth ;
Vincent, Hieronymus H. ;
Westerhof, Berend E. ;
Parati, Gianfranco ;
van Montfrans, Gert A. .
JOURNAL OF HYPERTENSION, 2007, 25 (04) :751-755
[6]   Large Artery Stiffness Assessment Using SphygmoCor Technology [J].
Butlin, Mark ;
Qasem, Ahmad .
PULSE, 2016, 4 (04) :180-192
[7]   Using the Global Burden of Disease Study to Assist Development of Nation-Specific Fact Sheets to Promote Prevention and Control of Hypertension and Reduction in Dietary Salt: A Resource From the World Hypertension League [J].
Campbell, Norm R. C. ;
Lackland, Daniel T. ;
Liu Lisheng ;
Niebylski, Mark L. ;
Nilsson, Peter M. ;
Zhang, Xin-Hua .
JOURNAL OF CLINICAL HYPERTENSION, 2015, 17 (03) :165-167
[8]   Formulas to Explain Popular Oscillometric Blood Pressure Estimation Algorithms [J].
Chandrasekhar, Anand ;
Yavarimanesh, Mohammad ;
Hahn, Jin-Oh ;
Sung, Shih-Hsien ;
Chen, Chen-Huan ;
Cheng, Hao-Min ;
Mukkamala, Ramakrishna .
FRONTIERS IN PHYSIOLOGY, 2019, 10
[9]   An iPhone Application for Blood Pressure Monitoring via the Oscillometric Finger Pressing Method [J].
Chandrasekhar, Anand ;
Natarajan, Keerthana ;
Yavarimanesh, Mohammad ;
Mukkamala, Ramakrishna .
SCIENTIFIC REPORTS, 2018, 8
[10]   Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method [J].
Chandrasekhar, Anand ;
Kim, Chang-Sei ;
Naji, Mohammed ;
Natarajan, Keerthana ;
Hahn, Jin-Oh ;
Mukkamala, Ramakrishna .
SCIENCE TRANSLATIONAL MEDICINE, 2018, 10 (431)