PulseCath iVAC 3L™ hemodynamic performance for simple assisted flow
被引:3
作者:
Niclauss, Lars
[1
]
von Segesser, Ludwig Karl
[1
]
机构:
[1] Univ Hosp CHUV, Dept Cardiovasc Surg, CH-1011 Lausanne, Switzerland
关键词:
PulseCath;
Left ventricular assist device;
Flow performance;
PULSATILE CATHETER PUMP;
D O I:
10.1510/icvts.2010.264051
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
The PulseCath iVAC 3L (TM) left ventricular assist device is an option to treat transitory left heart failure or dysfunction post-cardiac surgery. Assisted blood flow should reach up to 3 l/min. In the present in vitro model exact pump flow, depending on various frequencies and afterload was examined. Optimal flow was achieved with inflation/deflation frequencies of about 70-80/min. The maximal flow rate was achieved at about 2.5 l/min with a minimal afterload of 22 mmHg. Handling of the device was easy due to the connection to a standard intra-aortic balloon pump console. With increasing afterload (up to a simulated mean systemic pressure of 66 mmHg) flow rate and cardiac support are in some extent limited. (C) 2011 Published by European Association for Cardio-Thoracic Surgery. All rights reserved.