Cytotoxic, allergic and genotoxic activity of a nickel-titanium alloy

被引:287
作者
Wever, DJ
Veldhuizen, AG
Sanders, MM
Schakenraad, JM
vanHorn, JR
机构
[1] BIOSCAN,LAB MED DEVICE EVALUAT,BILTHOVEN,NETHERLANDS
[2] ACROMED BV,ROTTERDAM,NETHERLANDS
关键词
NiTi; memory metal; biocompatibility; biological safety;
D O I
10.1016/S0142-9612(97)00041-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The nearly equiatomic nickel-titanium (NiTi) alloy is known for its shape memory properties. These properties can be put to excellent use in various biomedical applications, such as wires for orthodontic tooth alignment and osteosynthesis staples. The aim of this study was to evaluate the short-term biological safety of the NiTi alloy. We carried out an end-point dilution minimal essential medium (MEM) extract cytotoxicity test, a guinea-pig sensitization test and two genotoxicity tests: the Salmonella reverse mutation test and the chromosomal aberration test. The NiTi alloy showed no cytotoxic, allergic or genotoxic activity, similar to the clinical reference control material AISI 316 LVM stainless steel. This promising biological behaviour was most likely due to a minimal release of ions and in that way a reflection of the good corrosion resistance of the NiTi alloy. Given these very good results, together with the good tissue compatibility as shown in several implantation studies in the literature, the NiTi alloy can be regarded as a biologically safe implant material with many promising clinical applications. (C) 1997 Elsevier Science Limited. All rights reserved.
引用
收藏
页码:1115 / 1120
页数:6
相关论文
共 24 条
  • [2] ASSAD M, 1994, P 40 ANN M ORTH RES, P591
  • [3] BENSMANN G, 1982, TECH MITT KRUPP FORS, V40, P123
  • [4] Buehler WJ., 1968, Ocean Engineering, V1, P105, DOI [DOI 10.1016/0029-8018(68)90019-X, 10.1016/0029-8018(68)90019-X, DOI 10.1016/0029]
  • [5] Castleman L.S., 1981, BIOCOMPATIBILITY CLI, V1, P129
  • [6] BIOCOMPATIBILITY OF NITINOL ALLOY AS AN IMPLANT MATERIAL
    CASTLEMAN, LS
    MOTZKIN, SM
    ALICANDRI, FP
    BONAWIT, VL
    JOHNSON, AA
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1976, 10 (05): : 695 - 731
  • [7] NITINOL INTRAVASCULAR STENT - RESULTS OF PRECLINICAL EVALUATION
    CRAGG, AH
    DEJONG, SC
    BARNHART, WH
    LANDAS, SK
    SMITH, TP
    [J]. RADIOLOGY, 1993, 189 (03) : 775 - 778
  • [8] CUTWRIGHT DE, 1973, ORAL SURG, V35, P578
  • [9] CELL-DAMAGE IN-VITRO FOLLOWING DIRECT-CONTACT WITH FINE PARTICLES OF TITANIUM, TITANIUM-ALLOY AND COBALT-CHROME-MOLYBDENUM ALLOY
    EVANS, EJ
    [J]. BIOMATERIALS, 1994, 15 (09) : 713 - 717
  • [10] THE BIOLOGIC EFFECTS OF IMPLANT MATERIALS
    GALANTE, JO
    LEMONS, J
    SPECTOR, M
    WILSON, PD
    WRIGHT, TM
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (05) : 760 - 775