Effect of framework material and vertical misfit on stress distribution in implant-supported partial prosthesis under load application: 3-D finite element analysis

被引:67
作者
Bacchi, Atais [1 ]
Xediek Consani, Rafael Leonardo [1 ]
Mesquita, Marcelo Ferraz [1 ]
Fernandes dos Santos, Mateus Bertolini [1 ]
机构
[1] Univ Estadual Campinas, Piracicaba Dent Sch, Dept Prosthodont & Periodont, BR-13414903 Piracicaba, SP, Brazil
关键词
biomechanics; dental implant; dental prosthesis; implant-supported; finite element analysis; osseointegration; OVERDENTURE-RETAINING BAR; BONE TISSUE; FIT; OSSEOINTEGRATION; ALLOYS;
D O I
10.3109/00016357.2012.757644
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. This study evaluated the influence of framework material and vertical misfit on stress created in an implant-supported partial prosthesis under load application. Materials and methods. The posterior part of a severely reabsorbed jaw with a fixed partial prosthesis above two osseointegrated titanium implants at the place of the second premolar and second molar was modeled using SolidWorks 2010 software. Finite element models were obtained by importing the solid model into an ANSYS Workbench 11 simulation. The models were divided into 15 groups according to their prosthetic framework material (type IV gold alloy, silver-palladium alloy, commercially pure titanium, cobalt-chromium alloy or zirconia) and vertical misfit level (10 mm, 50 mm and 100 mm). After settlement of the prosthesis with the closure of the misfit, simultaneous loads of 110 N vertical and 15 N horizontal were applied on the occlusal and lingual faces of each tooth, respectively. The data was evaluated using Maximum Principal Stress (framework, porcelain veneer and bone tissue) and a von Mises Stress (retention screw) provided by the software. Results. As a result, stiffer frameworks presented higher stress concentrations; however, these frameworks led to lower stresses in the porcelain veneer, the retention screw (faced to 10 mm and 50 mm of the misfit) and the peri-implant bone tissues. Conclusion. The increase in the vertical misfit resulted in stress values increasing in all of the prosthetic structures and peri-implant bone tissues. The framework material and vertical misfit level presented a relevant influence on the stresses for all of the structures evaluated.
引用
收藏
页码:1243 / 1249
页数:7
相关论文
共 44 条
[1]   Influence of Material of Overdenture-Retaining Bar with Vertical Misfit on Three-Dimensional Stress Distribution [J].
Abreu, Ricardo T. ;
Spazzin, Aloisio O. ;
Noritomi, Pedro Y. ;
Consani, Rafael Leonardo X. ;
Mesquita, Marcelo F. .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2010, 19 (06) :425-431
[2]  
Al-Turki LEE, 2002, INT J PROSTHODONT, V15, P38
[3]   Effect of Superstructure Materials and Misfit on Stress Distribution in a Single Implant-Supported Prosthesis: A Finite Element Analysis [J].
Assuncao, Wirley Goncalves ;
Gomes, Erica Alves ;
Ricardo Barao, Valentim Adelino ;
Delben, Juliana Aparecida ;
Tabata, Lucas Fernando ;
Capello de Sousa, Edson Antonio .
JOURNAL OF CRANIOFACIAL SURGERY, 2010, 21 (03) :689-695
[4]   Influence of different mucosal resiliency and denture reline on stress distribution in peri-implant bone tissue during osseointegration. A three-dimensional finite element analysis [J].
Bacchi, Atais ;
Consani, Rafael L. X. ;
Mesquita, Marcelo F. ;
dos Santos, Mateus B. F. .
GERODONTOLOGY, 2012, 29 (02) :E833-E837
[5]   OSSEOINTEGRATION AND ITS EXPERIMENTAL BACKGROUND [J].
BRANEMARK, PI .
JOURNAL OF PROSTHETIC DENTISTRY, 1983, 50 (03) :399-410
[6]  
Brunski JB, 1994, BIOL MECH TOOTH ERUP, P667
[7]   The response of bone in primates around unloaded dental implants supporting prostheses with different levels of fit [J].
Carr, AB ;
Gerard, DA ;
Larsen, PE .
JOURNAL OF PROSTHETIC DENTISTRY, 1996, 76 (05) :500-509
[8]   The effect of commercially pure titannium and alternative dental alloys on the marginal fit of one-piece cast implant frameworks [J].
de Torres, Erica Miranda ;
Silveira Rodrigues, Renata Cristina ;
Chiarello de Mattos, Maria da Gloria ;
Ribeiro, Ricardo Faria .
JOURNAL OF DENTISTRY, 2007, 35 (10) :800-805
[9]   Viscoelastic finite element analysis of an all-ceramic fixed partial denture [J].
DeHoff, PH ;
Anusavice, KJ ;
Götzen, N .
JOURNAL OF BIOMECHANICS, 2006, 39 (01) :40-48
[10]   Implant-Bone Interface Stress Distribution in Immediately Loaded Implants of Different Diameters: A Three-Dimensional Finite Element Analysis [J].
Ding, Xi ;
Zhu, Xing-Hao ;
Liao, Sheng-Hui ;
Zhang, Xiu-Hua ;
Chen, Hong .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2009, 18 (05) :393-402