A novel design methodology for Nth order, shielded longitudinal coils for NMR

被引:12
作者
Crozier, S
Dodd, S
Doddrell, DM
机构
[1] Centre for Magnetic Resonance, University of Queensland, St Lucia
关键词
D O I
10.1088/0957-0233/7/1/005
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A new method is presented for the design of nuclear magnetic resonance (NMR) shim coils. The design methodology is the simulated annealing of the conductor positions, where the error function to minimize contains spherical harmonic field errors and other coil properties. This very flexible method allows pure, compact, low-inductance coils to be designed. Active magnetic shielding has been incorporated in the design process, enabling these coils to be used for rapid, pulsed experiments without the generation of deleterious eddy currents. The details for longitudinal coil designs are presented here. A shielded Z(0) coil was built to test the design process and demonstrated good correlation with theoretical predictions.
引用
收藏
页码:36 / 41
页数:6
相关论文
共 22 条
[11]  
GOLAY MJE, 1971, Patent No. 3569823
[12]   ACCURATE SHIM-COIL DESIGN AND MAGNET-FIELD PROFILING BY A POWER-MINIMIZATION-MATRIX METHOD [J].
HOULT, DI ;
DESLAURIERS, R .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1994, 108 (01) :9-20
[13]  
Jackson J.D., 2001, Classical Electrodynmaics, VThird
[14]  
KIRKPATRICK S, 1983, SCIENCE, V220, P673
[15]   LOCALIZED VOLUME SELECTION TECHNIQUE USING AN ADDITIONAL RADIAL GRADIENT COIL [J].
LEE, SY ;
CHO, ZH .
MAGNETIC RESONANCE IN MEDICINE, 1989, 12 (01) :56-63
[16]   ACTIVE MAGNETIC SCREENING OF COILS FOR STATIC AND TIME-DEPENDENT MAGNETIC-FIELD GENERATION IN NMR IMAGING [J].
MANSFIELD, P ;
CHAPMAN, B .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1986, 19 (07) :540-545
[17]   EQUATION OF STATE CALCULATIONS BY FAST COMPUTING MACHINES [J].
METROPOLIS, N ;
ROSENBLUTH, AW ;
ROSENBLUTH, MN ;
TELLER, AH ;
TELLER, E .
JOURNAL OF CHEMICAL PHYSICS, 1953, 21 (06) :1087-1092
[18]   NEW SPATIAL LOCALIZATION METHOD USING PULSED HIGH-ORDER FIELD GRADIENTS (SHOT - SELECTION WITH HIGH-ORDER GRADIENT) [J].
OH, CH ;
HILAL, SK ;
CHO, ZH ;
MUN, IK .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (01) :63-70
[19]   MINIMUM ENERGY MRI GRADIENT COILS OF GENERAL GEOMETRY [J].
PISSANETZKY, S .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (07) :667-673
[20]   MAGNET FIELD PROFILING - ANALYSIS AND CORRECTING COIL DESIGN [J].
ROMEO, F ;
HOULT, DI .