Stress distribution of three NiTi rotary files under bending and torsional conditions using a mathematic analysis

被引:73
作者
Kim, T. O. [1 ]
Cheung, G. S. P. [2 ]
Lee, J. M. [3 ]
Kim, B. M. [3 ]
Hur, B. [1 ]
Kim, H. C. [1 ]
机构
[1] Pusan Natl Univ, Sch Dent, Dept Conservat Dent, Pusan 602739, South Korea
[2] Univ Hong Kong, Area Endodont, Fac Dent, Hong Kong, Hong Kong, Peoples R China
[3] Pusan Natl Univ, Div Precis Mfg Syst, Pusan 602739, South Korea
关键词
bending; cross-sectional geometry; finite-element analysis; NiTi rotary file; stress distribution; torsion; NICKEL-TITANIUM INSTRUMENTS; ROOT-CANAL INSTRUMENTS; STAINLESS-STEEL HAND; ENDODONTIC INSTRUMENTS; CLINICAL-USE; SHAPING ABILITY; PROFILE SYSTEMS; CROSS-SECTIONS; CURVED CANALS; ENGINE-DRIVEN;
D O I
10.1111/j.1365-2591.2008.01481.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
To compare and evaluate the stress distribution of three NiTi instruments of various cross-sectional configurations under bending or torsional condition using a finite-element analysis model. Three NiTi files (ProFile, ProTaper and ProTaper Universal) were scanned using Micro-CT to produce a three-dimensional digital model. The behaviour of the instrument under bending or torsional loads was analysed mathematically in software (ABAQUS V6.5-1), taking into consideration the nonlinear mechanical characteristic of NiTi material. ProFile showed the greatest flexibility, followed by ProTaper Universal and ProTaper. The highest stress was observed at the surface near the cutting edge and the base of (opposing) flutes during cantilever bending. Concentration of stresses was observed at the bottom of the flutes in ProFile and ProTaper Universal instruments in torsion. The stress was more evenly distributed over the surface of ProTaper initially, which then concentrated at the middle of the convex sides when the amount of angular deflection was increased. Incorporating a U-shaped groove in the middle of each side of the convex-triangular design lowers the flexural rigidity of the origin ProTaper design. Bending leads to the highest surface stress at or near the cutting edge of the instrument. Stress concentration occurs at the bottom of the flute when the instrument is subjected to torsion.
引用
收藏
页码:14 / 21
页数:8
相关论文
共 38 条
[1]   K3 endo, ProTaper, and ProFile systems: Breakage and distortion, in severely curved roots of molars [J].
Ankrum, MT ;
Hartwell, GR ;
Truitt, JE .
JOURNAL OF ENDODONTICS, 2004, 30 (04) :234-237
[2]   Evaluation of single-use rotary nickel-titanium instruments [J].
Arens, FC ;
Hoen, MM ;
Steiman, HR ;
Dietz, GC .
JOURNAL OF ENDODONTICS, 2003, 29 (10) :664-666
[3]   Progressive versus constant tapered shaft design using NiTi rotary instruments [J].
Bergmans, L ;
Van Cleynenbreugel, J ;
Beullens, M ;
Wevers, M ;
Van Meerbeek, B ;
Lambrechts, P .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (04) :288-295
[4]   Comparative analysis of torsional and bending stresses in two mathematical models of nickel-titanium rotary instruments: ProTaper versus ProFile [J].
Berutti, E ;
Chiandussi, G ;
Gaviglio, I ;
Ibba, A .
JOURNAL OF ENDODONTICS, 2003, 29 (01) :15-19
[5]   Impact of lubricant parameters on rotary instrument torque and force [J].
Boessler, Claudia ;
Peters, Ove A. ;
Zehnder, Matthias .
JOURNAL OF ENDODONTICS, 2007, 33 (03) :280-283
[6]   Shaping ability of ProTaper nickel-titanium files in simulated resin root canals [J].
Calberson, FLG ;
Deroose, CAJG ;
Hommez, GMG ;
De Moor, RJG .
INTERNATIONAL ENDODONTIC JOURNAL, 2004, 37 (09) :613-623
[7]  
Camps JJ, 1995, ENDOD DENT TRAUMATOL, V11, P270
[8]   A comparison of stainless steel hand and rotary nickel-titanium instrumentation using a silicone impression technique [J].
Chen, JL ;
Messer, HH .
AUSTRALIAN DENTAL JOURNAL, 2002, 47 (01) :12-20
[9]   Defects in ProTaper S1 instruments after clinical use: fractographic examination [J].
Cheung, GSP ;
Peng, B ;
Bian, Z ;
Shen, Y ;
Darvell, BW .
INTERNATIONAL ENDODONTIC JOURNAL, 2005, 38 (11) :802-809
[10]  
Clauder Thomas, 2004, Dent Clin North Am, V48, P87, DOI 10.1016/j.cden.2003.10.006