Dental Implants with Conical Implant-Abutment Interface: Influence of the Conical Angle Difference on the Mechanical Behavior of the Implant

被引:22
作者
Aguirrebeitia, Josu [1 ]
Abasolo, Mikel [1 ]
Vallejo, Javier [1 ]
Ansola, Ruben [1 ]
机构
[1] Univ Basque Country, ETSI BILBAO, Dept Mech Engn, Bilbao 48013, Spain
关键词
conical angle difference; conical implant; finite element model; sensitivity study; FINITE-ELEMENT-ANALYSIS; IN-VITRO EVALUATION; BACTERIAL LEAKAGE; PROSTHESIS; COMPONENTS; SYSTEM; DOGS; FIT;
D O I
10.11607/jomi.2775
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Misfit in the conical implant-abutment interface plays an important role on the mechanical behavior of the implant when masticatory forces are applied. The origin of the misfit adopted in this work is a conical angle difference between implant and abutment, which can be due to a combination of design decisions and manufacturing tolerances. The goal of this work was to investigate the effects of the implant-abutment conical angle difference in the following mechanical features: interfacial microgap, preload loss on the bolt, stress level in the bone, and abutment removal force and/or torque. Materials and Methods: A simplified three-dimensional nonlinear monoparametric finite element model of an OsseoSpeed TX 4.5 S 9-mm implant (Astra Tech) with a tapered implant-abutment interface was built to evaluate the variability of the mechanical features cited above with the conical angle difference, keeping constant the overall geometry, load and boundary conditions, material properties, frictional behavior, and mesh structure. Results: As the conical angle difference increased, the following effects were observed: the microgap decreased and remained almost constant for values over a given positive angle difference, the stress level in the bone increased sensitively, the removal force and/or torque needed to separate the abutment from the implant varied slightly, and the bolt preload loss increased. Conclusions: In light of the results provided, the conical angle difference in the implant-abutment interface had a significant influence on the overall mechanical behavior of the implant. Among the four mechanical features considered, the interfacial microgap and the bone stress were demonstrated to be the most sensitive to the conical angle difference, and therefore the most relevant when selecting an optimum value in the design process of a conical interface.
引用
收藏
页码:E72 / E82
页数:11
相关论文
共 50 条
  • [21] Compatibility of Nonoriginal Abutments With Implants: Evaluation of Microgap at the Implant-Abutment Interface, With Original and Nonoriginal Abutments
    Duraisamy, Revathi
    Krishnan, Chitra Shankar
    Ramasubramanian, Hariharan
    Sampathkumar, Jayakrishnakumar
    Mariappan, Saravanakumar
    Sivaprakasam, Azhagarasan Navarasampatti
    IMPLANT DENTISTRY, 2019, 28 (03) : 289 - 295
  • [22] Bacterial Colonization of the Implant-Abutment Interface (IAI) of Dental Implants with a Sloped Marginal Design: An in-vitro Study
    Koutouzis, Theofilos
    Gadalla, Hana
    Lundgren, Tord
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2016, 18 (01) : 161 - 167
  • [23] Influence of Diamondlike Carbon-Coated Screws on the Implant-Abutment Interface
    Vanoni Diez, Juliana Socas
    Brigagao, Vinicius Carvalho
    Cunha, Leonardo Goncalves
    Claro Neves, Ana Christina
    da Silva-Concilio, Lais Regiane
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (05) : 1055 - 1060
  • [24] Effect of double conical abutment screw on implant preload
    Arnetzl, Gerwin Vincent
    Payer, Michael
    Falkensammer, Frank
    Arnetzl, Gerwin
    CLINICAL ORAL IMPLANTS RESEARCH, 2016, 27 (05) : 553 - 557
  • [25] Effect of initial placement level and wall thickness on maintenance of the marginal bone level in implants with a conical implant-abutment interface: a 5-year retrospective study
    Yoo, Jaehyun
    Moon, Ik-Sang
    Yun, Jeong-Ho
    Chung, Chooryung
    Huh, Jong-Ki
    Lee, Dong-Won
    JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2019, 49 (03) : 185 - 192
  • [26] BACTERIAL COLONIZATION OF THE IMPLANT-ABUTMENT INTERFACE OF CONICAL CONNECTION WITH AN INTERNAL OCTAGON: AN IN VITRO STUDY USING REAL-TIME PCR
    Baj, A.
    Beltramini, G. A.
    Bolzoni, A.
    Cura, F.
    Palmieri, A.
    Scarano, A.
    Ottria, L.
    Gianni, A. B.
    JOURNAL OF BIOLOGICAL REGULATORS AND HOMEOSTATIC AGENTS, 2017, 31 (02) : 163 - 168
  • [27] Bacterial Leakage at Implant-Abutment Interface With Different Intermediate Materials
    Mohammadi, Farnoush
    Hajmousaei, Maryam
    Vaziri, Nastaran
    Arshad, Mahnaz
    JOURNAL OF ORAL IMPLANTOLOGY, 2019, 45 (06) : 451 - 455
  • [28] The influence of prosthetic crown height and implant-abutment connection design selection on the long-term implant-abutment stability: A laboratory study
    Bagegni, Aimen
    Spies, Benedikt C.
    Kern, Matthias
    Hazard, Derek
    Kohal, Ralf
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 113
  • [29] Effect of the implant-abutment interface on peri-implant tissues: A systematic review
    Bishti, Shaza
    Strub, Joerg R.
    Att, Wael
    ACTA ODONTOLOGICA SCANDINAVICA, 2014, 72 (01) : 13 - 25
  • [30] Mechanical Resistance of a 2.9-mm-Diameter Dental Implant With a Morse-Taper Implant-Abutment Connection
    Alberti, Alice
    Corbella, Stefano
    Francetti, Luca
    JOURNAL OF ORAL IMPLANTOLOGY, 2023, 49 (03) : 323 - 329