Electrochemical sensor for enzymatic lactate detection based on laser-scribed graphitic carbon modified with platinum, chitosan and lactate oxidase

被引:38
作者
Madden, Julia [1 ]
Vaughan, Eoghan [1 ]
Thompson, Michael [2 ]
O' Riordan, Alan [1 ]
Galvin, Paul [1 ]
Iacopino, Daniela [1 ]
Teixeira, Sofia Rodrigues [1 ]
机构
[1] Univ Coll Cork, Tyndall Natl Inst, Cork, Ireland
[2] Univ Toronto, Dept Chem, Toronto, ON, Canada
基金
爱尔兰科学基金会; 欧盟地平线“2020”;
关键词
Laser scribed graphitic carbon; Electrochemical biosensors; Lactate oxidase; Lactate sensing; Bendable substrates; CYCLIC VOLTAMMETRY; BLOOD LACTATE; ELECTRODES; BIOSENSOR;
D O I
10.1016/j.talanta.2022.123492
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We developed a flexible laser scribed graphitic carbon based lactate biosensor fabricated using a low cost 450 nm laser. We demonstrated a facile fabrication method involving electrodeposition of platinum followed by two casting steps for modification with chitosan and lactate oxidase. The biosensor demonstrated chronoamperometric lactate detection within a linear range from 0.2 mM to 3 mM, (R-2 > 0.99), with a limit of detection of 0.11 mM and a sensitivity of 35.8 mu A/mM/cm(2). The biosensor was successful in performing up to 10 consecutive measurements (one after the other) indicating good working stability (RSD <5%). Concerning storage stability, there was no decrease in signal response after 30 days of storage at 4 degrees C. Additionally, we demonstrate enzymatic lactate detection whilst the flexible polyimide substrates were fixed at a curvature (K) of 0.14 mm 1. No noticeable change in signal response was observed in comparison to calibrations obtained at a curvature of 0 mm 1, signifying potential opportunities for sensor attachment or integration with oral-care products such as mouth swabs. Both laser scribed graphitic carbon and Ag/AgCl modified-laser scribed graphitic carbon were successful as reference electrodes for chronoamperometric lactate measurements. Furthermore, using a three-electrode configuration on polyimide, lactate detection in both artificial saliva and sterile human serum samples was achieved for two spiked concentrations (0.5 mM and 1 mM).
引用
收藏
页数:8
相关论文
共 39 条
[11]   Substrate-Independent Laser-Induced Graphene Electrodes for Microfluidic Electroanalytical Systems [J].
Griesche, Christian ;
Hoecherl, Kilian ;
Baeumner, Antje J. .
ACS APPLIED NANO MATERIALS, 2021, 4 (03) :3114-3121
[12]   Accessing analytes in biofluids for peripheral biochemical monitoring [J].
Heikenfeld, Jason ;
Jajack, Andrew ;
Feldman, Benjamin ;
Granger, Steve W. ;
Gaitonde, Supriya ;
Begtrup, Gavi ;
Katchman, Benjamin A. .
NATURE BIOTECHNOLOGY, 2019, 37 (04) :407-419
[13]   Synthesis, characterization and non-enzymatic lactate sensing performance investigation of mesoporous copper oxide (CuO) using inverse micelle method [J].
Heo, Sung Gue ;
Yang, Won-Sik ;
Kim, Sungjin ;
Park, Young Min ;
Park, Kyoung-Tae ;
Oh, Soong Ju ;
Seo, Seok-Jun .
APPLIED SURFACE SCIENCE, 2021, 555
[14]   An Inkjet-Printed Flexible Non-Enzymatic Lactate Sensor for Clinical Blood Plasma Test [J].
Huang, Yi-Syuan ;
Chen, Kuan-Yu ;
Cheng, Yu-Ting ;
Lee, Chih-Kuo ;
Tsai, Hsiao-En .
IEEE ELECTRON DEVICE LETTERS, 2020, 41 (04) :597-600
[15]   Non-enzymatic electrochemical lactate sensing by NiO and Ni(OH)2 electrodes: A mechanistic investigation [J].
Kim, Sungjin ;
Kim, Kijun ;
Kim, Hyun-Jong ;
Lee, Ho-Nyun ;
Park, Tae Joo ;
Park, Young Min .
ELECTROCHIMICA ACTA, 2018, 276 :240-246
[16]   ELECTRON-TRANSFER KINETICS FROM CYCLIC VOLTAMMETRY - QUANTITATIVE DESCRIPTION OF ELECTROCHEMICAL REVERSIBILITY [J].
KLINGLER, RJ ;
KOCHI, JK .
JOURNAL OF PHYSICAL CHEMISTRY, 1981, 85 (12) :1731-1741
[17]   An extended method for the practical evaluation of the standard rate constant from cyclic voltammetric data [J].
Lavagnini, I ;
Antiochia, R ;
Magno, F .
ELECTROANALYSIS, 2004, 16 (06) :505-506
[18]   Flex-GO (Flexible graphene oxide) sensor for electrochemical monitoring lactate in low-volume passive perspired human sweat [J].
Lin, Kai-Chun ;
Muthukumar, Sriram ;
Prasad, Shalini .
TALANTA, 2020, 214
[19]   Laser-Induced Graphene-Based Platforms for Dual Biorecognition of Molecules [J].
Marques, Ana C. ;
Cardoso, Ana R. ;
Martins, Rodrigo ;
Sales, M. Goreti F. ;
Fortunato, Elvira .
ACS APPLIED NANO MATERIALS, 2020, 3 (03) :2795-2803
[20]  
Miadlikowska Jolanta, 2015, ACS APPL MAT AMP INT, V79, P132, DOI [10.1016/j.ympev.2014.04.003, DOI 10.1021/ACSAMI.5B06070]