High-Precision Wavelength Calibration of Wide-Band Monochromator

被引:2
作者
Zhang Zi-hui [1 ,2 ]
Wang Shu-rong [1 ]
Huang Yu [1 ]
Li Bo [1 ]
Yu Xiang-yang [1 ,2 ]
Lin Guan-yu [1 ]
机构
[1] Chinese Acad Sci, Changchun Inst Opt Fine Mech & Phys, State Key Lab Appl Opt, Changchun 130033, Peoples R China
[2] Chinese Acad Sci, Grad Univ, Beijing 100049, Peoples R China
关键词
Spectroscopy; Wavelength calibration; Peak search; Multi-level spectrum; Gaussian fit;
D O I
10.3964/j.issn.1000-0593(2012)10-2870-05
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Usually the monochromator is used to output monochromatic light to calibrate the space remote sensing spectrometer. In the present paper, the confidence of space remote sensing spectrometer is used as a standard to evaluate the precision of wide-band monochromator wavelength calibration. Through analysis of the accuracy of the instrument, the monochromator wavelength repeatability error and deviations was obtained respectively. And the intrinsic spectrum of the high pressure mercury lamp and the grating diffraction was used as calibration lines to avoid the error caused by replacing the light source. Through the special method of wavelength calibration to shorten the scan time, the Gaussian fitting was used to look for peaks of wavelength to reduce error. Finally, the relationship derived from polynomial fitting to measure the exact wavelengths' accuracy of the monochromator and calculate the calibration confidence of the space remote sensing spectrometer. Using this method, we can make wavelength accuracy of the 1. 5 M monochromator with wavelength band from 200 to 840 nm to reach to 0. 016 rim, then the confidence of the space remote sensing spectrometer can reach to 99. 82%.
引用
收藏
页码:2870 / 2874
页数:5
相关论文
共 11 条
[1]   High Precision All-Reflection Fourier Transform Imaging Spectrometer Spectral Calibration Using Homogeneous Broadening of the Wave Number Model [J].
Cui De-qi ;
Liao Ning-fang ;
Cao Wei-liang ;
Tan Bo-neng ;
Tian Li-xun .
SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31 (07) :1762-1766
[2]  
DONG Ning-ning, 2008, OPTICS PRECISION ENG, V16, P1661
[3]  
JIN Li-hong, 1998, J SICHUN U NATURAL S, V35, P546
[4]  
[李新 Li Xin], 2008, [光学学报, Acta Optica Sinica], V28, P902
[5]  
Lin Guanyu, 2010, Chinese Journal of Scientific Instrument, V31, P2668
[6]  
MA Hong, 2007, THEORY ERROR PRECISS, P45
[7]  
Quan-chen LI, 1999, PRINCIPLE SPECTROGRA, P209
[8]  
Shen H. C., 2006, OPTICAL INSTRUMENTS, V28, P51
[9]  
Wang Jian-wen, 2008, Shengwu Jiagong Guocheng, V6, P6
[10]  
Yang J., 2010, ELECT MEASUREMENT TE, V33, P78