Effects of stimulus configuration on psychophysical operating levels and on speech recognition with cochlear implants

被引:66
作者
Pfingst, BE
Zwolan, TA
Holloway, LA
机构
[1] Kresge Hearing Research Institute, Department of Otolaryngology, Univ. of Michigan Medical Center, Ann Arbor, MI 48109-0506
关键词
cochlear implant; detection threshold; electrode configuration; electrical stimulation; human; psychophysics; pulse duration; speech recognition;
D O I
10.1016/S0378-5955(97)00122-6
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Effects of electrode configuration and pulse duration on operating levels' and on speech recognition performance were studied in a group of 14 adult postlingually deaf human subjects with Nucleus cochlear implants. The operating levels (based on detection threshold and maximum comfortable loudness levels) for narrowly spaced bipolar (BP) stimulation were found to be about Il dB higher on average than those for widely spaced bipolar (BP+6) or monopolar (MP1) stimulation. Operating levels for common ground (CG) stimulation fell between those for BP and BP+6; the difference between BP and CG detection thresholds depended on pulse duration. Variation in detection thresholds and maximum comfortable loudness levels across the electrode array (electrodes 1-15) was larger for BP and CG stimulation than for BP+6 or MPI stimulation, suggesting narrower spread of activation for the BP and CG configurations despite the higher current levels. Speech recognition performance was tested using experimental processor configurations. Among the experimental electrode configurations tested (BP, CG, and BP+6), the highest speech recognition scores were obtained with the BP+6 configuration in many subjects. Effects of pulse duration on speech recognition were less consistent and usually smaller than the effects of electrode configuration. The results indicate that electrode configuration is an important variable determining speech recognition performance and suggest that restriction of the size of neural population activated by individual channels of the prosthesis is not necessarily advantageous.
引用
收藏
页码:247 / 260
页数:14
相关论文
共 32 条
[1]  
[Anonymous], 1983, Iowa cochlear implant tests
[2]  
[Anonymous], [No title captured], Patent No. [4,532,930, 4532930]
[3]  
Clark G M, 1988, Acta Otolaryngol Suppl, V448, P1
[4]  
CLARK GM, 1987, ADV OTORHINOLARYNGOL, V38, P1
[5]   A MODEL OF ELECTRICAL EXCITATION OF THE MAMMALIAN AUDITORY-NERVE NEURON [J].
COLOMBO, J ;
PARKINS, CW .
HEARING RESEARCH, 1987, 31 (03) :287-311
[6]  
GANTZ BJ, 1988, LARYNGOSCOPE, V98, P1100
[7]  
KENNEDY DW, 1987, LARYNGOSCOPE, V97, P42
[8]  
KILENY PR, 1992, AM J OTOL, V13, P117
[9]  
LEHNHARDT E, 1992, TRANSPLANTS IMPLANTS, P283
[10]   SOME CONSIDERATIONS OF MULTICHANNEL ELECTRICAL-STIMULATION OF THE AUDITORY-NERVE IN THE PROFOUNDLY DEAF - INTERFACING ELECTRODE ARRAYS WITH THE AUDITORY-NERVE ARRAY [J].
MERZENICH, MM ;
WHITE, M ;
VIVION, MC ;
LEAKEJONES, PA ;
WALSH, S .
ACTA OTO-LARYNGOLOGICA, 1979, 87 (3-4) :196-203