Stenosis of the main artery supplying an organ: Effect of end-organ vascular resistance on the poststenotic peak systolic velocity in an in vitro hydraulic model at Doppler US

被引:10
作者
Bude, RO
Larson, RG
Nichols, WW
Rubin, JM
机构
[1] Univ Michigan, Med Ctr, Dept Radiol, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Chem Engn, Ann Arbor, MI 48109 USA
[3] Univ Florida, Dept Med, Div Cardiol, Gainesville, FL USA
关键词
blood; flow dynamics; renal arteries; stenosis or obstruction; US; test objects;
D O I
10.1148/radiology.212.1.r99jl2879
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To test the hypothesis that increased end-organ vascular resistance reduces blood flow to the kidney, thus reducing the mean velocity in the renal artery and secondarily lowering the peak systolic velocity (PSV). MATERIALS AND METHODS: An in vitro hydraulic model with a pulsatile pump, blood-mimicking fluid, interchangeable stenoses, and variable compliance and resistance was used to investigate the relationship between end-organ vascular resistance and poststenotic PSV. RESULTS: Poststenotic PSV was mildly dependent on end-organ vascular resistance and decreased with increasing resistance. CONCLUSION: The results help explain some of the reported variability from using poststenotic PSV to detect hemodynamically significant renal arterial stenoses, but the effect is not great enough to completely explain the variability. Other factors not investigated in this study must be at work as well.
引用
收藏
页码:79 / 87
页数:9
相关论文
共 32 条
  • [1] NONINVASIVE DIAGNOSIS OF RENAL-ARTERY STENOSIS BY ECHO-DOPPLER VELOCIMETRY
    AVASTHI, PS
    VOYLES, WF
    GREENE, ER
    [J]. KIDNEY INTERNATIONAL, 1984, 25 (05) : 824 - 829
  • [2] BAXTER GM, 1995, CLIN RADIOL, V50, P618, DOI 10.1016/S0009-9260(05)83291-X
  • [3] RENAL-ARTERY STENOSIS - PROSPECTIVE EVALUATION OF DIAGNOSIS WITH COLOR DUPLEX US COMPARED WITH ANGIOGRAPHY - WORK IN PROGRESS
    BERLAND, LL
    KOSLIN, DB
    ROUTH, WD
    KELLER, FS
    [J]. RADIOLOGY, 1990, 174 (02) : 421 - 423
  • [4] PULSUS TARDUS - ITS CAUSE AND POTENTIAL LIMITATIONS IN DETECTION OF ARTERIAL-STENOSIS
    BUDE, RO
    RUBIN, JM
    PLATT, JF
    FECHNER, KP
    ADLER, RS
    [J]. RADIOLOGY, 1994, 190 (03) : 779 - 784
  • [5] RENAL-ARTERY STENOSIS - EVALUATION WITH COLOR DOPPLER FLOW IMAGING
    DESBERG, AL
    PAUSHTER, DM
    LAMMERT, GK
    HALE, JC
    TROY, RB
    NOVICK, AC
    NALLY, JV
    WELTEVREDEN, AM
    [J]. RADIOLOGY, 1990, 177 (03) : 749 - 753
  • [6] DUCK FA, 1990, PHYSICAL PROPERTIES, P161
  • [7] POSTTRANSPLANT RENAL-ARTERY STENOSIS - OUTPATIENT INTRAARTERIAL DSA VERSUS COLOR AIDED DUPLEX DOPPLER SONOGRAPHY
    DUDA, SH
    ERLEY, CM
    WAKAT, JP
    HUPPERT, PE
    LAUCHART, W
    RISLER, T
    CLAUSSEN, CD
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 1993, 16 (02) : 95 - 101
  • [8] FOLKOW B, 1971, CIRCULATION, P10
  • [9] DIAGNOSIS OF RENAL-ARTERY STENOSIS IN TRANSPLANTED KIDNEYS - VALUE OF DOPPLER WAVE-FORM ANALYSIS OF THE INTRARENAL ARTERIES
    GOTTLIEB, RH
    LIEBERMAN, JL
    PABICO, RC
    WALDMAN, DL
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (06) : 1441 - 1446
  • [10] DYNAMIC CORONARY STENOSIS
    GOULD, KL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1980, 45 (02) : 286 - 292