Wear of 36-mm BIOLOX® delta ceramic-on-ceramic bearing in total hip replacements under edge loading conditions

被引:56
作者
Al-Hajjar, Mazen [1 ]
Fisher, John [1 ,2 ]
Tipper, Joanne L. [1 ,2 ]
Williams, Sophie [1 ]
Jennings, Louise M. [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Leeds Teaching Hosp NHS Trust, Leeds Musculoskeletal Biomed Res Unit, Leeds LS2 9JT, W Yorkshire, England
基金
英国生物技术与生命科学研究理事会; 英国工程与自然科学研究理事会; 英国惠康基金;
关键词
Edge loading; malpositioning; ceramic-on-ceramic; hip replacement; microseparation; cup inclination angle; LONG-TERM WEAR; CUP ANGLE; ACETABULAR CUPS; JOINT SIMULATOR; FEMORAL HEADS; IN-VITRO; ALUMINA; ARTHROPLASTY; MICROSEPARATION; POSITION;
D O I
10.1177/0954411912474613
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ceramic-on-ceramic bearings have become of great interest due to the substantial improvements in the manufacturing techniques and material properties and due to polyethylene wear debris-induced osteolysis and the issues with metal wear debris and ion release by metal-on-metal bearings. Edge loading conditions due to translational malpositioning (microseparation conditions) have been shown to replicate clinically relevant wear mechanisms and increase the wear of ceramic-on-ceramic bearings; thus, it was necessary to test new bearing materials and designs under these adverse conditions. The aim of this study was to assess the effect of increasing head size on the wear of BIOLOX(R) delta ceramic-on-ceramic bearings under edge loading conditions due to rotational (steep cup inclination angle) and translational (microseparation) malpositioning. In this study, six 36-mm ceramic-on-ceramic bearings (BIOLOX delta, CeramTec, Germany) were tested under standard and edge loading conditions using the Leeds II hip simulator and compared to the 28-mm bearings tested and published previously under identical conditions. The mean wear rate under standard gait conditions was below 0.1 mm(3)/million cycles for both the 28-mm and the 36-mm ceramic-on-ceramic bearings, and increasing the inclination angle did not affect the wear rates. The introduction of microseparation to the gait cycle increased the wear rate of ceramic-on-ceramic bearing and resulted in stripe wear on the femoral heads. Under microseparation conditions, the wear rate of size 36-mm bearings (0.22 mm(3)/million cycles) was significantly higher (p = 0.004) than that for size 28-mm bearings (0.13 mm(3)/million cycles). This was due to the larger contact area for the larger bearings and deprived lubrication under edge loading conditions. The wear rate of BIOLOX delta ceramic-on-ceramic bearings under microseparation conditions was still very low (< 0.25 mm(3)/million cycles) compared to earlier generation ceramic-on-ceramic bearings (BIOLOX forte, 1.84 mm(3)/million cycles) and other bearing materials such as metal-on-metal bearings (2-8 mm(3)/million cycles).
引用
收藏
页码:535 / 542
页数:8
相关论文
共 37 条
[1]   Effect of cup inclination angle during microseparation and rim loading on the wear of BIOLOX® delta ceramic-on-ceramic total hip replacement [J].
Al-Hajjar, Mazen ;
Leslie, Ian J. ;
Tipper, Joanne ;
Williams, Sophie ;
Fisher, John ;
Jennings, Louise M. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2010, 95B (02) :263-268
[2]   Influence of cup orientation on the wear performance of metal-on-metal hip replacements [J].
Angadji, A. ;
Royle, M. ;
Collins, S. N. ;
Shelton, J. C. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H4) :449-457
[3]   Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant [J].
Bigsby, RJA ;
Hardaker, CS ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (03) :265-269
[4]  
BOUTIN P, 1971, PRESSE MED, V79, P639
[5]  
Bowsher JG, 2005, LARGE HEAD DIAMETERS, P1626
[6]   High strength and toughness alumina matrix composites by transformation toughening and 'in situ' platelet reinforcement (ZPTA) - The new generation of bioceramics [J].
Burger, W ;
Richter, HG .
BIOCERAMICS, 2000, 192-1 :545-548
[7]   Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty [J].
Chevillotte, Christophe ;
Pibarot, Vincent ;
Carret, Jean-Paul ;
Bejui-Hugues, Jacques ;
Guyen, Olivier .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (11) :1599-1604
[8]   Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements [J].
Clarke, IC ;
Good, V ;
Williams, P ;
Schroeder, D ;
Anissian, L ;
Stark, A ;
Oonishi, H ;
Schuldies, J ;
Gustafson, G .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H4) :331-347
[9]  
Fritsch EW, 1996, CLIN ORTHOP RELAT R, P129
[10]   Wear and biological activity of highly crosslinked polyethylene in the hip under low serum protein concentrations [J].
Galvin, A. L. ;
Tipper, I. L. ;
Jennings, L. M. ;
Stone, M. H. ;
Jin, Z. M. ;
Ingham, E. ;
Fisher, J. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H1) :1-10