Performance of a single ultrafine particle mass spectrometer

被引:42
作者
Phares, DJ [1 ]
Rhoads, KP [1 ]
Wexler, AS [1 ]
机构
[1] Univ Delaware, Dept Mech Engn, Newark, DE 19716 USA
关键词
D O I
10.1080/02786820252883829
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
We report on the performance of a rapid single particle mass spectrometer (RSMS-II), designed to obtain the size and composition of individual ultrafine particles. Particles are sized aerodynamically at the inlet using a dynamic focusing mechanism to transmit particles to the source region of a time-of-flight mass spectrometer. Since the target particles are too small to be detected optically, an excimer laser is pulsed at high frequency so that data are acquired only when a particle coincides with a laser pulse within the source region. The instrument is tested with sodium chloride and oleic acid particle standards of various sizes and the hit rate efficiency is monitored as the normalized number of particle hits per unit time. The hit rate efficiency depends on the particle flux through the active region of the laser beam, in addition to the particle size and composition, and may thus be used to determine the relative transmission efficiency and size selectivity of the inlet.
引用
收藏
页码:583 / 592
页数:10
相关论文
共 14 条