Influence of cage design on interbody fusion in a sheep cervical spine model

被引:60
|
作者
Kandziora, F [1 ]
Schollmeier, G [1 ]
Scholz, M [1 ]
Schaefer, J [1 ]
Scholz, A [1 ]
Schmidmaier, G [1 ]
Schröder, R [1 ]
Bail, H [1 ]
Duda, G [1 ]
Mittlmeier, T [1 ]
Haas, NP [1 ]
机构
[1] Humboldt Univ, Klinikum Charite, Strahlenklin & Poliklin, D-13353 Berlin, Germany
关键词
cervical spine; interbody cage; spinal fusion; sheep;
D O I
10.3171/spi.2002.96.3.0321
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. The purpose of this study was to compare the characteristics of interbody fusion achieved using an autologous tricortical iliac crest bone graft with those of a cylinder- and a box-design cage in a sheep cervical spine model. This study was designed to determine whether there are differences between three interbody fusion procedures in: 1) ability to preserve postoperative distraction., 2) biomechanical stability, and 3) histological characteristics of intervertebral bone matrix formation. Methods. Twenty-four sheep underwent C3-4 discectomy and fusion in which the following were used: Group 1, autologous tricortical iliac crest bone graft (eight sheep); Group 2, titanium cylinder-design cage filled with autologous iliac crest bone graft (eight sheep); and Group 3, titanium box-design cage filled with autologous iliac crest graft (eight sheep). Radiography was performed pre- and postoperatively and after 1, 2, 4, 8, and 12 weeks. At the same time points, disc space height, intervertebral angle, and lordosis angle were measured. After 12 weeks. the sheep were killed, and fusion sites were evaluated by obtaining functional radiographs in flexion and extension, Quantitative computerized tomography scans were acquired to assess bone mineral density, bone mineral content, and bone callus volume. Biomechanical testing was performed in flexion, extension, axial rotation, and lateral bending. Stiffness, range of motion, neutral zone, and elastic zone were determined. Histomorphological and histomorphometric analyses were performed, and polychrome sequential labeling was used to determine the time frame of new bone formation. Over a 12-week period significantly higher values for disc space height and intervertebral angle were shown in cage-treated sheep than in those that received bone graft. Functional radiographic assessment revealed significantly lower residual flexion-extens ion movement in sheep with the cylinder cage-fixed spines than in those that received bone graft group. The cylinder-design cages showed significantly higher values for bone mineral content, bone callus content, and stiffness in axial rotation and lateral bending than the other cages or grafts. Histomorphometric evaluation and polychrome sequential labeling showed a more progressed bone matrix formation in the cylindrical cage group than in both other groups. Conclusions. Compared with the tricortical bone graft, both cages showed significantly better distractive properties. The cylindrical cage demonstrated a significantly higher biomechanical stiffness and an accelerated interbody fusion compared with the box-design cage and the tricortical bone graft, The differences in bone matrix formation within both cages were the result of the significantly lower stress shielding on the bone graft by the cylinder-design cage.
引用
收藏
页码:321 / 332
页数:12
相关论文
共 50 条
  • [1] Experimental fusion of the sheep cervical spine.: Part I:: Effect of cage design on interbody fusion
    Kandziora, F
    Pflugmacher, R
    Scholz, M
    Schäfer, J
    Schollmeier, G
    Schnake, KJ
    Bail, H
    Duda, G
    Haas, NP
    CHIRURG, 2002, 73 (09): : 909 - 917
  • [2] Bioabsorbable interbody cages in a sheep cervical spine fusion model
    Kandziora, F
    Pflugmacher, R
    Scholz, M
    Eindorf, T
    Schnake, KJ
    Haas, NP
    SPINE, 2004, 29 (17) : 1845 - 1855
  • [3] Experimental fusion of the sheep cervical spine. Part I: Effect of cage design on interbody fusion [Experimentelle Spondylodese der Schafshalswirbelsäule: Teil 1: Der Effekt des Cage-Designs auf die intervertebrale Fusion]
    Kandziora F.
    Pflugmacher R.
    Scholz M.
    Schäfer J.
    Schollmeier G.
    Schnake K.J.
    Bail H.
    Duda G.
    Haas N.P.
    Der Chirurg, 2002, 73 (9): : 909 - 917
  • [4] Biomechanical comparison of cervical spine interbody fusion cages
    Kandziora, F
    Pflugmacher, R
    Schäfer, J
    Born, C
    Duda, G
    Haas, NP
    Mittlmeier, T
    SPINE, 2001, 26 (17) : 1850 - 1857
  • [5] Experimental fusion of the sheep cervical spine. Part II: Effect of growth factors and carrier systems on interbody fusion
    Kandziora, F
    Scholz, M
    Pflugmacher, R
    Krummrey, G
    Schollmeier, G
    Schmidmaier, G
    Schnake, KJ
    Duda, G
    Raschke, M
    Haas, NP
    CHIRURG, 2002, 73 (10): : 1025 - +
  • [6] Biomechanical comparison of bioabsorbable cervical spine interbody fusion cages
    Pflugmacher, R
    Schleicher, P
    Gumnior, S
    Turan, O
    Scholz, M
    Eindorf, T
    Haas, NP
    Kandziora, F
    SPINE, 2004, 29 (16) : 1717 - 1722
  • [7] Bioabsorbable Interbody Magnesium-Polymer Cage Degradation Kinetics, Biomechanical Stiffness, and Histological Findings From an Ovine Cervical Spine Fusion Model
    Daentzer, Dorothea
    Willbold, Elmar
    Kalla, Katharina
    Bartsch, Ivonne
    Masalha, Waseem
    Hallbaum, Maximiliane
    Hurschler, Christof
    Kauth, Theresa
    Kaltbeitzel, Daniel
    Hopmann, Christian
    Welke, Bastian
    SPINE, 2014, 39 (20) : E1220 - E1227
  • [8] Biodegradable cage. Osteointegration in spondylodesis of the sheep cervical spine
    Pflugmacher R.
    Eindorf T.
    Scholz M.
    Gumnior S.
    Krall C.
    Schleicher P.
    Haas N.P.
    Kandziora F.
    Der Chirurg, 2004, 75 (10): : 1003 - 1012
  • [9] Biodegradable cage. Osteointegration in spondylodesis of the sheep cervical spine
    Pflugmacher, R
    Eindorf, T
    Scholz, M
    Gumnior, S
    Krall, C
    Schleicher, P
    Haas, NP
    Kandziora, F
    CHIRURG, 2004, 75 (10): : 1003 - 1012
  • [10] Anterior cervical interbody fusion using polyetheretherketone cage filled with synthetic bone graft in acute cervical spine injury
    Hattou, L.
    Morandi, X.
    Lefebvre, J.
    Le Reste, P. -J.
    Riffaud, L.
    Henaux, P. -L.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2017, 103 (01) : 61 - 66