Clinical validation of CT image reconstruction with interior tomography

被引:3
作者
Ge, Gary [1 ]
Zhang, Jie [2 ]
Winkler, Michael [2 ]
Lumby, Cynthia [2 ]
Cong, Wenxiang [3 ]
Wang, Ge [3 ]
机构
[1] Univ Kentucky, Dept Radiat Med, Lexington, KY USA
[2] Univ Kentucky, Dept Radiol, Lexington, KY USA
[3] Rensselaer Polytech Inst, Biomed Imaging Ctr, Troy, NY USA
关键词
Active x-ray collimation; computed tomography (CT) image reconstruction; internal field of view in CT images; interior tomography; TUBE-CURRENT MODULATION; DOSE REDUCTION; BREAST; MDCT;
D O I
10.3233/XST-17329
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Active x-ray collimation is well adopted in radiography and fluoroscopy for radiation dose reduction and image quality improvement. The application of this concept in computed tomography (CT) is significantly limited due to the truncation of projection data. Generally, an internal field of view (FOV) inside an imaging object cannot be exactly reconstructed only from the truncated projection data. Recent research shows that given some prior information of the FOV image, interior tomography can provide a unique and stable solution for image reconstruction of an internal FOV. The objective of this study is to evaluate the performance of interior reconstruction based on patient datasets obtained from a clinical CT scanner with dual x-ray tubes, which simultaneously gives full projections and truncated projections. Image reconstructions are performed from full and truncated projection data for the comparison of image quality, respectively. The reconstructed CT images were reviewed by a radiologist and a resident. The evaluation results of two observers showed that CT images reconstructed with truncated projections met clinically diagnostic requirements and were comparable to clinical images. This study demonstrates that with the development of interior tomography, active x-ray collimation in the imaging plane can be readily employed in CT imaging to further reduce patient radiation and improve image quality.
引用
收藏
页码:303 / 309
页数:7
相关论文
共 25 条
[1]  
[Anonymous], 2009, Computed tomography: principles, design, artifacts, and recent advances
[2]  
Bushberg J.T., 2012, ESSENTIAL PHYS MED I, V3
[3]   Dose Reduction to Anterior Surfaces With Organ-Based Tube-Current Modulation: Evaluation of Performance in a Phantom Study [J].
Duan, Xinhui ;
Wang, Jia ;
Christner, Jodie A. ;
Leng, Shuai ;
Grant, Katharine L. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) :689-695
[4]  
Goldman AR, 2013, AJR, V200, P000
[5]   General Total Variation Minimization Theorem for Compressed Sensing Based Interior Tomography [J].
Han, Weimin ;
Yu, Hengyong ;
Wang, Ge .
INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2009, 2009
[6]   Radiation dose reduction to breast and thyroid during MDCT:: Effectiveness of an in-plane bismuth shield [J].
Hohl, C ;
Wildberger, JE ;
Süss, C ;
Thomas, C ;
Mühlenbruch, G ;
Schmidt, T ;
Honnef, D ;
Günther, RW ;
Mahnken, AH .
ACTA RADIOLOGICA, 2006, 47 (06) :562-567
[7]   Stability of the interior problem with polynomial attenuation in the region of interest [J].
Katsevich, E. ;
Katsevich, A. ;
Wang, G. .
INVERSE PROBLEMS, 2012, 28 (06)
[8]   Ordered-subset Split-Bregman algorithm for interior tomography [J].
Kong, Huihua ;
Liu, Rui ;
Yu, Hengyong .
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2016, 24 (02) :221-240
[9]   Radiation dose modulation techniques in the multidetector CT era: From basics to practice [J].
Lee, Chang Hyun ;
Goo, Jin Mo ;
Lee, Hyun Ju ;
Ye, Sung-Joon ;
Park, Chang Min ;
Chun, Eun Ju ;
Im, Jung-Gi .
RADIOGRAPHICS, 2008, 28 (05) :1451-1459
[10]   Strategies for Reducing Radiation Dose in CT [J].
McCollough, Cynthia H. ;
Primak, Andrew N. ;
Braun, Natalie ;
Kofler, James ;
Yu, Lifeng ;
Christner, Jodie .
RADIOLOGIC CLINICS OF NORTH AMERICA, 2009, 47 (01) :27-+