Response components of LiF:Mg,Ti around a moderated Am-Be neutron source

被引:0
作者
Méndez, R [1 ]
Iñiguez, MP
Barquero, R
Mañanes, A
Gallego, E
Lorente, A
Voytchev, M
机构
[1] Univ Valladolid, Dpto Fis Teor & Atom Nucl & Mol, E-47002 Valladolid, Spain
[2] Hosp Rio Hortega, Serv Protecc Radiol, Valladolid, Spain
[3] Univ Cantabria, Dept Fis Moderna, E-39005 Santander, Spain
[4] Univ Politecn Madrid, Dpto Ingn Nucl, E-28040 Madrid, Spain
关键词
D O I
暂无
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The responses of TLD-100. TLD-700 and TLD-600 thermoluminescence dosemeters to the radiation field inside a water tank enclosing an isotopic Am-241-Be neutron Source are analysed. Separate contributions coming from thermal neutrons, neutrons with energies above thermal and gamma rays to the total response of the three types of TLD are obtained, This is accomplished by assuming that the gamma responses for materials with different Li-6 enrichments are identical and that the neutron response of TLD-700 is negligible compared to TLD-100 and TLD-600, The last assumption is tested by Monte Carlo simulations of the neutron energy spectrum at the points where the TLDs are irradiated.