Standardizing Radiation Exposure during Cardiac Catheterization in Children with Congenital Heart Disease: Data from a Multicenter Brazilian Registry

被引:5
作者
Manica, Joao Luiz [1 ]
Duarte, Vanessa Oliveira [1 ]
Ribeiro, Marcelo [2 ]
Hartley, Adam [3 ]
Petraco, Ricardo [3 ]
Pedra, Carlos [2 ]
Rossi, Raul [1 ]
机构
[1] Fundacao Univ Cardiol IC FUC, Inst Cardiol, Ave Princesa Isabel 395, BR-90620000 Porto Alegre, RS, Brazil
[2] Hosp Coracao, Sao Paulo, SP, Brazil
[3] Hammersmith Hosp, Imperial Coll London, Natl Heart & Lung Inst, London, England
关键词
Diagnostic; Imaging/methods; Radiation Exposure Pathways; Heart Defects; Congenital; Cardiac Catheterization/methods; Child; DOSE BENCHMARKS; LARGE-VOLUME; INTERVENTION; PROGRAM;
D O I
10.36660/abc.20190012
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: In recent years the increasing number of interventional procedures has resulted in growing concerns regarding radiation exposure for patients and staff. The evaluation of radiation exposure in children is difficult due to the great variability in body weight. Therefore, reference levels of radiation are not well defined for this population. Objectives: To study and validate the ratio of dose-area product (DAP) to patient weight as a reference measurement of radiation for hemodynamic congenital heart disease procedures in children. Methods: This observational multicenter study uses data obtained from a Brazilian registry of cardiac catheterization for congenital heart disease from March 2013 to June 2014. Inclusion criteria were all patients aged <18 years old undergoing hemodynamic procedures for congenital heart disease, with recorded DAP doses. P-value <0.05 was considered as statistically significant. Results: This study evaluated 429 patients with median age and weight of 50 (10, 103) months and 15 (7, 28) kg, respectively. Median DAP was 742.2 (288.8, 1,791.5) mu Gy.m2. There was a good correlation between DAP and weight-fluoroscopic time product(rs=0.66). No statistically significant difference was observed in DAP/weight ratio between therapeutic and diagnostic procedures. There was a wide variation in the DAP/weight ratio among the therapeutic procedures (p<0.001). Conclusions: The DAP/weight ratio is the simplest and most applicable measurement to evaluate radiation exposure in a pediatric population. Although there is limited literature available, the doses obtained in the present study were similar to those previously found. Ongoing research is important to evaluate the impact of strategies to reduce radiation exposure in this population
引用
收藏
页码:1154 / 1160
页数:7
相关论文
共 15 条
[11]   Diagnostic reference levels and effective dose in paediatric cardiac catheterization [J].
Onnasch, D. G. W. ;
Schroeder, F. K. ;
Fischer, G. ;
Kramer, H-H .
BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (951) :177-185
[12]   Quantification of Temporal, Procedural, and Hardware-Related Factors Influencing Radiation Exposure During Pediatric Cardiac Catheterization [J].
Smith, Benjamin G. ;
Tibby, Shane M. ;
Qureshi, Shakeel A. ;
Rosenthal, Eric ;
Krasemann, Thomas .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 80 (06) :931-936
[13]  
Strauss Keith J, 2006, J Am Coll Radiol, V3, P686, DOI 10.1016/j.jacr.2006.04.008
[14]   Pilot program on patient dosimetry in pediatric interventional cardiology in Chile [J].
Ubeda, Carlos ;
Vano, Eliseo ;
Miranda, Patricia ;
Leyton, Fernando .
MEDICAL PHYSICS, 2012, 39 (05) :2424-2430
[15]   Characterization of radiation exposure and effect of a radiation monitoring policy in a large volume pediatric cardiac catheterization lab [J].
Verghese, George R. ;
McElhinney, Doff B. ;
Strauss, Keith J. ;
Bergersen, Lisa .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 79 (02) :294-301